Abstract
With an aging population, numerous assistive and monitoring technologies are under development to enable older adults to age in place. To facilitate aging in place predicting risk factors such as falls, and hospitalization and providing early interventions are important. Much of the work on ambient monitoring for risk prediction has centered on gait speed analysis, utilizing privacy-preserving sensors like radar. Despite compelling evidence that monitoring step length, in addition to gait speed, is crucial for predicting risk, radar-based methods have not explored step length measurement in the home. Furthermore, laboratory experiments on step length measurement using radars are limited to proof of concept studies with few healthy subjects. To address this gap, a radar-based step length measurement system for the home is proposed based on detection and tracking using radar point cloud, followed by Doppler speed profiling of the torso to obtain step lengths in the home. The proposed method was evaluated in a clinical environment, involving 35 frail older adults, to establish its validity. Additionally, the method was assessed in people's homes, with 21 frail older adults who had participated in the clinical assessment. The proposed radar-based step length measurement method was compared to the gold standard Zeno Walkway Gait Analysis System, revealing a 4.5cm/8.3% error in a clinical setting. Furthermore, it exhibited excellent reliability (ICC(2,k)=0.91, 95% CI 0.82 to 0.96) in uncontrolled home settings. The method also proved accurate in uncontrolled home settings, as indicated by a strong agreement (ICC(3,k)=0.81 (95% CI 0.53 to 0.92)) between home measurements and in-clinic assessments.
Introduction
As global populations age, "aging-in-place" strategies have become vital to reducing healthcare burdens. While gait speed is a common health metric, step length is often a more sensitive predictor of frailty, fall risk, and mortality. However, measuring step length accurately outside clinical settings is challenging. Pressure-sensitive mats are expensive and limited in scope, wearables suffer from low compliance, and cameras raise privacy concerns.
Radar offers a privacy-preserving alternative, but previous research has largely been confined to controlled laboratory settings with healthy subjects. This paper presents a system designed to measure the irregular, short-distance walking patterns of frail older adults in cluttered, real-world home environments.
Methodology
The system utilizes a Frequency Modulated Continuous Wave (FMCW) radar (specifically the TI IWR6843AOP) to track movement through a multi-stage pipeline:
- Detection and Tracking: Radar points are clustered using DBSCAN, associated via the Hungarian algorithm, and tracked in 3D using Kalman filtering.
- Track Segmentation: To maximize signal quality, the Ramer-Douglas-Peucker (RDP) algorithm simplifies walking paths into linear segments. The system prioritizes segments aligned with the radar’s radial axis, where Doppler sensitivity is highest.
- Torso Speed Profiling: Instead of tracking feet, which produce weak echoes, the system isolates the torso by filtering for points at an elevation of approximately 1.2m.
- Peak-to-Peak Analysis: Step length is derived from cyclical fluctuations in torso speed. As the center of mass (CoM) accelerates and decelerates during a gait cycle, the distance between these speed peaks corresponds to a single step.
Experimental Setup and Datasets
The study was conducted in two primary phases to ensure both accuracy and real-world applicability:
- Clinical Validation: 35 frail older adults performed various walking tasks (Normal, Fast, Narrow, Obstacle, and Dual-Task) over a 4-meter Zeno Walkway pressure mat, serving as the gold standard.
- Home Evaluation: 21 participants installed radar sensors in their homes (bedrooms, living rooms, or kitchens) for two weeks, capturing passive, uncontrolled walking data.
Results and Discussion
The proposed method demonstrated high accuracy and reliability across both environments:
- Accuracy: In clinical trials, the radar system achieved an average error of 4.5 cm (8.3%) for normal walking compared to the Zeno Walkway.
- Reliability: In home settings, the system showed excellent test-retest reliability with an ICC(2,k) of 0.91.
- Validity: A strong agreement (ICC = 0.81) was found between home-based passive measurements and in-clinic assessments, confirming that the "in the wild" data accurately reflects clinical status.
Key insights include the robustness of torso-based tracking over foot tracking and the importance of including acceleration/deceleration phases, which are common in short home walks. However, the system's performance is sensitive to home layout; paths perpendicular to the radar are harder to analyze, suggesting that future iterations might benefit from multi-sensor fusion to ensure radial coverage.
Conclusion
This research successfully transitions gait analysis from the lab to the home. By leveraging FMCW radar and torso speed profiling, the system provides a reliable, privacy-preserving method for long-term monitoring of step length in frail older adults, offering a scalable tool for early intervention and fall risk prediction.
Summary updated