Wearable Device for Blood Pressure Monitoring Based on a Novel Algorithm With a Few Morphological Feature Parameters of Pulse Waves

Di Wu, Chengqi Zhang, Guannan Peng, Yanke Zhang, Weixuan Shao, Junwei Zhang, Yu Peng, Chang Zeng, Long Mo, Zhengchun LiuIEEE Sensors Journal 2025

Abstract

We proposed an algorithm for blood pressure estimation based on a few morphological feature parameters of pulse waves to solve the problem of over-reliance on high-quality pulse waves (involving numerous pulse wave morphological feature parameters) of blood pressure estimation based on single-channel photoplethysmography (PPG). We successfully implemented the algorithm in a self-developed miniaturized wearable device. Statistical analysis revealed that in the algorithm, a parameter calculated from pulse pressure difference (DP) was highly correlated (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${R}^{{2}} = {0.942}$ </tex-math></inline-formula>) with systolic blood pressure (Ps). We defined this parameter as k. We used multiple linear regression to establish a mapping relationship between the morphological feature parameters of pulse waves and DP. This relationship had a correlation coefficient of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${R}^{{2}} = {0.667}$ </tex-math></inline-formula>, <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${P}~\lt 0.001$ </tex-math></inline-formula>. We built a miniaturized platform for blood pressure measurement and developed a reflective PPG sensor packaged with polydimethylsiloxane to enable comfortable wearable blood pressure measurement. In uncalibrated randomized experiments, we first evaluated the performance of the algorithm on separate datasets. A total of 116 objects showed a mean absolute error (MAE) of 11.15 and 8.29 mmHg for Ps and diastolic blood pressure (Pd), respectively. In a human experiment with 31 subjects, the errors for Ps and Pd were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$5.29~\pm ~10.92$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$- 2.77~\pm ~8.71$ </tex-math></inline-formula> mmHg, respectively. The results of multistate continuous blood pressure monitoring experiments improved after calibration was performed once. In a prolonged blood pressure monitoring experiment, the errors for the Ps and Pd of subjects were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.11~\pm ~7.05$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$0.19~\pm ~2.69$ </tex-math></inline-formula> mmHg, respectively. The results of both experiments conducted in this work were close to the international standard (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$5~\pm ~8$ </tex-math></inline-formula> mmHg) issued by the American Association for the Advancement of Medical Instrumentation. Our algorithm also showed promising results in continuous blood pressure monitoring experiments after exercise and under breath-holding conditions.
BibTeX
@article{wu2025wearable,
  author = {Di Wu and Chengqi Zhang and Guannan Peng and Yanke Zhang and Weixuan Shao and Junwei Zhang and Yu Peng and Chang Zeng and Long Mo and Zhengchun Liu},
  title = {{Wearable Device for Blood Pressure Monitoring Based on a Novel Algorithm With a Few Morphological Feature Parameters of Pulse Waves}},
  journal = {IEEE Sensors Journal},
  year = {2025},
  doi = {10.1109/jsen.2025.3545423},
}

Loading related content...

Figure