When a segmentation model trained on CT is pointed at MRI, or at a new scanner vendor, it fails badly. On cardiac scans, Dice falls from the high 80s to 38% without adaptation. In interactive tools, the clinician repairs these failures with clicks. Today those clicks are thrown away when the scan is closed, so the same systematic error has to be corrected again on the next patient. We propose to keep what the clicks teach, but to store it in a small, persistent "input lens" for each scanner or protocol rather than in the network's weights. The lens is a learned adjustment of the image's low-frequency content: the smooth, global part of the image that carries contrast and intensity style. It is fitted online from the clinician's corrections, and the segmentation network is never modified.
The bet is that much of the cross-modality gap is a problem of appearance, and appearance is shared by every scan from the same machine. Human corrections are then a far better training signal for that appearance adjustment than the model's own pseudo-labels. If the bet holds, the tool needs fewer clicks with each successive scan from a new scanner. The source model stays intact, so its performance on the original modality cannot degrade. Switching scanners means switching a 2,048-number lens. The same experiments also answer a scientific question no one has asked in this form: how much of the CT-to-MRI shift can be fixed in input space alone, measured in expert clicks?
Core Claim
A frozen interactive 3D segmentor needs only a per-scanner low-frequency Fourier prompt, trained online from clicks and accepted masks. With that alone, it recovers most of the click savings of online weight adaptation on cross-modality streams. It does so with no loss on the source domain and more robustness to sloppy acceptance of masks. On the same data, it clearly beats unsupervised input-space adaptation, because human corrections break the self-confirming pseudo-label loop that stalls that approach.
Each piece is borrowed; the combination and the question are new. FVP supplies the mechanism. It freezes a segmentor and optimizes about 2,048 parameters in the low-frequency Fourier amplitude of the input, raising abdominal CT→MRI Dice from 0.517 to 0.733. It is fully unsupervised and batch-mode, and it depends on thresholded self-pseudo-labels. You Point, I Learn (OAIMS) supplies the regime. It updates network weights from clicks during deployment, cutting BraTS T1 clicks-to-80%-Dice from 10.6 to 4.4 (companion report). Its limits are that it works on 2D slices, adapts within MRI sequences, forgets the source domain, and assumes accepted masks are correct. We differ from OAIMS in where the correction is stored: inside the input transform, not in the weights. We differ from FVP in what supervises it: human corrections, not self-training. The work also diverges from DC-TTA, which adapts SAM per image on natural and camouflaged scenes and resets between images. Here nothing resets within a scanner, and the targets are 3D medical volumes crossing modalities.
Research Program
The system has three parts. First, an interactive 3D U-Net trained on the source modality with click channels. Following Zhang et al., sparse clicks are turned into distance-decaying maps. We choose a plain 3D U-Net deliberately: AMOS found it beats Swin-UNETR on CT (89.04% vs 86.32% mDice), and a small backbone keeps per-click updates cheap. Second, a prompt bank keyed by scanner or protocol, read from image metadata. Each entry is a learnable multiplicative mask on the lowest-frequency band of the 3D Fourier amplitude, while phase is left untouched. Third, an online update rule that fits the prompt as the clinician works.
For target scan $x$ with Fourier amplitude $A(x)$ and phase $\Phi(x)$, the prompted input is
$$\tilde{x} = \mathcal{F}^{-1}\big(A(x)\odot(1 + M_{\text{low}}\odot p_s)\, e^{i\Phi(x)}\big),$$
where $p_s$ is the prompt for scanner $s$ and $M_{\text{low}}$ restricts it to a small central frequency window. The prompt is updated after each click and after each accepted mask:
$$p_s \leftarrow p_s - \eta \nabla_{p_s}\Big[\sum_{c} L_{\text{CCG}}(f(\tilde{x}, c), y_c) + \lambda\, w\, L(f(\tilde{x}), \hat{y})\Big].$$
Here $f$ is the frozen segmentor, $c$ ranges over clicks with their implied labels $y_c$, $L_{\text{CCG}}$ is OAIMS's click-centered Gaussian loss, and $\hat{y}$ is the accepted mask. The weight $w$ down-weights accepted masks whose boundary is inconsistent with the model's test-time-augmentation variance. This is a lightweight use of the uncertainty signal from post-hoc out-of-domain calibration, which flags doubt but never corrects it.
