Does Rater Disagreement Change Predicted Fracture-Fixation Fatigue?

Engineers who simulate how a fixed hip fracture or a spinal cage will hold up start from a 3D model of the bone. A person, or a program, traced that model by hand from a CT scan, marking where bone ends and soft tissue begins. Experts disagree about where that edge lies, especially on thin cortical walls and around hairline fracture lines. Yet the simulation studies that compare surgical implants treat one tracing of one patient as exact. Those studies then rank the implants by whether peak stress crosses a bone-damage threshold, often by margins of a few tens of megapascals.

This proposal asks a concrete question. When we replace that single tracing with the real spread of expert tracings, and replace the single static "worst-case" load with a repeated walking load whose damage adds up over many cycles, do the implant rankings and pass/fail verdicts still hold? We expect two answers. Rater disagreement should matter more for fatigue predictions than for static peak stress, because fatigue damage grows steeply with stress. And the disagreement that matters is local and structured, so neither a uniform outward/inward shift of the bone surface nor random jitter of mesh nodes will reproduce it. Both expectations are testable, and either could fail.

Core Claim

Inter-observer segmentation variability is a first-order source of uncertainty in fatigue-based implant comparisons. It is concentrated at thin, low-contrast bone regions, and it is amplified when the analysis moves from static peak stress to cyclic damage. The mechanism is simple. Under a power-law stress-life relation, a cycle at stress amplitude $\sigma_i$ survives $N_i = C\sigma_i^{-m}$ cycles, where $C$ and $m$ are material constants. Accumulated damage is then $D = \sum_i n_i/N_i$, with $n_i$ the number of cycles at amplitude $\sigma_i$. Differentiating gives

$$\frac{\delta D}{D} \approx m\,\frac{\delta \sigma}{\sigma},$$

so a relative stress change caused by a different contour is multiplied by the exponent $m$, which is large for bone and metals. A second effect is not captured by this formula. Multi-axial gait loads move the stress hot spot over the loading cycle, so it can sweep through exactly the thin walls where raters disagree most. A single static instant may miss this.

What is new here is the downstream target. We propagate a learned distribution of rater-specific bone contours through mesh generation into cyclic damage. We then measure how often clinical verdicts flip, and we test which cheap surrogate for rater disagreement, if any, reproduces the real mechanical spread.

Where This Sits in Prior Work

The closest existing bridge is Statistical Finite Element Model for Bone Shape and Biomechanical Properties. It built a principal-component model of shape and CT density across 11 proximal femurs and found a nearly six-fold stiffness span across ±2 standard deviations of the first mode under a static 1,600 N load. That paper motivates our work, since geometry clearly moves mechanics. It differs from ours in three ways. Its variation is between patients rather than between raters on one scan. Its global shape modes cannot represent local contour disagreements. And it uses static load only.

A group of single-patient implant studies constrains our design and supplies our test cases. Biomechanical evaluation of ortho-bridge system and proximal femoral nail antirotation in intertrochanteric fractures with lateral wall fracture based on finite element analysis reports lateral-wall peak stress of 86.8 MPa with a nail alone versus 59 MPa with a supplementary plate, against a 60 MPa damage threshold, under a static walking instant. Biomechanical evaluation of oblique lateral interbody fusion with various fixation methods for degenerative lumbar scoliosis: a finite element analysis considering different bone densities makes similar threshold calls on endplates. Both rest on one segmentation of one person. Neither tests repeated loading, even though the first explicitly attributes clinical failures to repeated loading during rehabilitation. The research file contains an illustration in which two observers' wall thicknesses map onto 59 versus 86.8 MPa. That illustration is hypothetical: those numbers come from two implants, not two raters. Measuring whether rater spread is actually that large is the point of this work.

The case that dynamic loading reveals effects static loading hides comes from Stress-Transfer Characteristics of Stress-Absorbing Elements in the IMZ Implant System. Part II. Shock-absorbing implant elements made no measurable difference under static axial load but redistributed stress under dynamic horizontal and tilting loads. We borrow the hypothesis, not the method.

