Invisible Drift: Using Perceptual Masking to Predict Which Radiomic Features Scanners Will Break

Radiomics turns a medical scan into numbers. It computes hundreds of hand-defined features of a tumor, such as how bright it is, how round it is, and how its gray levels are arranged in space, and then feeds those numbers to a predictive model. The texture features are the fragile part. Change the scanner, the radiation dose, or the reconstruction filter and they swing wildly, even when the images look the same to a radiologist. Adversarial machine learning has a well-developed account of exactly this situation. Human vision ignores noise placed in busy, textured regions (an effect called contrast masking), and attack papers use maps of local variance or local entropy to hide large perturbations there. This proposal turns that account into a measurement tool. For each lesion, we search for the perturbation a radiologist would not notice that moves each radiomic feature the most. We claim that this "masked sensitivity," computed from a single scan, predicts which features will fail to reproduce across real scanners and doses. Because it needs only one scan, it can stand in for the expensive multi-scanner or test-retest side studies that current stability-aware methods require.

Put simply, perceptual masking describes the set of changes that count as "the same image" to a human. Features that move a lot inside that set are precisely the ones scanner differences corrupt, because scanner differences also live in that set: reconstruction kernels and dose alter high-frequency texture while leaving anatomy and mean density intact. If this holds, a cheap per-lesion audit can replace phantom campaigns as the source of stability information. That audit can then feed a stability-aware training scheme instead of a hard rule that deletes features.

Core Claim

We make one predictive claim and one engineering claim. The predictive claim is that per-feature masked sensitivity, computed on one reference acquisition, rank-correlates with the per-feature cross-scanner reproducibility measured on a multi-scanner phantom. It should also correlate better than sensitivity under an unmasked perturbation of equal size. The engineering claim is that masked sensitivity can serve as the noise prior in a stability-aware training scheme and recover most of the benefit of priors measured from real repeat scans. No paper in the research file connects the two directions. The nearest work treats human-perceptual quality and radiomic fidelity as two separate, opposed scorecards and does not use one to predict the other.

Where This Sits in Prior Work

The masking half supplies the method. Sparse and Imperceivable Adversarial Attacks restricts pixel changes to regions of high local standard deviation and forbids them along edges; it also uses a gradient-free greedy search, which we need because radiomic features are not differentiable. Towards Imperceptible and Robust Adversarial Example Attacks Against Neural Networks weights perturbations by variance over a 3×3 neighborhood. Adversarial attacks hidden in plain sight builds a local-entropy strength map, and its user study (n=35, Cohen's d ≈ 2.29 against a standard iterative attack) shows that such maps really do reduce human detection. Sparse Adversarial Attack via Perturbation Factorization folds variance into an optimization objective. All four work on natural 2D images, target neural classifiers, and treat hiding noise as the goal. We keep their masks but point the search at a different target: deterministic feature extractors defined by the Image Biomarker Standardization Initiative (IBSI), the consensus reference for radiomic feature definitions. We also use the search as a diagnostic rather than an attack.

The radiomics half supplies the motivation and the ground truth. A comparative analysis of image harmonization techniques reports that unharmonized intensity features keep a concordance correlation coefficient (CCC, agreement between repeated measurements) of about 0.972 across dose and kernel changes, while texture features fall to about 0.500. It also reports that networks tuned for visual similarity do not restore feature agreement. That is the visual-versus-quantitative gap our proposal explains mechanistically. Impact of CT dose on AI performance supplies per-feature intraclass correlation coefficients (ICC, a reproducibility score) across 13 scanners and 5 doses on a liver phantom, with PyRadiomics averaging 0.8355 ± 0.1705. We use these as the target our predictor must match.

Two papers constrain the design. Optimal Statistical Incorporation of Independent Feature Stability Information shows two things: deleting unstable features costs over 5% on some targets because the most predictive features are often the least stable, and ignoring stability inflates estimated performance by nearly 20%. Its method, which the file calls DAFIT, injects measured noise into feature vectors during training. It needs a side study to measure that noise. We aim to replace the side study rather than the method. Assessing radiomics feature stability with simulated CT acquisitions already predicts stability from a physics simulator, so it is our most direct competitor. It models Poisson dose noise under standard analytic and iterative reconstructions, and it does not model vendor-specific kernel differences. Lung CT harmonization of paired reconstruction kernel images shows kernel effects are mostly high-frequency. It separates them with a uniform low-pass identity loss, which supports our choice to constrain perturbations to preserve low-frequency density.

Research Program

For a lesion image $x$ with region of interest $R$, we build a masking map $m(v)$ at each voxel $v$ inside $R$. We compute $m$ in two variants, either the local standard deviation or the local Shannon entropy over a small 3D neighborhood. Following Croce and Hein, we set $m$ to zero on strong edges and on a thin shell around the segmentation boundary. The boundary exclusion keeps the audit from conflating acquisition drift with segmentation drift. We also require perturbations to have near-zero low-frequency content, which we implement by projecting out the Gaussian-blurred component, so that mean Hounsfield-unit density is unchanged. The masked sensitivity of feature $k$ is then

$$S_k(x)=\max_{\delta:\ |\delta(v)|\le \epsilon\, m(v),\ G*\delta=0} \frac{|f_k(x+\delta)-f_k(x)|}{s_k},$$

where $f_k$ is the IBSI-compliant feature extractor, $\delta$ is the perturbation, $\epsilon$ is a global budget, $G*\delta$ is the blurred perturbation, and $s_k$ is the spread of feature $k$ across the cohort so that sensitivities are comparable across features. Texture matrices bin gray levels into discrete intervals, so gradients do not exist. We therefore maximize with a randomized greedy search in the style of CornerSearch, and cross-check it with a soft-binning differentiable surrogate. Volumetric search cost is a known limitation of these attacks, so we cap evaluations per lesion and report how the estimate converges as the cap grows.

