Telling Real Aortic Wall Change from Segmentation Drift in Steroid-Exposed Patients

Long-term corticosteroid therapy can thin the middle layer of the artery wall while blood tests suggest the patient is doing well. Case and cohort evidence links prolonged or high cumulative steroid exposure to aneurysm rupture and decades-later aneurysm complications. The natural way to watch for this damage is repeated CT angiography (CTA, contrast-enhanced CT of the blood vessels) together with three-dimensional deformation mapping. Deformation mapping registers a follow-up scan onto a baseline scan and reports, for each patch of the wall, how much the surface has grown. It already finds growth that diameter measurements miss. But it starts from a segmentation, an outline of the vessel wall drawn by a person or a network, and that outline is least reliable in exactly the places steroid damage should appear: thin, low-contrast wall and the boundary between wall and clot. If the wall loses contrast between scans, the outline can move even when the anatomy has not, and the growth map then reports change that never happened.

This proposal treats inter-observer variability as a property of change measurements, not of single images. The central observation is simple. When the same rater outlines both scans, that rater's systematic habit, such as habitually drawing the wall a little outside the true edge, largely cancels in the difference. Session-to-session inconsistency does not cancel, and neither does a shift in bias caused by changing image contrast. We will build a rater model whose latent "style" is shared across the two time points and whose bias is allowed to depend on local wall contrast. We will propagate its samples through deformation mapping to produce a per-region minimum detectable change, and test whether steroid-exposed aortas show growth beyond that threshold. Existing rater models only describe disagreement on one image. Existing deformation pipelines use one fixed outline. Neither can say whether a flagged patch of growth in a thinning wall is real.

Core Claim

For a wall patch $i$, write the measured surface-area change as $\Delta_i = g_i + e_i^{(2)} - e_i^{(1)}$. Here $g_i$ is the true growth and $e_i^{(t)}$ is the segmentation-induced error at time $t$. The uncertainty that matters is then

$$\mathrm{Var}(\Delta_i \mid g_i) = \mathrm{Var}(e_i^{(1)}) + \mathrm{Var}(e_i^{(2)}) - 2\,\mathrm{Cov}(e_i^{(1)}, e_i^{(2)}).$$

The covariance term is large when the same rater segments both scans under similar contrast. It shrinks when raters differ between scans, when a model trained on pooled raters is used, or when wall contrast changes between scans. The last case is the one iatrogenic thinning produces. Models that describe disagreement on a single image estimate only the first two terms, so they overstate uncertainty for consistent same-rater workflows. Pipelines that use one deterministic outline estimate none of the terms. Our claim has three parts. First, modeling the covariance explicitly, with rater bias that depends on contrast, gives correctly calibrated per-patch intervals for growth. Second, a model without it is badly miscalibrated in thrombus-laden and thin-wall regions. Third, a contrast-only change between scans produces spurious growth in deterministic pipelines that our model correctly flags as undetectable.

Grounding and Prior Work

The motivation comes from the vascular surveillance literature. Vascular Deformation Mapping for CT Surveillance of Thoracic Aortic Aneurysm Growth maps 3D wall growth by B-spline registration of serial CTA. It found growth in 24% of intervals, and 36% of those lay outside the maximally dilated segment. Its two-rater segmentations agreed at a Dice score (voxel overlap) of 0.97 ± 0.02, with area-ratio limits of agreement of −0.03 to 0.03. The translation to abdominal aneurysms, Vascular Deformation Mapping of Abdominal Aortic Aneurysm, failed in 45.5% of analyses, mainly because of wall and intraluminal-thrombus segmentation. These two results pull against each other, and we address that directly. The thoracic study suggests inter-observer effects are negligible. The abdominal study suggests they dominate. Our hypothesis is that global agreement numbers average away local disagreement concentrated in low-contrast patches. If our per-patch measurements show disagreement within ±0.03 even at thrombus and thin wall, the premise is wrong and we will report that. Towards personalised assessment of abdominal aortic aneurysm structural integrity shows that weak, highly compliant "islands" rarely sit at the maximum diameter. That supports local rather than diameter-based surveillance. It too computes strain on a single fixed mesh.