On the segmentation side, several papers supply the method for representing rater disagreement as a distribution rather than noise. Probabilistic Modeling of Inter- and Intra-observer Variability in Medical Image Segmentation (Pionono) gives each rater a Gaussian in a low-dimensional latent space. Its means capture systematic rater bias and its variances capture a rater's own inconsistency. Learning Calibrated Medical Image Segmentation via Multi-Rater Agreement Modeling and Uncertainty Quantification in Medical Image Segmentation with Multi-decoder U-Net produce calibrated uncertainty maps or rater-specific outputs. Calibrate the inter-observer segmentation uncertainty via diagnosis-first principle is the only one that judges segmentations by a downstream task, and that task is diagnostic classification. All four were validated on soft tissue, retina, skin or histology, and none ran a physical simulation downstream.

Benchmark of Segmentation Techniques for Pelvic Fracture in CT and X-ray: Summary of the PENGWIN 2024 Challenge documents genuine annotator disagreement over whether hairline lines separate fragments. It stops at geometric reduction planning and scores against a consolidated ground truth. QUBIQ notes that dense 3D uncertainty estimation is expensive. That cost constrains us to a generator that can sample many contours cheaply.

Research Program

The work has four stages.

The first stage collects data, which is unavoidable. The research file lists no publicly available multi-rater bone CT dataset with individual tracings. PENGWIN acknowledges multiple annotators but evaluates against a merged truth, and we could not confirm that per-rater masks are released. We will therefore have five raters independently segment two sets of scans. The first is about 30 pelvic fracture CT volumes drawn from the PENGWIN collection. The second is about 30 proximal femur CTs with intertrochanteric fractures involving the lateral wall. The raters span orthopedic surgeons, radiologists and trained engineers, and each re-traces a subset of the scans weeks later so we can measure within-rater inconsistency. This yields per-region maps of where disagreement concentrates, which is a result in its own right for bone.

The second stage builds the generator of plausible contours. We adapt a Pionono-style rater-latent model to 3D bone CT. To keep 3D sampling cheap, we run the latent model on narrow bands around a consensus surface rather than predicting the full volume. Each sample, defined by a rater identity and a latent draw, is turned into a surface and then a quadratic tetrahedral mesh. We use quadratic elements because linear tetrahedra become artificially stiff under bending-dominated loads. Element stiffness is assigned from CT density, as in the statistical femur model. A key detail is that density is read from the scan inside each sampled boundary, so moving the boundary also changes which voxels count as cortex.

The third stage defines the loads. The static reference reproduces the femur paper's walking instant: 2,469.6 N at the femoral head and 1,117.2 N from the abductor muscles. The cyclic case runs a full gait cycle of joint-reaction and muscle-force vectors scaled to the same peak, sampled at a few dozen instants. Each instant is solved as a linear elastic problem; inertia is negligible at walking frequency, so this is a quasi-static sweep. We then count stress cycles at each element and accumulate damage with the power-law relation above. The analysis covers a rehabilitation period of repeated cycles. For the pelvis, the same pipeline runs on representative screw fixations of PENGWIN fragments under gait loading of the hip socket.

The fourth stage compares implants under each source of uncertainty. Each fixation option (nail alone vs. nail plus plate; candidate pelvic constructs) is simulated across about 100 contour samples per scan. This gives distributions of peak stress, cyclic damage and fragment displacement rather than single numbers.

Baselines

The naive reference point is current practice: one segmentation per scan, standing in for the single Mimics-style tracing used in the femur and spine studies, analyzed under the static walking instant. Every uncertainty method is compared against the verdicts this single-mask pipeline produces. If those verdicts almost never flip when we add rater spread, the proposal's premise fails.

The strongest automated segmentation competitor is the boundary-core decomposition method from MIC-DKFZ that won PENGWIN. It reports a fragment-wise IoU of 0.930, a 95th-percentile Hausdorff distance (HD95) of 5.866 mm on CT, and a reduction-planning alignment error of 1.695 mm (PENGWIN summary). It is the strongest competitor because a reviewer will argue that a good automatic mask removes the need to model raters. We will retrain it on our consensus masks with the same data splits and GPU budget as our generator. Its mesh goes through the same meshing and loading pipeline. It "wins" if its single mechanical prediction lies near the center of the human-rater distribution and that distribution is narrow relative to the differences between implant designs. For the rater-modeling component, Pionono reports Cohen's kappa of 0.758 on Gleason 2019 (Pionono). That figure is on histology. For bone CT and for mechanical outputs, we found no reported result on Lacuna for propagating rater-specific segmentations into finite element stress or fatigue.

