A Unified Interaction Control Framework for Safe Robotic Ultrasound Scanning with Human-Intention-Aware Compliance

Xiangjie Yan, Shaqi Luo, Yongpeng Jiang, Mingrui Yu, Chen Chen, Senqiang Zhu, Gao Huang, Shiji Song, Xiang Li

Introduction

Ultrasound scanning is a physically demanding clinical task, often requiring sonographers to maintain precise probe orientation and contact force for extended periods. While robotic ultrasound (RUS) systems have been proposed to alleviate this ergonomic burden, the clinical environment presents significant challenges for human-robot collaboration. Doctors frequently need to intervene to adjust the probe or apply coupling gel, and the robot must safely handle accidental collisions or patient movements.

Existing systems typically manage these interactions through "hard switching" between discrete control modes. This approach often results in jerky transitions, instability, and potential safety risks. This paper presents a unified interaction control framework designed to provide smooth, fluid transitions between autonomous scanning and various human-interaction scenarios by distinguishing between intentional guidance and accidental contact.

Methodology

The proposed framework is built upon three core technical pillars: hierarchical compliance control, human-intention-aware weighting, and smooth mode transitions.

Hierarchical Compliance Control

The system utilizes a 7-DOF Franka Panda manipulator, controlled via a two-level hierarchy. The Main Task focuses on the end-effector (the ultrasound probe), maintaining its specific pose and a constant contact force against the patient's skin. The Secondary Task (or null-space) manages the robot's remaining degrees of freedom. By decoupling these tasks, the robot can adjust its body configuration—for instance, moving its "elbow" to avoid an obstacle—without disturbing the probe's position on the patient.

Human-Intention-Aware Weighting

Rather than using binary "if-else" logic to switch modes, the framework employs five continuous weighting factors ($a_h, a_p, a_f, a_n, a_b$). These factors are derived from real-time data provided by an Azure Kinect RGBD camera and an ATI force/torque sensor. They represent:

  • $a_h$: Proximity of a human hand to the probe (grasping intention).
  • $a_p$: Proximity of the probe to the patient.
  • $a_f$: Status of contact force maintenance.
  • $a_n$: External forces detected on the robot's body (null-space).
  • $a_b$: Proximity of a human to the robot's body.

Smooth Mode Transitions and Stability

These weights allow the robot to blend seamlessly between six operational modes: Scanning, Human-Guiding, Null-space Avoiding, Null-space Contacting, Waiting, and Recovery. Because the weights change continuously, the transitions are mathematically smooth. To ensure the system remains stable during these transitions, the authors implemented "energy tank" methods, which prove that the system remains passive and will not spontaneously gain energy, even as stiffness and damping parameters fluctuate.

Experimental Validation

The framework was evaluated using a real-world experimental setup involving a 7-DOF Franka Panda, an ATI mini40 force/torque sensor, and an Azure Kinect camera. A user study was conducted with nine male volunteers to validate the system's robustness during carotid artery examinations across different body types.

Key Results

  • Precision and Stability: During autonomous scanning, the robot maintained a stable contact force of approximately 10N. The tracking error remained below 0.04m, even when the subjects moved or turned their heads.
  • Seamless Interaction: The system demonstrated high sensitivity to human intention. When a doctor approached to apply gel, the robot transitioned to "Human-Guiding Mode" instantly and resumed scanning without error once the doctor released the probe.
  • Collision Rejection: A critical finding was the robot's ability to handle accidental impacts. When the robot's arm was bumped during a scan, the null-space stiffness dropped to zero to absorb the impact. Crucially, the force and position of the ultrasound probe on the patient remained almost entirely unaffected, demonstrating the effectiveness of the hierarchical decoupling.

Discussion and Future Work

While the framework successfully demonstrates fluid interaction and safety, several areas for improvement remain. The current method of inferring intention relies primarily on distance and force thresholds. Future iterations could incorporate more sophisticated human intention estimation, such as gesture recognition or physiological signals, to make interactions feel more natural.

Additionally, the experimental environment was controlled; a crowded surgical suite with multiple personnel and equipment would require more robust computer vision to distinguish between different types of obstacles. Finally, expanding the framework to cover more complex anatomical regions—where respiratory motion or deeper tissue contact is required—will be essential for broader clinical adoption.

Summary updated

Abstract

The ultrasound scanning robot operates in environments where frequent human-robot interactions occur. Most existing control methods for ultrasound scanning address only one specific interaction situation or implement hard switches between controllers for different situations, which compromises both safety and efficiency. In this paper, we propose a unified interaction control framework for ultrasound scanning robots capable of handling all common interactions, distinguishing both human-intended and unintended types, and adapting with appropriate compliance. Specifically, the robot suspends or modulates its ongoing main task if the interaction is intended, e.g., when the doctor grasps the robot to lead the end effector actively. Furthermore, it can identify unintended interactions and avoid potential collision in the null space beforehand. Even if that collision has happened, it can become compliant with the collision in the null space and try to reduce its impact on the main task (where the scan is ongoing) kinematically and dynamically. The multiple situations are integrated into a unified controller with a smooth transition to deal with the interactions by exhibiting human-intention-aware compliance. Experimental results validate the framework’s ability to cope with all common interactions including intended intervention and unintended collision in a collaborative carotid artery ultrasound scanning task.
BibTeX
@inproceedings{yan2024unified,
  author = {Xiangjie Yan and Shaqi Luo and Yongpeng Jiang and Mingrui Yu and Chen Chen and Senqiang Zhu and Gao Huang and Shiji Song and Xiang Li},
  title = {{A Unified Interaction Control Framework for Safe Robotic Ultrasound Scanning with Human-Intention-Aware Compliance}},
  booktitle = {IEEE/RSJ International Conference on Intelligent Robots and Systems},
  year = {2024},
  url = {https://openreview.net/forum?id=pJim3STf3z},
}
Figure