The clinical regime is supplied by the iatrogenic literature. Tang et al. describe high-dose immunosuppression thinning the media of existing aneurysms while inflammatory markers settle. Shimooka et al. report aneurysm-related stroke after two decades of steroids. Skolnik et al. and Hew et al. place harm-associated cumulative exposure at 500–1000 mg prednisolone-equivalent, and report that roughly a quarter of Australian inhaled-steroid asthma users reach such doses. These papers motivate the target population and give a defensible exposure stratification. None of them contains imaging measurements we can reuse.

The method draws on multi-rater segmentation models. Pionono gives each rater a Gaussian in latent space. Its mean captures systematic bias and its variance captures intra-rater inconsistency, which is exactly the decomposition the covariance term needs. It reports Cohen's kappa of 0.758 on Gleason 2019 against 0.750 for the strongest baseline, but only on 2D histology and single images. The Latent Doctor Model infers a rater embedding from a support set and conditions disagreement on the image. We reuse its image-conditioned prior to let bias depend on wall contrast. Variational Inference for Quantifying Inter-observer Variability supplies the simpler shared-latent variant used as an ablation. QUBIQ constrains our design: plain ensembles often match generative models at reproducing consensus probability maps, so any added structure must earn its keep against an ensemble. Rethinking Intracranial Aneurysm Vessel Segmentation shows that high-Dice outlines can be topologically unusable. We therefore add its centerline-Dice loss (overlap of vessel centerlines) so that sampled outlines remain registrable surfaces. DiFF argues that the target of segmentation should be set by the downstream task. We adopt that view by calibrating against change measurements rather than per-image overlap, without using its 2D classification machinery.

Research Program

The first component is an annotation study, because no dataset in the file has multi-rater outlines of the same aorta at two time points. From an institutional longitudinal surveillance CTA archive, we select scan pairs from patients with and without documented chronic systemic steroid exposure. Cumulative dose is computed from pharmacy records and split at 500 and 1000 mg prednisolone-equivalent. Four to five raters each outline the lumen, outer wall and thrombus on both scans of every pair. Each rater repeats a subset weeks later, which measures intra-rater variance directly. The design deliberately includes same-rater pairs, cross-rater pairs and repeat sessions, so every term in the variance decomposition is observed empirically.

The second component is the model. A 3D segmentation backbone in the style of the SEG.A. ensemble is extended with a Pionono-style latent rater variable $z_r$. The variable is sampled once per rater and shared across both time points, which induces the covariance term. Its mean is shifted by a small network that takes the local wall-to-background contrast, so that contrast loss between scans produces bias drift rather than pure noise. A centerline-Dice term keeps samples topologically valid. Training uses single-time-point multi-rater data (QUBIQ CT tasks, plus our annotations) and the paired annotations for the coupling.

The third component is propagation. Each joint sample of baseline and follow-up outlines passes through the same deformation-mapping registration. This yields a distribution of per-patch area ratios and a minimum detectable change, defined as the 95th percentile of $|\Delta_i|$ under the model when $g_i = 0$. Growth is declared only where the observed change exceeds that threshold.

Baselines

The naive reference is the clinical one: maximum axial diameter change, with its reported 1–5 mm inter-observer error as cited in the thoracic deformation-mapping paper. We run it on the same pairs to show which flagged regions it can and cannot see.

The strongest existing segmentation-to-growth pipeline is deformation mapping driven by a single high-performing automatic outline. We implement the outline with the Auto3DSeg SegResNet ensemble from Aorta Segmentation from 3D CT in MICCAI SEG.A. 2023 Challenge, which reports a hidden-test Dice of 0.920 and a 95th-percentile Hausdorff distance (near-worst boundary distance) of 6.013 mm. This is the strongest competitor because it is what a deployable pipeline would use today. Its ensemble members also give a free disagreement estimate. It is retrained on the same training data and given the same compute and hyperparameter budget as our model. Its growth intervals come from member spread.