Case-specific errors, such as a lesion's odd margin or a rater's habit, still get fixed by clicks. They cannot easily contaminate the prompt, because a global low-frequency change cannot express them. This asymmetry is the design's answer to the rater-versus-scanner confusion that MedUHIP raises: a prompt has little room to memorize one clinician's idiosyncrasies. We also measure the limit of the design. The residual gap left after the prompt converges is our estimate of the "content" part of the shift, which input space cannot reach.
Baselines
The naive reference point is the source-trained interactive 3D U-Net used as-is on the target stream. Clicks correct each scan, but nothing is remembered. This gives the no-memory click curve that everything must beat. On MM-WHS, the non-interactive version of this model is reported at 38.1% mean Dice for CT→MRI in Structure-Driven UDA.
The strongest competitor on the question we pose is OAIMS-style online weight adaptation. It uses the same backbone, the same click simulator, the same CCG loss and the same accepted-mask step, but updates all weights (and, in a variant, only the decoder). It is the only published method that turns clinician clicks into persistent learning under distribution shift. Its reported figures are 4.4 versus 10.6 clicks to 80% Dice on BraTS T1 (OAIMS) and convergence within 10 to 40 images (OAIMS). We reproduce the BraTS FLAIR→T1 setting first to confirm our implementation lands near those numbers, then run both methods on identical scan orderings. Learning rate and steps per click get equal tuning budgets on a held-out validation stream. We beat it if, by scan 20 of each target stream, our clicks to 85% Dice lie within 1 click of OAIMS while source-domain Dice stays within 0.5 point of the original. OAIMS is expected to drop on the source domain. Winning outright on clicks is not required. Matching clicks while eliminating forgetting is the claim.
The strongest autonomous competitor is unsupervised FVP run offline on the full unlabeled target set, then used as the starting prompt for a non-learning interactive model. It is reported at 0.733 Dice on abdominal CT→MRI in the FVP work. On MM-WHS, no reported FVP result was found on Lacuna. On the same MM-WHS split, Structure-Driven UDA reports 74.8% for CT→MRI. On CrossMoDA the top unsupervised entries reach 87.0% vestibular schwannoma and 85.7% cochlea median Dice (CrossMoDA 2021). Our method has no offline target data at scan zero, so we beat FVP if its prompts reach better zero-click Dice than FVP's after at most 10 corrected scans.
The key ablation replaces human-supervised prompt updates with FVP's own pseudo-label updates under identical online streaming. This isolates whether the gain comes from human signal or merely from online fitting. A second ablation removes the uncertainty weight $w$, to test robustness to bad accepted masks. A third ablation adds LoRA-style weight adapters on top of the prompt, to measure what content-level adaptation adds beyond input space.
Why not just…?
Why not just let OAIMS update the weights across scans? This is the real threat, and the proposal is built to settle it rather than argue it. OAIMS already persists across images within a stream, and weights have strictly more capacity than a 2,048-number prompt. On CT→MRI, where tissue contrast relations change rather than merely rescale, weights may simply win on clicks. The case for the prompt rests on two measurable properties. The first is source retention: OAIMS itself reports forgetting of the source distribution. The second is fragility when users accept imperfect masks. We test both directly with interleaved-scanner streams and with a simulated user who accepts at 75% Dice. If OAIMS matches us on forgetting under a simple rehearsal fix and beats us on clicks, we will report that the input-space restriction is not worth it.
