Use of Expected Utility (EU) to Evaluate Artificial Intelligence-Enabled Rule-Out Devices for Mammography Screening

Kwok Lung Fan, Yee Lam Elim Thompson, Weijie Chen, Craig K. Abbey, F. W. SamuelsonarXiv 2023 · DOI

TL;DR

Evaluating AI "rule-out" devices in mammography using standard sensitivity and specificity is fundamentally flawed because these metrics move in opposite directions by design. This paper proposes using Expected Utility (EU)—a metric that weights the clinical value of a true positive against the cost of a false negative—to provide a single, unambiguous measure of whether an AI actually improves screening performance.

What Problem Does This Paper Address?

In breast cancer screening, the workload for radiologists is massive (40 million mammograms annually in the U.S. alone). AI "rule-out" devices aim to help by autonomously removing low-risk images from the reading queue.

The evaluation challenge is the "Ambiguity Trap." In a retrospective simulation, if an AI rules out a case that a radiologist would have correctly identified as cancer, sensitivity drops. If it rules out a case that would have been a false alarm (false positive), specificity rises. Because rule-out devices almost always do both, researchers are left with a trade-off: is a 1% gain in specificity worth a 0.5% drop in sensitivity? Standard metrics can't answer this, making it difficult for regulators and clinicians to determine if a device is truly "effective."

Methods

The authors propose two alternative frameworks to resolve this ambiguity:

  1. PPV/NPV (Predictive Values): These measures can be more statistically powerful than sensitivity/specificity, but they require "ground truth" for every patient—including those the radiologist dismissed—which is rarely available in real-world screening data.
  2. Expected Utility (EU): This is the paper's primary focus. EU assigns a "utility" (value) to each possible outcome (True Positive, False Positive, etc.).
  3. Relative Utility ($U_{Rel}$): This represents how many times more "valuable" it is to find a cancer than it is "costly" to have a false alarm.
  4. Iso-utility lines: By plotting these on an ROC curve, researchers can see if an AI's operating point sits above or below the "break-even" line of current human performance.
  5. Practical Advantage: Crucially, the authors show that EU can be calculated using only recall rates and cancer detection rates, which are standard metrics tracked by hospitals even when full ground truth is missing.

Datasets

The authors re-analyzed data from two major recent publications to test their EU framework:

  • U.S. Study (Yala et al., 2019): A dataset of 26,540 mammograms from 7,176 patients with a 0.7% cancer prevalence.
  • European Study (Larsen et al., 2022): A massive dataset of 122,969 mammograms from 47,877 patients in Norway, reflecting a double-reading workflow.
  • Baseline Utility Data: To set the "value" of finding a cancer, they used historical data from the DMIST study (U.S.) and Otten et al. (Europe) to estimate that U.S. radiologists operate at a $U_{Rel}$ of ~162, while European double-reading operates at ~111.

Key Insights and Ideas

  • Sensitivity is a Floor: In a rule-out scenario, the combined AI+Radiologist sensitivity can only stay the same or go down compared to a radiologist alone. This makes "non-inferiority" the best-case scenario for sensitivity, which feels counter-intuitive to developers.
  • The Single-Metric Solution: EU collapses the Se/Sp trade-off into a single number. If the EU of the AI-enabled workflow is higher than the baseline, the device provides a net benefit to the population.
  • Regional Differences Matter: The "value" of a diagnosis isn't universal. Because European screening uses double-reading and has lower recall rates, the baseline utility and the impact of AI are different than in the U.S. single-reader system.
  • The "Safe Zone": The study suggests that ruling out a small fraction of patients (e.g., <30%) results in a negligible drop in EU, suggesting these devices could safely reduce workload without significantly harming overall screening quality.

Key Results

  • U.S. Re-analysis: Despite the original study's claims of "improved performance," the EU analysis showed no statistically significant improvement at any of the reported AI thresholds. The probability that the AI workflow actually had higher utility than the baseline was only about 72%—not enough for a definitive "win."
  • European Re-analysis: As the AI ruled out more patients, the EU consistently dropped. The AI removed enough "true positive" cases that the benefit of fewer false alarms couldn't compensate for the missed cancers.
  • Workload vs. Performance: While the AI didn't "improve" the clinical outcome, it did maintain near-baseline utility while potentially reducing the radiologist's workload by 30%. This shifts the value proposition from "better medicine" to "better efficiency."

Limitations and Future Work

  • Static Thresholds: The model assumes radiologists don't change their behavior when they know an AI has already filtered the "easy" cases. In reality, radiologists might become more suspicious of the remaining cases, shifting their own internal thresholds.
  • Utility Subjectivity: The entire EU framework depends on the $U_{Rel}$ value (e.g., 162). If society or a specific hospital values finding a cancer even more highly, the AI's performance would look worse.
  • Subgroup Analysis: The paper notes that utility might vary across patient subgroups (e.g., those with dense breast tissue). Future work should involve calculating relative utility baselines for these specific populations to ensure equitable AI performance.

Summary updated

BibTeX
@misc{fan2023use,