Many screening programs do not ask "is there a lesion?" once. They ask "has this lesion grown?" at every round. Japanese hemodialysis patients were scanned at fixed intervals. Small kidney masses under active surveillance are reimaged every six to twelve months. Lung screening repeats low-dose CT yearly. In each case, the decision to intervene depends on an estimated growth rate crossing a threshold. That estimate silently assumes that the change between two scans is biological. It is not, if the patient was scanned on a different machine, with a different slice thickness or reconstruction filter. A model trained at one hospital learns that hospital's scanners. It then reads a protocol change as a volume change, and the result is a false growth alarm or a missed progression.
This note proposes using each patient's own anatomy as the measuring stick. Stable reference tissues appear in both scans: cortical bone, fat, blood in the aorta, air in the airway. If these "look different" between two exams, the difference is scanner-induced, because these tissues do not grow. We train a growth estimator at a single site under one constraint: its output must carry no statistical dependence on those reference-tissue shifts. The constraint is justified because, in a scheduled screening program, which scanner a patient lands on is unrelated to how fast their tumor grows. Any correlation between estimated growth and scanner-induced shift is therefore measurement error by definition, and it can be penalized away using only one institution's data.
Core Claim
A lesion-growth estimator trained on serial screening CT from a single site can stay calibrated when the scanner changes between rounds, without multi-center data. Two ingredients make this possible. The first is supervision from pairs whose true growth is zero. The second is a penalty that removes any dependence of predicted growth on within-patient reference-tissue shifts and acquisition metadata, which together we call "anchors."
The method is borrowed from anchor regression as extended to nonlinear models by Londschien et al.. In their ICU work, exogenous variables tied to the environment are used to penalize residuals that covary with them. That work operated on tabular streams across nine hospital databases. Here, the anchors are not hospital indicators. They are per-patient, per-exam-pair physical measurements, which is what lets the idea work when only one hospital is available for training.
Concretely, let $g(x_1, x_2)$ be the predicted log volume ratio of a lesion between exams $x_1$ and $x_2$. Let $\Delta a$ be the vector of anchor shifts between the two exams: changes in median attenuation of reference tissues, a reference-structure size ratio, and slice-thickness and filter indicators. Let $\mathcal{Z}$ be the set of pairs known to have zero true growth. We minimize
$$ \mathcal{L} = \mathcal{L}_{\text{seg}} + \lambda_0 \sum_{(x_1,x_2)\in\mathcal{Z}} g(x_1,x_2)^2 + \gamma \left\| \widehat{\mathrm{Cov}}\big(g(x_1,x_2), \Delta a\big) \right\|^2 , $$
where $\mathcal{L}_{\text{seg}}$ is a standard lesion segmentation loss, $\lambda_0$ weights the zero-growth term, and $\gamma$ weights the decorrelation penalty, estimated over all training pairs in a batch. The third term is the anchor-regression idea in covariance form. It does not ask the model to ignore scanner differences in the image. It asks the final growth number to be unpredictable from them.
Why This Is Not Already Done
The closest systems are the large opportunistic non-contrast CT screeners. PANDA trained on 3,208 patients from one center and still generalized across 10 external centers. It did this with an organ-localization cascade, a joint segmentation-classification head and hard-example mining. GRAPE instead relied on 16 centers for validation plus a preprocessing harmonization step. Both answer a single-timepoint detection question, and both are motivating evidence that single-site training can transfer for detection.
Neither ingests prior scans, and neither addresses the measurement problem that surveillance creates. Growth thresholds such as those used in the DISSRM registry depend on differences of two measurements. Scanner bias does not average out of such a difference; it is the difference.
NLSTseg supplies the cautionary evidence. A 2D U-Net on low-dose lung screening CT reached 0.95 training IoU but only 0.42 on held-out scans, and it confused vessels with nodules. That instability, applied to two timepoints, compounds into spurious growth. Wang et al. and McGough et al. document that scanner contrast and slice thickness shift CT-derived measurements. That is the error this proposal targets, but neither offers a correction.
Research Program
The primary data source is the National Lung Screening Trial low-dose CT archive. It has the properties this question needs: repeated scheduled screening rounds, many acquisition sites, and documented variation in scanner manufacturer, slice thickness and tube voltage. NLSTseg pixel-level lesion masks provide segmentation supervision for its 605 patients.
Zero-growth pairs come from three sources:
- Same-acquisition pairs. Where an exam was reconstructed more than once with different filters or slice thicknesses, the reconstructions show the same tumor at the same instant, so true growth is exactly zero.
- Simulated pairs. These are built by resampling a scan along the slice axis and adding dose-appropriate noise.
- Stable-nodule pairs. These come from nodules that remained benign through follow-up. They provide only approximately zero growth and are used only for evaluation.
Anchors are measured by an organ segmenter, kept separate from the lesion model:
- median attenuation of subcutaneous fat, tracheal air and cortical vertebral bone, plus blood in the aorta;
- the ratio of a vertebral body dimension across exams, which captures resolution and partial-volume effects;
- acquisition metadata from the image headers.
We deliberately avoid muscle and organ volumes as anchors, because weight loss and illness change them biologically.
The training regime enforces single-institution conditions. All training pairs come from one site, or in a second setting from one scanner manufacturer. Evaluation pairs come from other sites and, critically, include pairs where the scanner changed between rounds.