The reference that defines success is "real raters through the pipeline": deformation mapping run on every human outline pair in a held-out patient set. We beat the single-outline baseline if two things hold. First, our intervals achieve near-nominal coverage of held-out human-pair area ratios in thrombus and thin-wall patches where the ensemble's coverage falls clearly below nominal. Second, on contrast-only phantoms we keep false growth calls near zero while the baseline's rate is clearly above zero.

Two ablations isolate the key ingredient. The first draws independent latent samples at the two time points. This equals a Pionono or shared-VAE model applied per image and removes the covariance term. The second keeps the shared rater variable but removes contrast dependence. If neither ablation loses calibration relative to the full model, the temporal coupling is not needed and we will say so.

Why not just…?

Why not just have several radiologists segment each scan and run the pipeline on every pair? That is the gold standard, and we use it to evaluate. It does not scale to routine surveillance. At the 3–20 seconds per frame reported for interactive refinement in Automatic Explainable Segmentation of Abdominal Aortic Aneurysm, multiple raters over hundreds of slices per scan is impractical. The model's job is to reproduce this reference from one outline. Coverage against held-out human pairs measures whether it does.

Why not just use deep-ensemble disagreement? QUBIQ shows ensembles are strong at matching single-image consensus maps, so this is the most serious threat. Ensemble spread reflects model uncertainty and has no notion of a rater habit that persists across time points. It also does not represent a bias that shifts systematically when contrast drops. The ensemble baseline and the contrast-only phantom experiment decide this directly. If ensemble spread matches our coverage in low-contrast patches and also flags phantom changes, we will drop the latent rater model in favor of the ensemble.

Why not just add random boundary jitter, such as dilating or eroding the outline by a millimetre? Independent jitter assumes errors are spatially uniform and uncorrelated across scans. That overstates uncertainty when the same rater outlines both scans and understates it where contrast changes. We include tuned jitter, with magnitude fit to the multi-rater annotations, as a further baseline. It is beaten only if its per-patch coverage is wrong in opposite directions for stable and low-contrast patches.

Why not trust the thoracic study's ±0.03 agreement and skip all of this? Those limits are aggregate figures from relatively well-defined thoracic boundaries. The 45.5% abdominal failure rate points the other way. Our first measurement is the per-patch disagreement map in the annotation study. If it stays within that range everywhere, the project shrinks to a confirmatory note.

Proposed Experiments

The first decisive check is a contrast-only phantom. On baseline scans, we lower wall and thrombus attenuation toward the surrounding tissue, mimicking the contrast loss expected from media thinning, while leaving geometry unchanged. We then run every pipeline against the original. True growth is zero by construction. We compare spurious growth area and false-call rate across pipelines, as a function of the size of the contrast drop.

The second check is calibration on held-out patients. For each pipeline we measure the empirical coverage of 90% intervals against human-pair area ratios. Results are stratified by patch type (clear wall, thin wall, thrombus interface) and by pairing type (same rater versus cross rater).

The third check is a zero-change control. We use public 4D-CTA cardiac phases from the Zenodo cohort used for the structural integrity index, registering end-diastolic frames from the same session. The fraction of the surface flagged as growth should be near zero.

The fourth analysis is exploratory and clinical. In the steroid-stratified cohort, we compare the fraction of wall area showing growth beyond the minimum detectable change across exposure strata, adjusting for baseline diameter, interval length and inflammatory aortitis. Underlying vasculitis is a confounder by indication that observational data cannot fully remove. We do not expect a reported result here, since no reported result was found on Lacuna for imaging-measured wall change by cumulative steroid dose. The value of the study is that whatever difference appears will be stated with a defensible uncertainty, rather than being possibly an artifact of fading contrast.