Why not just run unsupervised adaptation (FVP or CycleGAN-style translation) offline and skip learning from the human? It needs a collection of unlabeled target scans before the first patient, which a new protocol does not have. Even with that collection, the best CrossMoDA methods still trail full supervision by 4–5% Dice on tumors (CrossMoDA 2021), and FVP tops out at 0.733 on abdominal CT→MRI. The pseudo-label ablation above shows whether human signal, rather than streaming, explains any gain.
Why not just collect the corrected masks and retrain in batch, as AMOS did for its annotations? The AMOS pipeline of seed labels, model proposals and radiologist correction works well offline. At a large enough label count it will beat any prompt. We therefore run a matched-label comparison: nnU-Net-style fine-tuning on the first $k$ accepted masks, evaluated for $k$ in {5, 10, 20, 40}. The question is whether our method reaches equal clicks sooner with no retraining cycle and no weight change. If batch fine-tuning at $k=10$ dominates, the online prompt's value reduces to deployment convenience, and we will say so.
Why not just use a promptable foundation model zero-shot? The evidence says this fails exactly where shift is hardest. Zero-shot SAM falls to 0.0688 Dice on some medical targets (benchmark repository). Training-free GazeRefine reaches only 80.28% on prostate MRI. We still include a SAM-based interactive baseline for click counts on the same streams.
Proposed Experiments
We use four settings. MM-WHS supplies CT→MRI and MRI→CT, which probes the asymmetry reported by Structure-Driven UDA (85.8% vs 74.8%). AMOS supplies CT→MRI and also MRI cross-vendor streams, where AMOS reports over 20% Dice drops. CrossMoDA supplies ceT1→hrT2, using its public annotated hrT2 validation data only for scoring. BraTS FLAIR→T1/T2 supplies the OAIMS reproduction.
Clicks come from a standard simulated user that places each click at the center of the largest error region. Real clinicians are not used for the main results. A small reader study with two radiologists on 20 AMOS MRI scans checks that the simulated conclusions hold with real noise.
Each method sees target scans in five random orders. The metrics are clicks to 80% and 85% Dice per scan as a function of stream position, Dice after 3 clicks, failure rate at a 20-click cap, zero-click Dice at stream end, and source-domain Dice before and after the stream. An interleaved stream alternating two MRI vendors tests whether per-scanner prompts avoid the interference that single-model adaptation suffers.
The decisive checks come in three parts. If the human-supervised prompt does not beat the pseudo-label prompt by at least 2 clicks at scan 20 on MM-WHS CT→MRI, the human signal is not doing the work, and the central claim fails. If the prompt's click curve plateaus far above OAIMS's on CT→MRI but matches it on vendor-to-vendor MRI, we will have mapped where input space suffices and where it does not, and the directional plateau gap becomes the measured content component of the shift. If OAIMS loses more than 3 Dice points on the source while our method loses none, and the click curves are within 1 click, the proposal's trade holds.
References
- FVP: Fourier Visual Prompting for Source-Free Unsupervised Domain Adaptation of Medical Image Segmentation
- You Point, I Learn: Online Adaptation of Interactive Segmentation Models for Handling Distribution Shifts in Medical Imaging
- DC-TTA: Divide-and-Conquer Framework for Test-Time Adaptation of Interactive Segmentation
- DINs: Deep Interactive Networks for Neurofibroma Segmentation in Neurofibromatosis Type 1 on Whole-Body MRI
- AMOS: A Large-Scale Abdominal Multi-Organ Benchmark for Versatile Medical Image Segmentation
- Improved Post-hoc Probability Calibration for Out-of-Domain MRI Segmentation
- MedUHIP: Towards Human-In-the-Loop Medical Segmentation
- Structure-Driven Unsupervised Domain Adaptation for Cross-Modality Cardiac Segmentation
- CrossMoDA 2021 challenge: Benchmark of cross-modality domain adaptation techniques for vestibular schwannoma and cochlea segmentation
- GazeRefine: Expert Gaze as a Test-Time Prompt for Training-Free Medical Image Segmentation