Abstract
Cryogen-free dilution refrigerators generally simplify low temperature research but some types of samples, including superconducting qubits and other nanoelectronic devices, are affected by environmental heat sources such as stray photons or residual helium. We present the design and performance of a hermetic cell installed on the mixing chamber plate of a cryogen-free dilution refrigerator. The performance was quantified by measuring the dependence of the resonance frequency of a mechanical oscillator installed inside the cell on the mixing chamber temperature down to 10 mK. We found the expected logarithmic temperature dependence of the resonance frequency down to the lowest temperatures, demonstrating that the efficiency of the hermetic shield is significantly better than that of a simpler shield with no visible gaps.
Introduction
Cryogen-free ("dry") dilution refrigerators have become the standard for ultra-low temperature research due to their ease of automation and independence from liquid helium supplies. However, these systems often lack the robust "Inner Vacuum Can" (IVC) characteristic of traditional "wet" refrigerators. This design difference can expose sensitive samples—such as superconducting qubits or nanomechanical resonators—to parasitic heating. Even when the refrigerator's mixing chamber plate reaches 10 mK, the sample itself may remain significantly warmer due to stray infrared photons from higher-temperature stages or residual helium gas carrying heat across the vacuum.
Design and Methodology
To ensure samples reach true thermal equilibrium, we developed a compact hermetic cell made of high-purity copper. This cell creates a localized, vacuum-tight environment directly on the mixing chamber plate.
Key features of the design include:
- Indium Seals: Indium wire gaskets provide a vacuum-tight seal between copper flanges at millikelvin temperatures.
- Mechanical Reinforcement: Stainless steel rings are employed to distribute bolt pressure evenly, preventing the copper flanges from warping.
- Pinch-Seal Technique: A copper capillary tube is used to evacuate the cell and is then permanently sealed using a manual vise-grip tool, eliminating the need for bulky permanent pumping lines.
Performance was validated using a Double Paddle Oscillator (DPO), a high-precision mechanical resonator. In amorphous materials, the resonance frequency of such a device follows a predictable logarithmic dependence on temperature. If the DPO's frequency follows this curve down to 10 mK, it confirms the sample is perfectly thermalized.
Results and Performance
The effectiveness of the hermetic cell was compared against several other shielding configurations, including an unshielded setup and a "well-sealed" shield closed with copper tape.
As shown in the data, only the hermetic cell maintained the expected logarithmic frequency shift down to the base temperature of 10 mK. Even the "well-sealed" shield allowed enough radiation leakage to cause significant heating. The study also ruled out mass loading from residual helium gas, as no corresponding frequency shifts were observed, identifying stray photons as the primary culprit for parasitic heating.
Notably, the hermetic cell in a dry fridge provided superior thermalization compared to a standard setup in a traditional wet fridge, which lacked a dedicated internal radiation shield.
Discussion and Conclusion
The results demonstrate that achieving true 10 mK sample temperatures in cryogen-free systems requires more than just "good enough" shielding. The vacuum-tight indium seal of the hermetic cell provides the necessary thermal decoupling from environmental heat sources.
Future developments will focus on addressing heat transfer through microwave wiring, which remains a challenge for quantum electronics. Applying this hermetic cell design to superconducting qubits could potentially lead to longer coherence times by mitigating the impact of non-equilibrium quasiparticles generated by stray radiation.