When radiologists read a cancer scan, they routinely mark each tracked tumor with two lines: its longest diameter on one axial slice and the longest diameter perpendicular to it. This convention is called RECIST (Response Evaluation Criteria in Solid Tumors). The marks are stored in every hospital's image archive as a byproduct of normal care. Prior work has shown that these marks can train a 3D lesion segmenter without anyone drawing a full outline. Such segmenters still break when moved to a new hospital with different scanners, contrast timing, and slice thickness. The usual fix is to collect outlines at the new site, which is exactly the expert labor the weak labels were meant to avoid.
Our observation is that the new hospital has the same kind of free marks. We propose to adapt a segmenter trained only on RECIST marks, source-free, to an external hospital, using that hospital's own archived marks as the only supervision. "Source-free" means only the trained weights leave the original institution, not its images. The model is tuned so that its predicted 3D masks agree with the recorded diameters. The same agreement score, computed on held-out marks, also serves as a way to choose and audit the adapted model at a site that has no ground-truth outlines. No dense mask is required anywhere, at either site, for training or for model selection.
Core Claim
Recorded diameters constrain a lesion's extent, orientation, and position on one slice. They do so tightly enough to act as a differentiable adaptation loss and as a label-free validation metric at a new site. A weak-only source model adapted this way should close most of its cross-hospital gap. The comparison points are a model trained from scratch on the target's marks and a model fine-tuned on masks derived from those marks. The expected advantage should be largest when the target site has few measured lesions, and on slices far from the measured one.
Where This Sits
Three strands of prior work set up the problem.
Weak-label segmenters that are never tested elsewhere. Accurate Weakly Supervised Deep Lesion Segmentation on CT Scans: Self-Paced 3D Mask Generation from RECIST expands a 2D RECIST slice into 3D masks through slice-by-slice self-paced training. It reports 76% 3D Dice on DeepLesion and 70.2% on a lymph node set, and it notes that errors grow away from the seed slice. Weakly Supervised Lesion Co-Segmentation on Ct Scans uses RECIST to seed GrabCut and trains a paired co-segmentation network, reaching 90.3% Dice on 2D slices. Both train and test within one institution's data. We reuse their mask-generation recipe to build our source model, and then ask what happens after deployment elsewhere.
Adaptation methods that still need dense source labels or new target labels. Scribble-based Domain Adaptation via Co-segmentation transfers T1-c to T2 MRI using 150 fully outlined source scans plus newly drawn target scribbles, reaching 83.4% Dice. Weakly-Supervised Cross-Domain Adaptation for Endoscopic Lesions Segmentation uses full source masks and target image-level tags. Both methods motivate using weak target labels. They differ from our setting in two ways. They require dense source outlines, and their target labels must be commissioned, whereas ours already sit in the archive.
Source-free adaptation of weak localizers and its failure modes. Source-Free Domain Adaptation of Weakly-Supervised Object Localization Models for Histology found two problems in histology. First, adaptation that improves classification degrades localization. Second, self-training collapses when starting accuracy is low. Adaptation of Weakly Supervised Localization in Histopathology by Debiasing Predictions counters the collapse by unlearning overconfident predictions. It remains dependent on its own class activation maps and on a sensitive forgetting ratio.
These papers constrain our design. They warn that self-training on the model's own beliefs is unsafe under shift, and that model selection without target localization labels is itself an unsolved problem. Our design answers both warnings with an external geometric anchor, the recorded measurement, instead of the model's own predictions.
Improved post-hoc probability calibration for out-of-domain MRI segmentation documents severe overconfidence under protocol shift, with expected calibration error above 15% for bSSFP to LGE. This is why we cannot use model confidence for checkpoint selection. AMOS reports drops of more than 20% on unseen scanners. That supports treating scanner change as the central threat.
Research Program
Source model. We first train a 3D lesion segmenter on DeepLesion using only its real RECIST marks, following the self-paced procedure of Cai et al. This gives a weak-only model that reproduces a published result. Only its weights are shipped to the target site.
