A Micro-Power EEG Acquisition SoC With Integrated Feature Extraction Processor for a Chronic Seizure Detection System
Lists
Only you see this
Abstract
Abstract—This paper presents a low-power SoC that performs EEG acquisition and feature extraction required for continuous detection of seizure onset in epilepsy patients. The SoC corresponds to one EEG channel, and, depending on the patient, up to 18 channels may be worn to detect seizures as part of a chronic treatment system. The SoC integrates an instrumentation amplifier, ADC, and digital processor that streams features-vectors to a central device where seizure detection is performed via a machine-learning classifier. The instrumentation-amplifier uses chopper-stabilization in a topology that achieves high input-impedance and rejects large electrode-offsets while operating at 1 V; the ADC employs power-gating for low energy-per-conversion while using static-biasing for comparator precision; the EEG feature extraction processor employs low-power hardware whose parameters are determined through
BibTeX
@article{kumar2010micro,
author = {Naveen Kumar and Ali Shoeb and Jose L. Bohorquez and Joel Dawson and John V. Guttag and Anantha P. Chandrakasan},
title = {{A Micro-Power EEG Acquisition SoC With Integrated Feature Extraction Processor for a Chronic Seizure Detection System}},
journal = {IEEE J. Solid State Circuits},
year = {2010},
url = {https://openreview.net/forum?id=zqkPxp8xGp},
}Loading related content...