An MRI scanner makes its images by switching magnetic gradient coils on and off very fast. The coils sit inside a strong static field, so every switch shakes the scanner's structure. That shaking produces the loud knocking patients hear, and it induces voltages and field ripples that jitter the intended sampling path. OPTIKS showed that this can largely be avoided in software. It keeps the sampling path fixed and changes only how fast the scanner moves along it. It does this by gradient descent through a measured acoustic transfer function, a curve that says how strongly the hardware rings at each frequency. The result moves waveform energy away from the structure's resonances. The reported effects on clinical 3T scanners were large: up to 94.8% less back-electromotive force (the voltage the vibrating coil induces back on itself), 91.1% less field oscillation, up to 9.22 dB lower peak acoustic noise, and up to 11.4% faster readouts.
That transfer function, however, is measured once, typically on a cool machine, and is treated as fixed. Gradient coils heat substantially during demanding protocols. Heat changes the stiffness and damping of the coil assembly and its epoxy, so the resonances the optimizer carefully avoided may move toward the waveform's energy. This is a simulation-to-hardware gap in miniature: the optimizer's internal model of the hardware is accurate at calibration and goes stale in operation. We propose to measure how far the resonances actually drift during a sustained scanning session. We then fit a small temperature-dependent model of them and precompute a library of timing solutions indexed by coil temperature, which the scanner switches between as it heats. The proposal also includes an explicit test of whether a single solution made robust to the whole temperature range does just as well. If it does, that simpler solution should win.
Core Claim
We claim three things. First, the noise and back-EMF reductions OPTIKS predicts from a cold calibration will degrade measurably over a long, high-duty session. Second, the degradation can be explained by a few resonance peaks shifting in frequency and damping as a smooth function of coil temperature. Third, timing solutions chosen to match the current temperature recover most of the lost benefit. They should also keep more of the speed gain than one conservative solution designed for the worst case over all temperatures.
The new element is not differentiable timing optimization, which OPTIKS supplies. It is the decision to treat the hardware model as a function of a measured operating condition, and to settle that model's fidelity directly against hardware. This also removes OPTIKS's 5–15 minute optimization time from the critical path, because the library is computed before the session and the scanner only looks up an entry while scanning.
Why this gap, and what the neighbouring work says
The evidence that resonances move between model and operation comes from outside MRI. In machine tools, Chan et al. compared finite-element predictions of a rotary table's natural frequencies with hammer-impact measurements. They found 8.5%, 8.3% and 1.4% errors on the first three modes, and they validated only at no load. They then programmed the controller to skip speeds near those modes. That is the same "avoid the dead zone" strategy OPTIKS uses, and it carries the same blind spot: operating conditions shift the modes away from where the avoidance was computed. Chumakov et al. suppressed vibration-induced beam broadening in a cryogenic monochromator by fixing coolant pump frequencies in non-resonant bands. Their work again assumes a stationary resonance map. These papers motivate the proposal. They show that frequency avoidance works and that it depends on a resonance map that may not hold.
The method borrows its shape from the computing side of the simulation-to-silicon gap. Walker et al. fixed gem5's power estimates by adding a physically motivated, temperature-dependent leakage term fitted from on-chip sensors, reaching 2.8% mean absolute percentage error across unseen workloads. Their lesson is that a low-dimensional physical model with temperature as an explicit input beats both the idealized simulator and unconstrained regression. We apply the same move to the acoustic transfer function. Two other papers instead constrain the design by representing the obvious alternative, which is to inject measured variability into the optimization rather than model it explicitly. Rasch et al. trained networks under measured analog-hardware noise, and Air Learning trained drone policies under profiled onboard latency, cutting the flight-time gap from 37.73% to 0.5%. Their success is why the robust-over-temperature baseline below is a serious competitor and not a strawman.
Research Program
The work has three stages. Each one can end the project honestly.
Stage one: measure the drift. On a clinical 3T system with a phantom, we run a sustained high-duty heating protocol, for example repeated spiral and circular echo-planar readouts. We periodically interleave short probe waveforms: chirps or band-limited noise of a few seconds. At each probe we record the acoustic response with an MR-safe microphone, the back-EMF, and the coil temperature reported by the scanner's thermal monitoring. This yields transfer-function snapshots across the temperature range a real session reaches. If the dominant peaks move by less than the spectral width OPTIKS already leaves around them, the project stops here. That would be a useful negative result: static calibration suffices.
Stage two: model the drift. We fit each snapshot with a sum of resonant modes and let the mode parameters vary smoothly with temperature:
$$H(f;T)=\sum_{k=1}^{K}\frac{a_k(T)}{f_k(T)^2-f^2+i\,f\,f_k(T)/Q_k(T)}$$
Here $H$ is the transfer function at frequency $f$ and coil temperature $T$. $K$ is the number of dominant modes. $f_k$, $Q_k$ and $a_k$ are each mode's centre frequency, sharpness, and gain, and each is taken as a low-order polynomial in $T$. We test this model on held-out temperatures and on a held-out session against two alternatives: the cold-calibrated static curve, and nearest-snapshot lookup.