Measurement-consistency loss. At the target site, each archived measurement provides three things: a slice index, long-axis endpoints $a, b$, and short-axis endpoints $c, d$. Let $p(x)$ be the model's soft foreground probability at pixel $x$ on the measured slice. Let $u$ be the unit vector along the long axis.
We define a soft extent along $u$ as the gap between probability-weighted soft maximum and soft minimum projections:
$$E_u(p) = \tau \log \sum_x p(x)\, e^{\langle x, u\rangle/\tau} \; - \; \big(-\tau \log \sum_x p(x)\, e^{-\langle x, u\rangle/\tau}\big),$$
Here $\tau$ is a temperature controlling how sharply the soft max approximates a hard max. $E_v$ is defined the same way for the short-axis direction $v$.
The loss has three parts:
$$\mathcal{L} = \big|E_u(p) - \|a-b\|\big|_\epsilon + \big|E_v(p) - \|c-d\|\big|_\epsilon + \lambda\,\mathcal{L}_{\text{seg}},$$
In this loss, $|\cdot|_\epsilon$ is an absolute error that ignores deviations smaller than a tolerance $\epsilon$. We set $\epsilon$ from published reader-to-reader variability in RECIST measurements, so the loss does not chase measurement noise. The term $\mathcal{L}_{\text{seg}}$ is a cheap point-and-line term: pixels on both measured segments should be foreground, and pixels a few millimetres beyond each endpoint along the lines should be background.
On slices away from the measured one, we add a single constraint: predicted lesion extent must shrink smoothly with distance from the measured slice. This targets the slice-direction error growth that Cai et al. report. The loss never compares the model against its own pseudo-masks. That is the deliberate difference from the self-training schemes that collapsed in the histology work.
Label-free selection and auditing. We hold out 20% of the target site's measured lesions. On those lesions we compute the same diameter and endpoint agreement from the adapted model's predictions. This score is used for checkpoint selection and for choosing $\tau$, $\epsilon$, and $\lambda$. Per-lesion disagreement also flags cases that a clinician should review. The central empirical question is whether this score ranks checkpoints in the same order as true Dice does.
Baselines
Do nothing. The naive reference is the unadapted DeepLesion source model applied directly to each target.
Upper bound. A fully supervised 3D U-Net trained on the target's dense masks gives the ceiling.
Strongest competitor: from-scratch weak training. The self-paced RECIST pipeline of Cai et al. is retrained from scratch at the target site on the target's own marks. It uses the same free signal and needs no source model, so it is the real threat. Cai et al. report 76% mean 3D Dice on DeepLesion and 70.2% on the lymph node set, against 71.0% for full supervision there. I found no reported result on Lacuna for RECIST-only segmentation on LiTS or KiTS tumors. We will therefore rerun this pipeline ourselves under the same compute and tuning budget, with its hyperparameters chosen on the same held-out marks.
Ablation isolating the key ingredient. The source model is fine-tuned on GrabCut masks generated from the target marks, as in the co-segmentation initialization of Agarwal et al. This uses identical information but converts measurements into hard pseudo-masks instead of a tolerance-aware geometric loss.
Label-free source-free adaptation. Batch-normalization statistic updating and entropy-based self-training, in the style of SHOT from the histology SFDA benchmark, are run without the marks. This measures what the marks add.
Prompted foundation model. A training-free option uses the marks as box and point prompts to a frozen general vision backbone, in the manner of GazeRefine.
All methods receive the same target images, the same marks, and the same selection protocol on held-out marks. Beating the strongest competitor means higher mean 3D Dice on both targets, with non-overlapping bootstrap intervals at small and medium target sizes. On slices more than 10 mm from the measured slice, the gain must hold rather than shrink.
Why not just…?
