BrainZ-BP: A Non-invasive Cuff-less Blood Pressure Estimation Approach Leveraging Brain Bio-impedance and Electrocardiogram

Bufang Yang, Le Liu, Wenxuan Wu, Mengliang Zhou, Hongxing Liu, Xinbao Ning

Abstract

Accurate and continuous blood pressure (BP) monitoring is essential to the early prevention of cardiovascular diseases. Non-invasive and cuff-less BP estimation algorithm has gained much attention in recent years. Previous studies have demonstrated that brain bio-impedance (BIOZ) is a promising technique for non-invasive intracranial pressure (ICP) monitoring. Clinically, treatment for patients with traumatic brain injuries (TBI) requires monitoring the ICP and BP of patients simultaneously. Estimating BP by brain BIOZ directly can reduce the number of sensors attached to the patients, thus improving their comfort. To address the issues, in this study, we explore the feasibility of leveraging brain BIOZ for BP estimation and propose a novel cuff-less BP estimation approach called BrainZ-BP. Two electrodes are placed on the forehead and occipital bone of the head in the anterior-posterior direction for brain BIOZ measurement. Various features including pulse transit time and morphological features of brain BIOZ are extracted and fed into four regression models for BP estimation. Results show that the mean absolute error, root mean square error, and correlation coefficient of random forest regression model are 2.17 mmHg, 3.91 mmHg, and 0.90 for systolic pressure estimation, and are 1.71 mmHg, 3.02 mmHg, and 0.89 for diastolic pressure estimation. The presented BrainZ-BP can be applied in the brain BIOZ-based ICP monitoring scenario to monitor BP simultaneously.

Introduction

Continuous blood pressure (BP) monitoring is a cornerstone of cardiovascular disease management and critical care, particularly for patients with traumatic brain injuries (TBI). While inflatable cuffs are the standard for non-invasive measurement, they are intermittent and can cause patient discomfort. Invasive arterial lines provide continuous data but carry risks of infection and vascular damage.

Recent research has focused on cuff-less solutions, primarily utilizing Pulse Transit Time (PTT) derived from electrocardiogram (ECG) and photoplethysmogram (PPG) sensors. However, PPG is sensitive to skin pigmentation and ambient light. Bio-impedance (BIOZ), which measures tissue impedance changes due to blood volume fluctuations, offers a robust alternative. This paper introduces BrainZ-BP, a novel method that estimates systemic blood pressure directly from brain BIOZ, offering a potential path toward integrated ICP and BP monitoring using a single sensor set.

System Architecture and Methodology

The BrainZ-BP system utilizes a multi-stage pipeline for signal acquisition and processing. As shown in the system architecture, the approach integrates a brain BIOZ module and an ECG module.

The BrainZ-BP system architecture, including the signal acquisition modules for brain BIOZ and ECG, data preprocessing, and the BP estimation pipeline.

Signal Acquisition

Brain BIOZ is measured using a two-electrode setup placed on the forehead and the occipital bone. A safe, low-amplitude 10 kHz current is injected to capture impedance variations caused by cerebral blood flow. Simultaneously, a standard Lead-I ECG is recorded from the wrists to provide a timing reference for the cardiac cycle.

Preprocessing and Feature Extraction

Raw signals are filtered using Savitzky-Golay and FIR bandpass filters to remove respiratory and motion artifacts. From the cleaned waveforms, a comprehensive set of 42 features is extracted. These include traditional PTT-based features—the delay between the ECG R-peak and the BIOZ pulse arrival—as well as detailed morphological and complexity features.

Detailed view of the ECG and brain BIOZ waveforms, highlighting the extracted PTT and morphological features used for BP estimation.

Experimental Results

The study involved 13 healthy participants, generating a dataset of 1,942 synchronized recordings. To ensure a robust range of BP values, subjects performed physical exercise before measurements. Four regression models were evaluated: Linear Regression (LR), Support Vector Machine (SVM), Decision Tree (DT), and Random Forest (RF).

The Random Forest model demonstrated superior performance, achieving a Mean Absolute Error (MAE) of 2.17 mmHg for Systolic Blood Pressure (SBP) and 1.71 mmHg for Diastolic Blood Pressure (DBP). These results comfortably meet the British Hypertension Society (BHS) "Grade A" standard (MAE ≤ 5 mmHg) and the Association for the Advancement of Medical Instrumentation (AAMI) requirements.

Discussion and Conclusion

A key finding of this research is that morphological features of the brain BIOZ signal, such as pulse width and slope ratios, contribute significantly to estimation accuracy, often exceeding the importance of PTT alone. The selection of a 10 kHz frequency was found to be optimal for balancing signal quality and skin-electrode impedance.

While the current model requires personalized calibration, BrainZ-BP represents a significant step toward non-invasive, continuous BP monitoring. Its primary application lies in neuro-critical care, where it could simplify patient monitoring by leveraging existing brain-monitoring electrodes to provide real-time blood pressure data. Future work will focus on developing generalized models and validating the system in clinical populations with fluctuating intracranial pressure.

Summary updated

Abstract

Accurate and continuous blood pressure (BP) monitoring is essential to the early prevention of cardiovascular diseases. Non-invasive and cuff-less BP estimation algorithm has gained much attention in recent years. Previous studies have demonstrated that brain bio-impedance (BIOZ) is a promising technique for non-invasive intracranial pressure (ICP) monitoring. Clinically, treatment for patients with traumatic brain injuries (TBI) requires monitoring the ICP and BP of patients simultaneously. Estimating BP by brain BIOZ directly can reduce the number of sensors attached to the patients, thus improving their comfort. To address the issues, in this study, we explore the feasibility of leveraging brain BIOZ for BP estimation and propose a novel cuff-less BP estimation approach called BrainZ-BP. Two electrodes are placed on the forehead and occipital bone of the head in the anterior-posterior direction for brain BIOZ measurement. Various features including pulse transit time and morphological features of brain BIOZ are extracted and fed into four regression models for BP estimation. Results show that the mean absolute error, root mean square error, and correlation coefficient of random forest regression model are 2.17 mmHg, 3.91 mmHg, and 0.90 for systolic pressure estimation, and are 1.71 mmHg, 3.02 mmHg, and 0.89 for diastolic pressure estimation. The presented BrainZ-BP can be applied in the brain BIOZ-based ICP monitoring scenario to monitor BP simultaneously.