Mri Raw Data: Rf-Spoiled Radial Flash Sequence Without Cardiac Gating And During Free Breathing

Florian Knöll, Christian Clason, Kristian Bredies, Martin Uecker, Rudolf StollbergerZenodo 2011 · DOI · OpenAlex

Abstract

RF-spoiled radial FLASH sequence without cardiac gating and during free breathing<br> Sequence parameters:<br> TR 2.0ms<br> TE 1.3 ms<br> Flip angle 8°<br> Matrix Size 128x128 (two-fold oversampling, resulting in 256 sample points for each radial spoke)<br> Radial Spokes: 125<br> In Plane Resolution 2mmx2mm<br> Slice Thickness 8mm<br> 3T System<br> 32 channel body array coil, compressed to 12 virtual channels using svd based coil compression<br> ote: This data set was acquired with a radial acquisition. The corresponding k-Space trajectory is also included (256x125 matrix k). The Non-Uniform-FFT Toolbox is needed for reconstruction of this data set.<br> Florian Knoll ([email protected]) Date: 2.2.2011
BibTeX
@inproceedings{knoll2011mri,
  author = {Florian Knöll and Christian Clason and Kristian Bredies and Martin Uecker and Rudolf Stollberger},
  title = {{Mri Raw Data: Rf-Spoiled Radial Flash Sequence Without Cardiac Gating And During Free Breathing}},
  booktitle = {Zenodo},
  year = {2011},
  doi = {10.5281/zenodo.800511},
}