Why not just train the RECIST pipeline from scratch at the new hospital, since it has marks anyway? This is the most serious alternative, and it may win once a site has thousands of measured lesions. Our bet is that a typical external site has tens to hundreds of measured lesions of a given type. In that regime, a model initialized on DeepLesion's roughly 32,000 marked slices and corrected by geometry should dominate. The experiment settles this directly: we sweep the number of target marks from 25 to all available and plot both methods. If the curves cross early, the honest conclusion is that adaptation matters only for small sites, and we will report the crossover.
Why not just turn the target marks into GrabCut masks and fine-tune? This uses the same information with less machinery. Agarwal et al. found that classical graph-based refinement traded recall for precision, which suggests hard pseudo-masks inherit systematic boundary errors. Our loss uses only the geometric facts the radiologist actually recorded, within a noise tolerance. The ablation compares the two head-to-head. If they tie, the contribution reduces to the selection score, and we will say so.
Why not just run standard source-free adaptation without any labels? In histology, pseudo-label methods entered error-accumulation loops from low starting accuracy (benchmark). They collapsed to chance on hard CAMELYON17 centers until explicit debiasing recovered 86.2% (SFDA-DeP). These methods also offer no trustworthy stopping rule under shift. We include them to measure that failure in 3D CT rather than assume it.
Why not just prompt a frozen foundation model with the marks? GazeRefine's prompted frozen features fall from 88.10% Dice on colour endoscopy to 80.28% on low-contrast prostate MRI. Low-contrast liver and kidney tumors on CT are the harder case. Such a model also prompts only the measured slice, so it offers nothing for off-slice volume. It remains a useful baseline for the measured slice.
Proposed Experiments
Datasets. The source is DeepLesion with its real RECIST marks. The targets are LiTS liver tumors and KiTS kidney tumors, both multi-center abdominal CT collections that release tumor masks. The lymph node CT set serves as a near-domain sanity check against Cai et al.'s published 70.2%.
Simulated marks. Target marks are synthesized from dense masks on the slice of largest tumor area, following the RECIST convention. Because synthesized marks are cleaner than real ones, we perturb endpoints and slice choice at levels matching published inter-reader variability. We then report results as a function of that noise.
Metrics. We report 3D Dice, Dice on the measured slice, Dice binned by distance from the measured slice, long-axis diameter error, and the correlation between predicted and true volume, where Cai et al. report $R^2 = 0.78$ on DeepLesion.
Decisive checks. Three checks determine whether the proposal holds up.
The first is the target-size sweep against from-scratch training, described above.
The second is the pseudo-mask ablation.
The third is a selection test. Across all checkpoints and hyperparameter settings, we compute the Spearman correlation between the held-out measurement-agreement score and true 3D Dice. We then compare the Dice achieved by selecting with that score against oracle selection by true Dice and against selection by prediction entropy. The selection claim stands only if agreement-based selection lands within a few Dice points of the oracle while entropy-based selection does not. If the correlation is weak, that claim fails, and we will say the loss helps but the audit does not.
A final stress test repeats the pipeline with target marks at twice the realistic noise. This tells us whether the tolerance $\epsilon$ is doing real work.
References
- Accurate Weakly Supervised Deep Lesion Segmentation on CT Scans: Self-Paced 3D Mask Generation from RECIST
- Weakly Supervised Lesion Co-Segmentation on Ct Scans
- Scribble-Based Domain Adaptation via Co-segmentation
- Weakly-Supervised Cross-Domain Adaptation for Endoscopic Lesions Segmentation
- Source-Free Domain Adaptation of Weakly-Supervised Object Localization Models for Histology
- Adaptation of Weakly Supervised Localization in Histopathology by Debiasing Predictions
- Improved Post-hoc Probability Calibration for Out-of-Domain MRI Segmentation
- AMOS: A Large-Scale Abdominal Multi-Organ Benchmark for Versatile Medical Image Segmentation
- GazeRefine: Expert Gaze as a Test-Time Prompt for Training-Free Medical Image Segmentation