Abstract: Venipuncture is a common step in clinical scenarios, and is with highly practical value to be automated with robotics. Nowadays, only a few on-shelf robotic systems are developed, however, they can not fulfill practical usage due to varied reasons. In this paper, we develop a compact venipuncture robot -- VeniBot, with four parts, six motors and two imaging devices. For the automation, we focus on the positioning part and propose a Dual-In-Dual-Out network based on two-step learning and two-task learning, which can achieve fully automatic regression of the suitable puncture area and angle from near-infrared (NIR) images. The regressed suitable puncture area and angle can further navigate the positioning part of VeniBot, which is an important step towards a fully autonomous venipuncture robot. Validation on 30 VeniBot-collected volunteers shows a high mean dice coefficient (DSC) of 0.7634 and a low angle error of 15.58° on suitable puncture area and angle regression respectively, indicating its potentially wide and practical application in the future.
Introduction
Venipuncture is one of the most common medical procedures, performed over 1.4 billion times annually in the U.S. alone. Despite its frequency, it is surprisingly difficult; success rates can drop below 50% when dealing with elderly patients, infants, or patients in shock. Failed attempts lead to hematomas, nerve damage, and increased infection risk. While some robotic assistants exist, they are often bulky or require significant human intervention. The core technical challenge is not just "seeing" a vein, but identifying a suitable puncture area—a segment that is straight, long, and thick enough—while ignoring bifurcations, sharp curves, or noisy skin features.
The VeniBot Hardware Platform
The researchers developed VeniBot, a compact robotic system designed to automate the positioning and puncturing process. The hardware consists of four main modules:
- Imaging: Uses a Near-Infrared (NIR) camera to see through skin and an ultrasound probe for depth.
- Positioning: A 4-Degree-of-Freedom (DoF) system (X, Y, Z translation and Z-axis rotation) that aligns the needle.
- Puncturing: The mechanism that executes the needle insertion.
- Supporting: The base frame.
Dual-In-Dual-Out (DIDO) Network Architecture
Standard "Single-In-Single-Out" networks often fail to handle the abstract concept of "suitability" in vein selection. To overcome this, the authors proposed the Dual-In-Dual-Out (DIDO) architecture, which utilizes two-step and two-task learning.
In the two-step learning phase, a network first learns to segment all visible veins. A second network then takes both the original NIR image and the generated vein mask as inputs. This allows the AI to see both the "big picture" of the vein structure and the fine-grained skin textures needed to judge quality.
Simultaneously, the two-task learning approach enables the network to predict both the specific mask for the best puncture area and the longitudinal angle of the vein. To ensure stability during training, the authors transformed the angle into a continuous mathematical representation, avoiding the erratic behavior often seen in circular coordinate systems.
Key Results and Discussion
The proposed DIDO network significantly outperformed standard deep learning baselines using a dataset of 900 NIR images from 30 volunteers:
- Puncture Area Regression: Achieved a Mean Dice Coefficient (DSC) of 0.7634, compared to basic models that struggled to reach 0.56.
- Angle Accuracy: Achieved a low angle error of 15.58°.
- Efficiency: The entire training process took only about one hour on a consumer-grade GPU (NVIDIA GTX 1050Ti), making it computationally accessible.
While VeniBot excels at positioning the needle above the skin, this study focused primarily on that stage. Future work will involve the full integration of the positioning and puncturing systems into a single, seamless autonomous workflow, alongside testing on more diverse patient populations to ensure clinical reliability.