Calibration of Two Validated SARS-CoV-2 Pseudovirus Neutralization Assays for COVID-19 Vaccine Evaluation

Yunda Huang, Oleg Borisov, Jia Jin Kee, Lindsay N. Carpp, Terri Wrin, Suqin Cai, Marcella Sarzotti‐Kelsoe, Charlene McDanal, Amanda Eaton, Rolando Pajón, John Hural, Christine M. Posavad, Katherine Gill, Shelly Karuna, Lawrence Corey, M. Juliana McElrath, Peter B. Gilbert, Christos J. Petropoulos, David C. Montefiori

Abstract

Vaccine-induced neutralizing antibodies (nAbs) are key biomarkers considered to be associated with vaccine efficacy. In United States Government-sponsored phase 3 efficacy trials of COVID-19 vaccines, nAbs are measured by two different validated pseudovirus-based SARS-CoV-2 neutralization assays, with each trial using one of the two assays. Here we describe and compare the nAb titers obtained in the two assays. We observe that one assay consistently yielded higher nAb titers than the other when both assays were performed on the World Health Organization’s anti-SARS-CoV-2 immunoglobulin International Standard, COVID-19 convalescent sera, and mRNA-1273 vaccinee sera. To overcome the challenge this difference in readout poses in comparing/combining data from the two assays, we evaluate three calibration approaches and show that readouts from the two assays can be calibrated to a common scale. These results may aid decision-making based on data from these assays for the evaluation and licensure of new or adapted COVID-19 vaccines.

Introduction

In the global effort to develop and deploy COVID-19 vaccines, neutralizing antibodies (nAbs) have emerged as a critical correlate of protection. These measurable biomarkers provide essential data for evaluating vaccine efficacy and supporting regulatory decisions. However, a significant challenge arose during the U.S. government-sponsored Phase 3 trials: the use of different laboratory assays. Specifically, trials utilized either an assay developed at Duke University or one from Monogram Biosciences.

While both are validated pseudovirus-based assays, they do not produce equivalent results. The Monogram assay consistently reports antibody titers that are 3 to 5 times higher than those from the Duke assay. This systematic discrepancy hinders the direct comparison of different vaccine platforms (e.g., mRNA vs. viral vector) and complicates "immunobridging" efforts—the process of approving boosters or pediatric doses based on antibody levels rather than large-scale clinical trials.

Methods

The researchers investigated the technical basis for the discrepancy between the Duke and Monogram assays. Both assays utilize "pseudoviruses"—harmless viral vectors displaying the SARS-CoV-2 Spike protein—and a luciferase reporter system. A key difference identified was the method of introducing the TMPRSS2 protein, which facilitates viral entry into cells.

To harmonize the data, three calibration approaches were evaluated:

  1. WHO Standard (Approach 1): Utilizing the World Health Organization’s International Standard (IU/mL) to calculate a simple multiplicative conversion factor.
  2. Bivariate Normal (BIVN) (Approach 2): A statistical model that aligns the two scales while accounting for an estimated 30% measurement error in both laboratories.
  3. Linear Regression (Approach 3): A standard mathematical regression to map Duke’s results directly onto the Monogram scale.

The study utilized three datasets: the WHO International Standard (1,000 IU/mL), a convalescent panel of 248 serum samples, and a vaccinee panel of 90 samples from the Moderna mRNA-1273 Phase 1 trial.

Results

The analysis confirmed that the difference between the labs was systematic. Monogram titers were approximately 2.8 times higher for ID50 (the dilution inhibiting 50% of the virus) and 5.2 times higher for ID80 compared to Duke titers across convalescent samples. This discrepancy is likely due to how TMPRSS2 is expressed in the cell lines, highlighting the sensitivity of biological assays to minor protocol variations.

Calibration successfully bridged this gap. Approach 2 (BIVN) provided the highest level of agreement, achieving a Concordance Correlation Coefficient (CCC) of 0.87 for ID50 titers. Significantly, Approach 1 (using the WHO International Standard) performed nearly as well as the more complex models. This finding supports the use of the WHO standard for global harmonization, as it allows labs to convert "titer units" into a universal "International Units" (IU/mL) scale. The calibration factors remained consistent across both naturally infected and vaccinated samples, indicating the robustness of the method.

Discussion and Conclusion

This study provides a "Rosetta Stone" for COVID-19 vaccine data, allowing for the direct comparison of results from different clinical trials. By demonstrating that systematic differences between validated assays can be corrected, the researchers have established a framework for more efficient vaccine evaluation.

While the study focused on the G614 variant and the mRNA platform, the methodology can be adapted for new variants of concern, such as Delta and Omicron, and other vaccine technologies. Future work will involve validating these methods with viral vector and protein subunit vaccines and comparing pseudovirus results with live-virus neutralization assays. Ultimately, these calibration tools are essential for the rapid development and licensure of next-generation COVID-19 vaccines.

Summary updated

Abstract

Vaccine-induced neutralizing antibodies (nAbs) are key biomarkers considered to be associated with vaccine efficacy. In United States Government-sponsored phase 3 efficacy trials of COVID-19 vaccines, nAbs are measured by two different validated pseudovirus-based SARS-CoV-2 neutralization assays, with each trial using one of the two assays. Here we describe and compare the nAb titers obtained in the two assays. We observe that one assay consistently yielded higher nAb titers than the other when both assays were performed on the World Health Organization’s anti-SARS-CoV-2 immunoglobulin International Standard, COVID-19 convalescent sera, and mRNA-1273 vaccinee sera. To overcome the challenge this difference in readout poses in comparing/combining data from the two assays, we evaluate three calibration approaches and show that readouts from the two assays can be calibrated to a common scale. These results may aid decision-making based on data from these assays for the evaluation and licensure of new or adapted COVID-19 vaccines.
BibTeX
@inproceedings{huang2021calibration,
  author = {Yunda Huang and Oleg Borisov and Jia Jin Kee and Lindsay N. Carpp and Terri Wrin and Suqin Cai and Marcella Sarzotti‐Kelsoe and Charlene McDanal and Amanda Eaton and Rolando Pajón and John Hural and Christine M. Posavad and Katherine Gill and Shelly Karuna and Lawrence Corey and M. Juliana McElrath and Peter B. Gilbert and Christos J. Petropoulos and David C. Montefiori},
  title = {{Calibration of Two Validated SARS-CoV-2 Pseudovirus Neutralization Assays for COVID-19 Vaccine Evaluation}},
  booktitle = {Scientific},
  year = {2021},
  doi = {10.1038/s41598-021-03154-6},
}
Figure