The budget $\epsilon$ is fixed by a detection experiment rather than by hand. Radiologists perform a two-alternative forced-choice task (shown two images, pick the perturbed one) on original versus perturbed CT slices under standard window settings. We choose the largest $\epsilon$ at which detection stays near chance. This anchors the envelope to clinical indistinguishability, as Göpfert et al. did for natural images. We will report how sensitive the results are to $\epsilon$. If only a small reader panel is available, we will also report results at several budgets defined by a perceptual proxy.

The second stage feeds $S_k$ in as the per-feature noise scale of DAFIT-style augmentation, in place of the variance DAFIT measures from test-retest scans. We then train radiomic classifiers with this augmentation.

Baselines

The naive reference point is a radiomic model trained with no stability information. For the prediction task, the naive predictor is feature-type membership alone ("texture unstable, intensity stable"), which already reflects the 0.972 versus 0.500 CCC split. Beating it requires a significant within-texture-family rank correlation with measured ICC, and this is the test that matters most. It is the strictest test because the family split alone could otherwise produce a high overall correlation.

The strongest competitor for predicting stability is the ASTRA-based physics simulator of Flouris et al. We run it on the same single reference acquisition, give it the same compute per lesion, and tune its dose and reconstruction settings on the same held-out scanners that we use to tune $\epsilon$. The file reports no global ICC figure for it (no reported result found on Lacuna for simulator-predicted ICC correlation). We therefore reimplement it and compare Spearman correlation against measured per-feature ICC. Beating it means higher correlation on the cross-scanner and cross-kernel components of variation. Merely matching it on dose-only variation would count as a tie, not a win.

The strongest competitor for downstream training is DAFIT supplied with noise measured from actual repeat scans. This is an upper reference, because it uses the very side study we aim to replace. We will count matching it to within its own seed variance as success. We also include hard ICC-threshold filtering, for which DAFIT reports a drop of over 5% on targets such as sphericity. For deep-feature context we report CT-FM, whose reported mean ICC is 0.9347 ± 0.0420 in the dose study. The key ablation replaces the masked set with an unmasked perturbation of identical total L2 norm, and then with uniform Gaussian image noise. If masking carries no information, these will predict ICC as well as masked sensitivity does. Secondary ablations compare the σ-map and entropy variants and remove the edge and boundary exclusions.

Why not just…?

Why not just run the physics simulator? This is the real threat. Flouris et al. show that simulated acquisitions reproduce phantom stability rankings. Their simulator models dose noise under generic reconstructions, not the proprietary kernel behavior that dominates cross-vendor drift, which is where Lung CT harmonization of paired reconstruction kernel images still finds residual effects. Our decomposition experiment below settles it. We split measured ICC into its dose component and its scanner/kernel component. If the simulator wins on both, the masking audit adds nothing for CT, and we will say so. It would then remain useful only for MRI or PET, where no comparable simulator is standard.

Why not just measure stability on a phantom once and reuse it? Phantom ICCs describe phantom tissue at the scanners that took part in the study. Features that DAFIT finds most predictive are lesion-specific: sphericity and heterogeneity depend on the patient's lesion. We test transfer directly. We ask whether phantom-derived ICC or per-lesion masked sensitivity better predicts kernel drift on patient lesions in the National Lung Screening Trial (NLST) paired-kernel data.

Why not just use a foundation model that is already stable? CT-FM's higher ICC is reported, but the same paper finds it also classifies dose best. That means acquisition information survives inside its features, and high ICC does not imply insensitivity. Explicit IBSI features also remain the standard for interpretable biomarkers. We report CT-FM as context rather than as a replacement.

Why not just add unstructured image noise as augmentation? Uniform noise spends its budget evenly, including in smooth tissue where real kernel changes do little. Our equal-norm unmasked ablation tests exactly this. If uniform noise predicts ICC as well as masked noise, the masking hypothesis is falsified.

Proposed Experiments

The primary test uses the CT4Harmonization-Multicentric phantom dataset of Impact of CT dose on AI performance: 1,378 series from 13 scanners at 5 doses, with liver regions for normal tissue, cysts, hemangioma, and metastasis. We compute masked sensitivity for every IBSI feature using only the 14 mGy series of one scanner. We then correlate it with the ICC measured across the remaining scanners and doses, holding out scanners for tuning. We report correlation within the texture families that analyze gray-level co-occurrence, run lengths, and zone sizes, and we report dose-driven and scanner-driven variance separately.

The second test uses NLST paired kernels, including GE BONE versus STANDARD, the pair for which Lung CT harmonization of paired reconstruction kernel images reports percent-emphysema RMSE falling from 9.91 to 0.84 after harmonization. Here we test whether per-lesion masked sensitivity on the STANDARD image predicts each feature's actual shift under BONE reconstruction.

The third test is downstream. We train liver-tissue classifiers on radiomic features with leave-scanner-out evaluation on the phantom, and nodule-attribute models on LIDC-IDRI following the DAFIT protocol. We compare naive training, hard filtering, measured-noise DAFIT, simulator-noise DAFIT, and masked-sensitivity DAFIT, using identical classifiers and tuning budgets. The metrics are held-out accuracy and AUC, plus the gap between internal and external performance, which DAFIT identifies as the roughly 20% overestimate to close.

The decisive result is a positive within-texture rank correlation that exceeds both the equal-norm unmasked ablation and the physics simulator on the scanner/kernel component of variation. That result would show masking identifies which features scanners break, not merely that texture is fragile. A null result on the unmasked ablation would falsify the central claim cleanly, and we will report it as such.