Biocompatibility characterisation of CMOS-based Lab-on-Chip electrochemical sensors for in vitro cancer cell culture applications

Melina Beykou, Vicky Bousgouni, Nicolas Moser, Pantelis Georgiou, Chris Bakal

Abstract

Lab-on-Chip electrochemical sensors, particularly Ion-Sensitive Field-Effect Transistors (ISFETs), are emerging as powerful tools for point-of-care diagnostics and real-time monitoring of the tumour microenvironment. These sensors leverage standard Complementary Metal Oxide Semiconductor (CMOS) technology, using the Silicon Nitride ($Si_3N_4$) passivation layer as a sensing surface. However, the biocompatibility of $Si_3N_4$ remains poorly understood. This study quantitatively evaluates the adhesion, morphogenesis, proliferation, and mechano-responsiveness of normal (MCF10A) and cancer (MDA-MB-231) cells on $Si_3N_4$ surfaces. Our findings reveal that while raw $Si_3N_4$ decreases cell adhesion and mechanical signaling (YAP/TAZ activation), it paradoxically increases proliferation markers (Ki67). Crucially, we demonstrate that surface engineering with extracellular matrix components, specifically Poly-L-ornithine and Laminin (PLOL), can restore cellular phenotypes to levels comparable to standard culture plastic, validating these sensors for future cell-based assays.

Introduction

The integration of electrochemical sensors with CMOS technology has paved the way for scalable, high-resolution platforms capable of monitoring physiological parameters like pH in real-time. ISFETs are particularly suited for detecting the acidic shifts characteristic of tumour microenvironments. Despite their potential, a critical gap exists: the interface between the inorganic microchip surface and organic cellular life. Most CMOS chips are passivated with $Si_3N_4$, a material that may induce cellular stress or alter biological signaling. To ensure that data collected by these sensors reflects true biological phenomena rather than artifacts of surface-induced stress, a comprehensive biocompatibility characterisation is required.

Methodology

To bridge the gap between semiconductor engineering and cell biology, researchers compared cellular behavior on standard tissue culture plastic against a 4,368-pixel ISFET array.

Cell Models and Surface Engineering

The study utilised two distinct mammary cell lines: the triple-negative breast cancer line MDA-MB-231 and the non-tumorigenic MCF10A line. To improve the "hospitality" of the $Si_3N_4$ surface, several coatings were tested, including Collagen Type I, Fibronectin (FN), and a combination of Poly-L-ornithine and Laminin (PLOL).

Imaging and Analysis

High-throughput confocal microscopy and Total Internal Reflection Fluorescence (TIRF) microscopy were employed to visualize focal adhesions—the protein complexes cells use to grip a substrate. Data analysis was supported by an AI-driven pipeline that used a linear classifier to categorize cell morphology (round, elongated, or star-shaped) and quantify protein expression across thousands of individual cells.

Key Insights

The Impact of Surface Topography

The physical architecture of the ISFET pixels—small rectangular structures—was found to guide cell growth. Cells tended to elongate parallel to the pixel axes, suggesting that the micro-topography of the circuit influences morphogenesis. This "scaffolding" effect must be accounted for when interpreting morphological data from chip-based assays.

Mechanical Signaling and Proliferation

On untreated $Si_3N_4$, cells exhibited a significant decrease in the activation of YAP/TAZ, transcription regulators that serve as internal "tension gauges." This suggests that cells perceive the raw microchip surface as less adhesive or mechanically "softer" than standard plastic. Interestingly, while fewer cells adhered to the chip, those that did showed a 37% higher Ki67/Hoechst ratio, indicating increased proliferation. This suggests that the semiconductor environment may trigger a compensatory growth response in surviving cells.

The PLOL Rescue Effect

Surface treatment proved essential for normalizing cell behavior. While untreated $Si_3N_4$ saw only 50% of the cell attachment found on plastic, coatings like PLOL and Collagen increased cell numbers by 2x to 3x. Specifically, PLOL was identified as a "magic bullet," restoring YAP/TAZ levels and growth rates to levels nearly identical to those observed on standard laboratory plastic.

Results Summary

  • Adhesion: Raw $Si_3N_4$ significantly hinders cell attachment; however, ECM coatings can surpass the adhesiveness of standard plastic.
  • Morphology: Focal adhesions on the chip were smaller in area, indicating that cells must "stretch" more to maintain their grip on the semiconductor surface.
  • Sensitivity: Non-tumorigenic cells (MCF10A) showed less sensitivity to the $Si_3N_4$ surface than cancer cells, highlighting the need for cell-specific surface optimization.

Conclusion and Future Work

This study establishes that while the native $Si_3N_4$ surface of CMOS sensors alters cell biology, these effects can be mitigated through strategic surface engineering. By applying PLOL coatings, researchers can create a physiologically relevant environment that supports normal cell signaling and growth. Future work will focus on optimizing pixel layouts to minimize structural constraints and applying these validated sensors to monitor real-time drug responses in complex cancer models. This characterisation is a vital step toward moving Lab-on-Chip technology from the engineering lab into the clinical oncology suite.

Summary updated

Abstract

Abstract Lab-on-Chip electrochemical sensors, such as Ion-Sensitive Field-Effect Transistors (ISFETs), are being developed for use in point-of-care diagnostics, such as pH detection of tumour microenvironments, due to their integration with standard Complementary Metal Oxide Semiconductor technology. With this approach, the passivation of the CMOS process is used as a sensing layer to minimise post-processing, and Silicon Nitride (Si 3 N 4 ) is the most common material at the microchip surface. ISFETs have the potential to be used for cell-based assays however, there is a poor understanding of the biocompatibility of microchip surfaces. Here, we quantitatively evaluated cell adhesion, morphogenesis, proliferation and mechano-responsiveness of both normal and cancer cells cultured on a Si 3 N 4 , sensor surface. We demonstrate that both normal and cancer cell adhesion decreased on Si 3 N 4 . Activation of the mechano-responsive transcription regulators, YAP/TAZ, are significantly decreased in cancer cells on Si 3 N 4 in comparison to standard cell culture plastic, whilst proliferation marker, Ki67, expression markedly increased. Non-tumorigenic cells on chip showed less sensitivity to culture on Si 3 N 4 than cancer cells. Treatment with extracellular matrix components increased cell adhesion in normal and cancer cell cultures, surpassing the adhesiveness of plastic alone. Moreover, poly-l-ornithine and laminin treatment restored YAP/TAZ levels in both non-tumorigenic and cancer cells to levels comparable to those observed on plastic. Thus, engineering the electrochemical sensor surface with treatments will provide a more physiologically relevant environment for future cell-based assay development on chip.
BibTeX
@inproceedings{beykou2024biocompatibility,
  author = {Melina Beykou and Vicky Bousgouni and Nicolas Moser and Pantelis Georgiou and Chris Bakal},
  title = {{Biocompatibility characterisation of CMOS-based Lab-on-Chip electrochemical sensors for in vitro cancer cell culture applications}},
  booktitle = {Biosensors},
  year = {2024},
  doi = {10.1016/j.bios.2024.116513},
}
Figure