Stage three: build and deploy the library. For each temperature bin, the existing OPTIKS optimizer is run against $H(f;T)$, keeping its nerve-stimulation barrier. The scanner then selects a solution at sequence or repetition boundaries from its live temperature reading. Because the geometric path is unchanged and only the timing differs, reconstruction needs only the per-repetition timing. We will check that switching introduces no discontinuity artifacts.
Baselines
The naive reference is conventional time-optimal timing with no knowledge of the transfer function. We also include global gradient de-rating, which OPTIKS's 11.4% speed gain was measured against. These set the floor for noise and the ceiling for scan time.
The strongest existing method is OPTIKS itself, with a transfer function calibrated cold, before the session, run from its public code. Its reported numbers on these trajectories are the targets the cold solution should reproduce at the start of the session: up to 94.8% back-EMF reduction, 91.1% field-oscillation reduction, and 9.22 dB peak-noise reduction (OPTIKS). There is no reported result on Lacuna for OPTIKS performance under sustained coil heating. Establishing that number is the first deliverable.
The most threatening competitor is a robust OPTIKS variant. It is a single timing solution optimized against the worst case, or the average, over all transfer functions measured across the temperature range, in the spirit of Rasch et al. and Air Learning.
Fairness rules apply across all methods. Every method uses the same trajectories, nerve-stimulation limits, phantom, heating protocol and scanner. Every optimized method gets the same optimizer, iteration budget and probe data. Each is evaluated by measured, not predicted, acoustic noise and back-EMF at matched temperatures.
Two ablations isolate the key ingredients. In the first, the library is indexed by elapsed time rather than measured temperature. This tests whether temperature is the right input. In the second, the library is built from raw nearest-snapshot curves rather than the parametric model. This tests whether the model adds anything beyond the data.
The win condition has two parts. Across the hot half of the session, the library must hold measured peak noise and back-EMF closer to their cold-calibration values than static OPTIKS does. It must also keep a shorter readout than the robust variant. Doing better than static OPTIKS alone is not enough.
Why not just…?
Why not just re-measure the transfer function and re-run OPTIKS during the session? Re-measuring is cheap, but re-optimizing takes 5–15 minutes per sequence, so the solution would always lag the hardware's current state. Uploading new timings to a running scanner is also operationally awkward. The re-measure-and-reoptimize result does, however, serve as an upper bound in our analysis. If the library comes within noise of it, the library has captured everything re-optimization would.
Why not just optimize one solution that is robust across all temperatures? This is a real threat, and it may be the right answer. If the resonances move only slightly, a single robust solution costs almost no speed and the library is needless complexity. The comparison of readout duration at matched noise settles this directly. We commit to reporting the robust variant as the recommendation if the speed difference is within measurement noise.
Why not just randomize over hardware parameters, as Active Domain Randomization does? Active Domain Randomization learns a sampler over many rigid-body physics parameters because that space is too large to cover uniformly. Our parameter space is a single measured temperature with a handful of modes, so direct measurement replaces sampling. Randomization reduces to the robust baseline above, which is already tested.
Why not just cool the coils harder or de-rate? Cooling capacity is fixed by the installed hardware. De-rating is exactly the slow baseline OPTIKS improves on. Both are reported as reference points, not solutions.
Proposed Experiments
The decisive experiment is a sustained session on a phantom. It runs a heating protocol through the full temperature range and repeats variable-density spiral, rosette and circular echo-planar readouts under each method, interleaved and randomized in order so that heating affects all methods equally.
We measure four quantities. The first is peak and A-weighted sound pressure level as a function of coil temperature. The second is back-EMF amplitude. The third is readout duration. The fourth is the predicted-minus-measured noise gap, which is the direct measure of the simulation-to-hardware gap for each method's internal model.
A second check repeats the drift fit on a different day to test whether temperature alone predicts the transfer function. If other factors such as coil age or room conditions matter, a per-session recalibration step becomes necessary.
The final check reconstructs quantitative relaxation-time maps from a fingerprinting acquisition under each method. It confirms that timing switches leave map agreement with a reference scan intact, matching OPTIKS's claim of unchanged diagnostic fidelity.
Pre-registered outcomes determine the conclusion. If stage one shows negligible drift, static calibration is vindicated. If drift is large but the robust variant matches the library on speed, the robust variant is the practical recommendation. Only if both conditions fail does the temperature-indexed library earn its added complexity.
References
- OPTIKS: Optimized Gradient Properties Through Timing in K-Space
- Integrating numerical techniques and predictive diagnosis for precision enhancement in roller cam rotary table
- Performance of a cryogenic silicon monochromator under extreme heat load
- Accurate and stable empirical CPU power modelling for multi- and many-core systems
- Hardware-aware training for large-scale and diverse deep learning inference workloads using in-memory computing-based accelerators
- Air Learning: a deep reinforcement learning gym for autonomous aerial robot visual navigation
- Active Domain Randomization