Augmented reality‐based feedback for technician‐in‐the‐loop C‐arm repositioning

Mathias Unberath, Javad Fotouhi, Jonas Hajek, Andreas Maier, Greg Osgood, Russell G. Taylor, Mehran Armand, Nassir Navab

Abstract

This paper presents a marker-free Augmented Reality (AR) system designed to assist X-ray technicians in repositioning C-arm imaging devices during surgery. By utilizing a head-mounted display to capture and visualize 3D "ghost" poses of previous C-arm positions, the system aims to eliminate "trial-and-error" radiation exposure. Experimental results demonstrate that AR guidance can reduce the number of scout X-rays to zero while maintaining clinically acceptable angular accuracy, significantly lowering radiation risk for patients and surgical staff.

Introduction

In minimally invasive orthopedic procedures, surgeons rely on mobile C-arm X-ray machines to verify implant placement. Achieving specific anatomical views (e.g., "inlet" or "obturator" views) requires precise repositioning of the C-arm, which has multiple degrees of freedom. Because internal anatomy is not visible externally, technicians often perform "fluoro hunting"—taking multiple scout X-rays to find the correct position. This iterative process increases both surgical time and cumulative radiation dose. Current tracking solutions often require expensive external hardware or invasive markers, which are impractical in fast-paced clinical environments.

Methodology

The proposed "technician-in-the-loop" solution employs a marker-free AR approach. The system utilizes an optical see-through head-mounted display (OST HMD) equipped with an integrated infrared (IR) depth sensor.

Coordinate systems and transformations used for AR-based C-arm localization and guidance.

As shown in the coordinate system diagram above, the system uses Simultaneous Localization and Mapping (SLAM) to track the technician's head movement relative to the world coordinate system ($^T T_W$). When a desired C-arm pose is achieved, the IR sensor captures a 3D point cloud of the device's surface, which is stored as a reference pose ($^P T_{IR}$). This allows the system to "bookmark" specific views without requiring external tracking infrastructure.

Guidance and Visualization

To restore a previously saved view, the system projects the stored 3D point cloud—a virtual "ghost"—back into the technician's field of view. The technician then manually adjusts the physical C-arm until it aligns with the virtual green overlay.

AR visualization workflow: (a) recorded pose in red, (b) target 'ghost' pose in green, and (c) final alignment of the physical C-arm with the virtual ghost.

This visual feedback loop treats the technician as an active controller, relying on human perception to achieve alignment rather than fully automated robotics. This approach prevents cognitive errors, such as restoring the wrong anatomical view, by providing a clear visual target.

Results and Discussion

The system was evaluated using an anthropomorphic pelvis phantom across six clinical C-arm poses. Key findings include:

  • Radiation Reduction: The conventional manual approach required an average of 2.76 scout X-rays per view. With AR guidance, this was reduced to zero, as the technician could align the device visually.
  • Accuracy: The AR system achieved an angular error of 1.54° ± 0.92°, comparable to the manual method (1.23° ± 0.45°). While positional error was higher (51.6 mm vs. 16.7 mm), angular accuracy is generally considered more critical for correct anatomical projection in C-arm imaging.

Conclusion

This marker-free AR system demonstrates the potential to eliminate scout X-rays during C-arm repositioning. By leveraging the technician's visual perception through a "ghosting" interface, the system provides a flexible, low-cost alternative to traditional tracking methods. Future work will focus on improving positional accuracy through algorithms like Iterative Closest Point (ICP) and validating the system in dynamic clinical environments with moving staff and surgical drapes.

Summary updated

Abstract

Interventional C-arm imaging is crucial to percutaneous orthopedic procedures as it enables the surgeon to monitor the progress of surgery on the anatomy level. Minimally invasive interventions require repeated acquisition of X-ray images from different anatomical views to verify tool placement. Achieving and reproducing these views often comes at the cost of increased surgical time and radiation. We propose a marker-free 'technician-in-the-loop' Augmented Reality (AR) solution for C-arm repositioning. The X-ray technician operating the C-arm interventionally is equipped with a head-mounted display system capable of recording desired C-arm poses in 3D via an integrated infrared sensor. For C-arm repositioning to a target view, the recorded pose is restored as a virtual object and visualized in an AR environment, serving as a perceptual reference for the technician. Our proof-of-principle findings from a simulated trauma surgery indicate that the proposed system can decrease the 2.76 X-ray images required for re-aligning the scanner with an intra-operatively recorded C-arm view down to zero, suggesting substantial reductions of radiation dose. The proposed AR solution is a first step towards facilitating communication between the surgeon and the surgical staff, improving the quality of surgical image acquisition, and enabling context-aware guidance for surgery rooms of the future.
BibTeX
@inproceedings{unberath2018augmented,
  author = {Mathias Unberath and Javad Fotouhi and Jonas Hajek and Andreas Maier and Greg Osgood and Russell G. Taylor and Mehran Armand and Nassir Navab},
  title = {{Augmented reality‐based feedback for technician‐in‐the‐loop C‐arm repositioning}},
  booktitle = {Healthcare},
  year = {2018},
  doi = {10.1049/htl.2018.5066},
}
Figure