Abstract
This paper presents a marker-free Augmented Reality (AR) system designed to assist X-ray technicians in repositioning C-arm imaging devices during surgery. By utilizing a head-mounted display to capture and visualize 3D "ghost" poses of previous C-arm positions, the system aims to eliminate "trial-and-error" radiation exposure. Experimental results demonstrate that AR guidance can reduce the number of scout X-rays to zero while maintaining clinically acceptable angular accuracy, significantly lowering radiation risk for patients and surgical staff.
Introduction
In minimally invasive orthopedic procedures, surgeons rely on mobile C-arm X-ray machines to verify implant placement. Achieving specific anatomical views (e.g., "inlet" or "obturator" views) requires precise repositioning of the C-arm, which has multiple degrees of freedom. Because internal anatomy is not visible externally, technicians often perform "fluoro hunting"—taking multiple scout X-rays to find the correct position. This iterative process increases both surgical time and cumulative radiation dose. Current tracking solutions often require expensive external hardware or invasive markers, which are impractical in fast-paced clinical environments.
Methodology
The proposed "technician-in-the-loop" solution employs a marker-free AR approach. The system utilizes an optical see-through head-mounted display (OST HMD) equipped with an integrated infrared (IR) depth sensor.
As shown in the coordinate system diagram above, the system uses Simultaneous Localization and Mapping (SLAM) to track the technician's head movement relative to the world coordinate system ($^T T_W$). When a desired C-arm pose is achieved, the IR sensor captures a 3D point cloud of the device's surface, which is stored as a reference pose ($^P T_{IR}$). This allows the system to "bookmark" specific views without requiring external tracking infrastructure.
Guidance and Visualization
To restore a previously saved view, the system projects the stored 3D point cloud—a virtual "ghost"—back into the technician's field of view. The technician then manually adjusts the physical C-arm until it aligns with the virtual green overlay.
This visual feedback loop treats the technician as an active controller, relying on human perception to achieve alignment rather than fully automated robotics. This approach prevents cognitive errors, such as restoring the wrong anatomical view, by providing a clear visual target.
Results and Discussion
The system was evaluated using an anthropomorphic pelvis phantom across six clinical C-arm poses. Key findings include:
- Radiation Reduction: The conventional manual approach required an average of 2.76 scout X-rays per view. With AR guidance, this was reduced to zero, as the technician could align the device visually.
- Accuracy: The AR system achieved an angular error of 1.54° ± 0.92°, comparable to the manual method (1.23° ± 0.45°). While positional error was higher (51.6 mm vs. 16.7 mm), angular accuracy is generally considered more critical for correct anatomical projection in C-arm imaging.
Conclusion
This marker-free AR system demonstrates the potential to eliminate scout X-rays during C-arm repositioning. By leveraging the technician's visual perception through a "ghosting" interface, the system provides a flexible, low-cost alternative to traditional tracking methods. Future work will focus on improving positional accuracy through algorithms like Iterative Closest Point (ICP) and validating the system in dynamic clinical environments with moving staff and surgical drapes.