Two cheap surrogates for rater disagreement are the real competitors on the uncertainty question, and both get the same budget of 100 mechanical runs per scan. The first is a uniform surface offset: the consensus surface is grown and shrunk by an amount chosen to match the measured inter-rater spread in mean surface distance. The second is random, spatially smoothed Gaussian displacement of surface nodes, with the correlation length tuned on held-out raters. A third geometric baseline is the inter-patient principal-component approach of the statistical femur model, with mode amplitudes fit to the rater data. It tests whether global shape modes can stand in for local disagreement. The rater-conditioned generator beats these baselines only if its distribution of mechanical outputs matches held-out real raters' outputs significantly better. We measure this match with energy distance and coverage of the held-out rater's peak stress and damage.

The ablation that isolates the key ingredient is static against cyclic loading on the same contour samples, together with a rater-identity-removed generator that keeps only a pooled latent distribution. These separate the contribution of dynamic loading from the contribution of structured rater bias.

Why not just…?

Why not just use a strong automatic segmenter and one consensus mask? A deterministic mask gives one answer with no indication of how fragile that answer is. The multi-rater literature reports that such models are overconfident exactly at ambiguous boundaries (MRNet). Even the PENGWIN winner's 5.866 mm HD95 is large relative to a cortical wall a few millimeters thick. Still, this is an empirical question, and the automatic-segmenter baseline above answers it directly. If verdict-flip rates under rater spread turn out near zero, the consensus mask is enough, and we will say so.

Why not just shift the surface uniformly, or jitter mesh nodes randomly? This is the most serious threat. If a uniform offset matched to the measured spread reproduces the mechanical distribution, that offset is the practical recommendation and the learned generator is unnecessary. Our reason to doubt it is that rater disagreement is local and systematic: some raters consistently under-segment low-contrast planes, as Pionono's bias terms show. A uniform offset over-perturbs well-defined regions and under-perturbs the thin wall where the hot spot sits. Random jitter also tends to create kinks that produce artificial stress spikes. The surrogate-matching experiment settles this. We commit in advance to reporting the offset as sufficient if its energy distance to held-out raters falls within the generator's confidence interval.

Why not just run a worst-case static load with a safety factor? If the ranking of peak static stress across contour samples matches the ranking of cyclic damage, static analysis remains adequate and the dynamic stage adds cost without insight. The IMZ result (Sato et al.) shows that static loading can hide effects visible under off-axis dynamic loads. The power-law relation predicts amplification of contour-induced stress changes. Neither is proof for bone fixation, so we measure the rank correlation and the flip rates directly under both loading regimes.

Proposed Experiments and Decisive Checks

The first decisive check is the observer-versus-design ratio. For the femur case, we compute the spread (interquartile range) of lateral-wall peak stress and cyclic damage across real raters' contours. We divide it by the difference between the two implant designs computed on the consensus mask. A ratio near or above one, together with a verdict-flip rate above roughly 10% of rater samples crossing the 60 MPa threshold, would show that single-segmentation implant comparisons like those in the femur and spine studies are not reliable at their reported margins. A small ratio would show that they are reliable despite rater disagreement. Both outcomes would be publishable.

The second check is amplification. We compare the coefficient of variation of the damage sum with that of static peak stress across the same samples, and test whether the measured ratio approaches the exponent $m$. We also check whether the location of maximum damage moves into high-disagreement regions more often under gait loading than under the static instant. This is the direct test of the hypothesis that dynamic loading magnifies rater disagreement.

The third check is surrogate sufficiency. Leaving one rater out at a time, we ask which of four sources best predicts that rater's mechanical outputs: the rater-conditioned generator, the uniform offset, smoothed Gaussian jitter, or global principal-component modes. Outputs are compared by energy distance and 90% interval coverage. Repeating the analysis on the pelvic fracture scans tests whether the conclusions carry over from a thin-wall problem to a fragment-connectivity problem, where raters disagree about topology and not just boundary position.

Segmentation quality is reported throughout with Dice and HD95, so that mechanical conclusions can be read against familiar overlap scores. Whether overlap metrics predict mechanical disagreement at all is itself measured, as the rank correlation between per-sample HD95 and per-sample damage deviation.