A Protocol for Producing Quality Control Materials for Hematology Testing Using Porcine and Goose Blood v2

N. H. Vo, Hoanh Tran, Tung Le, Thuy Hang Nguyen, D.d. Vu, My T. Thai, T.t.p. Nguyen

TL;DR

This protocol provides a cost-effective, locally adaptable method for producing hematology internal quality control (IQC) materials using porcine and goose blood. By employing a two-step chemical stabilization process, the researchers created simulated white blood cell (WBC) samples that remain stable for over 60 days, offering a sustainable alternative to expensive, short-lived commercial controls in resource-limited settings.

What Problem Does This Paper Address?

In clinical hematology, Internal Quality Control (IQC) is essential to ensure that automated blood counters are providing accurate results. However, most laboratories rely on commercial IQC materials that are expensive, must be imported, and have a very limited shelf life (often less than a month).

For laboratories in Low- and Middle-Income Countries (LMICs), these factors create a significant financial and logistical barrier to maintaining high diagnostic standards. There is a dire need for a "home-grown" solution: a method to produce stable, reliable control materials using locally available biological sources that can mimic human blood parameters on automated analyzers.

Methods

The authors detail two distinct experiments to create "simulated leukocytes" (white blood cell analogs) using animal blood sourced from legal commercial slaughterhouses.

  1. Goose Erythrocyte-Derived Simulated Leukocytes:
    • The Logic: Avian red blood cells are nucleated, which allows them to mimic the size and complexity of human white blood cells on many automated counters.
  2. Process: Goose blood is collected in CPDA-1 anticoagulant. The red blood cells (erythrocytes) are isolated via centrifugation and washed with Phosphate-Buffered Saline (PBS). They are then fixed using a stabilizing solution of 25% glutaraldehyde and neomycin sulfate (an antibiotic to prevent microbial growth) for one hour. The final product is suspended in 0.85% NaCl.
  3. Porcine Leukocyte-Derived Simulated Leukocytes:
    • The Logic: Using actual mammalian white blood cells but stabilizing them to prevent degradation.
  4. Process: Porcine blood undergoes a controlled hemolysis step (using a hemolysis solution and 0.1% acetic acid at 40°C) to remove the red blood cells, leaving the leukocytes behind. These leukocytes are then subjected to a "two-step stabilization": they are washed and then fixed overnight in a glutaraldehyde-based solution.

The protocol emphasizes the importance of gentle stirring and precise temperature control (40°C during porcine hemolysis) to preserve cellular morphology.

Datasets

As this is a laboratory protocol rather than a data-mining study, "datasets" refer to the biological sources and the hardware used for validation:

  • Biological Sources: Blood from healthy geese (>5 months old) and porcine blood from commercial sources.
  • Validation Platforms: The method was primarily optimized for the Abbott Cell-Dyn Emerald 22 analyzer.
  • Cross-Platform Validation: The materials were also tested and validated on Sysmex and Horiba systems to ensure broader applicability.

Key Insights and Ideas

  • The "Two-Step" Advantage: The core innovation is the chemical fixation process. By using glutaraldehyde to cross-link proteins, the researchers "freeze" the cellular structure, preventing the natural decay that usually limits the shelf life of blood controls.
  • Stability is King: While fresh blood controls might last a few days, this protocol achieves a stability window of 60 days when stored at 2–8°C. This is a game-changer for labs that cannot afford frequent shipments of commercial reagents.
  • Accessibility: By using slaughterhouse by-products (pig and goose blood), the raw materials are essentially zero-cost, making the protocol highly sustainable for local centers for standardization.
  • Universal Design: Although optimized for one specific analyzer (Abbott), the successful validation on Sysmex and Horiba suggests the simulated cells are robust enough to be recognized correctly by different impedance and optical technologies used in various hematology suites.

Key Results

  • Longevity: The prepared IQC materials maintain stable WBC parameters for at least 60 days.
  • Morphology Preservation: The chemical fixation successfully preserves the size and internal complexity of the cells, allowing automated analyzers to categorize them as leukocytes consistently.
  • Cost Reduction: The paper presents this as a "low-cost" alternative, though specific per-unit cost comparisons are not detailed, the reliance on local animal blood and standard lab chemicals (glutaraldehyde, PBS) implies a fraction of the cost of commercial equivalents.

Limitations and Future Work

  • Parameter Specificity: The abstract and protocol focus heavily on WBC (White Blood Cell) parameters. It is not explicitly discussed how well these materials simulate Red Blood Cell (RBC) indices or Platelet (PLT) counts over the same 60-day period.
  • Manual Labor: While the materials are cheap, the protocol requires significant "hands-on" time (overnight incubation, multiple wash steps), which requires trained laboratory personnel.
  • Future Directions: The authors suggest that while this version (V.1) is optimized for specific analyzers, further iterations could refine the stabilization process to better mimic differential counts (Neutrophils, Lymphocytes, etc.) or extend the shelf life even further.

Summary updated

Abstract

Commercial hematology internal quality control (IQC) materials are often costly, imported, and have limited shelf life, which poses a significant challenge for laboratories in low- and middle-income countries (LMICs). This protocol presents a low-cost, locally adaptable method for preparing IQC materials targeting white blood cell (WBC) parameters. Using porcine and goose blood sourced from legal commercial slaughterhouses, the samples are chemically fixed through a two-step stabilization process to preserve cellular morphology and WBC stability over 60 days. The method has been optimized for the Abbott Cell-Dyn Emerald 22 analyzer and validated across other platforms, including Sysmex and Horiba systems, making it applicable in diverse laboratory settings.
BibTeX
@misc{vo2025protocol,
  author = {N. H. Vo and Hoanh Tran and Tung Le and Thuy Hang Nguyen and D.D. Vu and My T. Thai and T.T.P. Nguyen},
  title = {{A Protocol for Producing Quality Control Materials for Hematology Testing Using Porcine and Goose Blood v2}},
  year = {2025},
  doi = {10.17504/protocols.io.n2bvjex8ngk5/v2},
}

Loading related content...

Figure