A kidney arm would test transfer to renal surveillance, the application that motivates the problem. It requires serial non-contrast abdominal CT from a partner institution. The file lists no public longitudinal renal imaging set; KiTS19 is contrast-enhanced, single-timepoint and single-center. KiTS19 is used only to pretrain the kidney segmenter.
The independence assumption needs testing before it can carry the method. In a fixed-schedule trial, scanner assignment should be unrelated to tumor biology. It fails if suspicious patients are preferentially reimaged on newer diagnostic scanners. We test it directly by checking whether anchor shifts predict baseline lesion size, smoking history or eventual diagnosis. If they do, we restrict the penalty to pairs where they do not, and we report the restriction.
Baselines
Naive reference. This is the standard surveillance practice: segment the lesion independently at each timepoint and subtract volumes. Segmentation uses a 3D nnU-Net pipeline with its usual resampling and intensity clipping, the recipe that won KiTS19 with tumor Dice 0.851 and composite Dice 0.912 Heller et al.. On low-dose lung screening, the nearest reported segmentation number is the 2D U-Net test IoU of 0.42 from NLSTseg. No reported result was found on Lacuna for false-growth rates or growth-measurement repeatability under scanner change on NLST. These baselines therefore have to be measured, not quoted.
Strongest existing competitor. This is the same nnU-Net trained with center-aware adversarial domain augmentation CADA. CADA is the best single-source generalization method in the file, reaching 83.57% on CIFAR-10-C against roughly 76% for prior augmentation methods. It is the right competitor because it targets exactly our training condition: one source domain, unseen targets. We add the slice-resampling and noise augmentations used to build our simulated pairs, so that the competitor sees the same artificial scanner variation. We also add PANDA-style hard-example mining on false growth calls from the training site.
All methods share the same backbone, training site, epochs and a matched hyperparameter search budget, with checkpoints selected on held-out pairs from the training site only. Beating CADA means a lower false-growth alarm rate on cross-scanner stable pairs at matched sensitivity to true growth, with non-overlapping bootstrap intervals.
Ablations. These separate the two ingredients: the zero-growth term alone, the anchor penalty alone, and the anchor penalty with metadata-only anchors (no tissue measurements). The last one tests whether the patient's own anatomy adds anything beyond header fields.
Why not just…?
Why not just normalize intensities to a reference tissue, for example by rescaling attenuation so the aorta matches across exams? This is the most serious threat, and it is run as a baseline. A linear intensity correction can fix contrast offsets. Volume bias from slice thickness, however, comes from partial-volume blurring at the lesion edge. It depends on lesion size and shape, so a global intensity rescale cannot remove it. The experiments settle this by stratifying false-growth rates by lesion volume. If reference normalization matches our method on lesions under 5 cm³, which make up 500 of the 715 NLSTseg lesions, then the learned penalty is unnecessary and we will say so.
Why not just train with heavy augmentation? CADA is that alternative, run at full strength. The concern is that augmentation makes segmentations look robust while leaving small systematic biases. Those biases cancel within an exam but not across a scanner switch. The cross-scanner versus same-scanner gap in false-growth rate measures this directly.
Why not just collect multi-center serial data, as PANDA and GRAPE did for validation? Those consortia required 6,239 and 18,160 external scans for single-timepoint tasks. Serial data with scanner switches is far scarcer. Londschien et al. also found that external data can be worth as few as about 100 local patients when sites differ strongly. We quantify the trade-off by adding increasing numbers of external-site pairs to the naive baseline. We report how many are needed to match our single-site model, expressed in their "equivalent patient" units.
Why not just let radiologists measure diameters, as surveillance protocols do? Manual diameters carry the same partial-volume bias plus reader variability. A subset of cross-scanner stable pairs will be read by radiologists to place the human error rate on the same scale.
Proposed Experiments
The decisive check is on same-acquisition reconstruction pairs from sites never seen in training, where true growth is exactly zero. We report the 95% limits of agreement of predicted log volume ratio and the fraction of pairs crossing a clinical growth threshold of 25% volume increase. The method succeeds if its limits of agreement on cross-filter or cross-thickness pairs approach those of same-protocol pairs, while the baselines' limits widen.
The second check is on scheduled screening rounds with a scanner switch. We measure the false-growth rate on stable benign nodules, and the area under the curve for separating nodules later diagnosed malignant from those that stayed benign using growth alone. Both are reported separately for scanner-switch and no-switch pairs.
The third check asks whether the penalty did what it claims. On held-out sites, the residual correlation between predicted growth and anchor shifts should be near zero for our model and clearly nonzero for the baselines. Sensitivity to true growth should not drop, since a model that predicts zero everywhere would also pass the first two checks.
If the kidney arm's partner data materializes, the same three checks are repeated on serial non-contrast renal CT. There they are applied at the 0.5 cm/year growth threshold used to trigger intervention in small renal mass surveillance.
References
- Domain Generalization and Adaptation in Intensive Care with Anchor Regression
- Large-scale pancreatic cancer detection via non-contrast CT and deep learning
- AI-based large-scale screening of gastric cancer from noncontrast CT imaging
- Five-year Analysis of a Multi-institutional Prospective Clinical Trial of Delayed Intervention and Surveillance for Small Renal Masses: The DISSRM Registry
- NLSTseg: A Pixel-level Lung Cancer Dataset Based on NLST LDCT Images
- Integrating No.3 lymph nodes and primary tumor radiomics to predict lymph node metastasis in T1-2 gastric cancer