An open-source software platform for translational photoacoustic research and its application to motion-corrected blood oxygenation estimation

Thomas Kirchner, Janek Gröhl, Franz Sattler, Moritz S. Bischoff, Angelika Laha, Marco Nolden, Lena Maier-Hein

Introduction

Photoacoustic (PA) imaging is a powerful functional imaging modality that utilizes laser pulses to generate ultrasound waves, enabling the visualization of blood oxygenation ($sO_2$) deep within biological tissues. Despite its potential, translational research in this field is often hindered by two primary challenges: the prevalence of proprietary, closed-source software silos and the significant impact of motion artifacts during multispectral data acquisition. Because $sO_2$ calculation requires sequential imaging at multiple wavelengths, physiological movements such as arterial pulsation or patient respiration can lead to severe misalignment and inaccuracies in oxygenation maps.

System Architecture and Methodology

To address these hurdles, researchers developed a real-time, multispectral hybrid photoacoustic and ultrasonic (PAUS) imaging platform integrated into the Medical Imaging Interaction Toolkit (MITK). This open-source framework utilizes a modular architecture that decouples hardware control (lasers and data acquisition) from processing and application layers, facilitating hardware interoperability and reproducible research.

Experimental setup of the real-time hybrid PAUS system, showing the MITK workstation, ultrasound DAQ, and fast-tuning laser system.

A key innovation of this platform is its approach to motion correction. The system acquires an ultrasound (US) image immediately following each PA image. While PA images vary significantly in intensity across different laser wavelengths, making direct registration difficult, the interleaved US images provide consistent structural data. By applying the Farnebäck optical flow algorithm to these US "anchor" frames, the system calculates motion vectors that are then used to warp and align the corresponding PA images before spectral unmixing.

Experimental Validation

The system was validated through in vivo experiments on healthy human volunteers, focusing on peripheral vessels (radial and ulnar arteries) and the carotid artery. The carotid study, performed by a vascular surgeon, involved complex sequences of 84 multispectral sets across five wavelengths. This high-volume data collection demonstrated the platform's capability to handle clinically relevant imaging scenarios in real-time, with processing speeds reaching approximately 120ms per sequence.

Key Results and Discussion

The implementation of US-based motion correction yielded significant improvements in image quality and data precision. In carotid artery imaging, motion correction effectively suppressed artifacts that otherwise distorted the vessel's appearance. Quantitatively, the precision of arterial blood $sO_2$ measurements improved dramatically, with the Interquartile Range (IQR) dropping from 50% to 25% after correction.

A photoacoustic image at 831nm (left) and the resulting spectral unmixing map (right) showing estimated blood oxygenation levels in tissue.

Functional imaging results showed physiological consistency, clearly distinguishing between arterial and venous oxygenation. In peripheral vessel studies, the system measured an average arterial oxygenation ($SaO_2$) of 72% ± 7% and a venous oxygenation ($SvO_2$) of 38% ± 9%. While the arterial values are lower than the expected physiological norm of ~98% due to light attenuation (fluence effects), the clear differentiation between vessel types validates the system's functional utility.

Conclusion and Future Work

By providing an open-source, standardized platform within MITK, this research offers a catalyst for the broader photoacoustic community to build upon and reproduce translational results. Future work will focus on implementing sophisticated fluence compensation algorithms to correct the systematic underestimation of $sO_2$ and bring measurements closer to absolute physiological values. This platform represents a significant step toward moving photoacoustic imaging from specialized laboratories into routine clinical practice.

Summary updated

Abstract

Photoacoustic (PA) imaging systems based on clinical linear ultrasound arrays have become increasingly popular in translational PA research. Such systems can be more easily integrated in a clinical workflow due to the simultaneous access to ultrasonic imaging and their familiarity of use to clinicians. In contrast to more complex setups, hand held linear probes can be applied to a large variety of clinical use cases. However, most translational work with such scanners is based on proprietary development and as such not accessible to the community. In this contribution, we present a custom-built, hybrid, multispectral, real-time photoacoustic and ultrasonic imaging system with a linear array probe that is controlled by software developed within the Medical Imaging Interaction Toolkit (MITK) a highly customizable and extendable open-source software platform. Our software offers direct control of both the laser and the ultrasonic system and may serve as a starting point for various translational research projects and developments. To demonstrate the applicability of the platform, we used it to implement a new method for blood oxygenation estimation in the presence of non-rigid inter-frame motion caused by pulsing arteries. Initial results from experiments with healthy human volunteers demonstrate the suitability of the method with the sample clinical application of imaging the common carotid artery as well as peripheral extremity vessels.