Upper Limb Compensatory Strategies Across Tasks with an Ultrasound-Controlled Prosthesis

Hanadi Hakami, Bruno Grandi Sgambato, Matthew Banger, Dario Farina, Alison H. McGregorResearch Square 2025 · DOI

Abstract

Objective: Upper-limb amputation often leads to compensatory trunk and shoulder movements, increasing the risk of secondary musculoskeletal complications. This issue is exacerbated by the limited wrist functionality in current prosthetic systems, which restrict natural movement patterns during daily activities. Here, we investigate how limitations in the degrees of freedom (DoF) of the prosthetic wrist influence compensatory upper body motion during functionally relevant tasks. Methods: Three transradial amputees, experienced in the use of both body-powered and myoelectric prostheses, were fitted with custom sockets and a prosthetic hand controlled via a 32-channel A-mode ultrasound interface. Eight able-bodied participants served as controls. All participants performed three standardized tasks—drinking, lightbulb insertion, and the Clothespin Relocation Test—while trunk and shoulder kinematics and kinetics were recorded using motion capture and surface electromyography. Results: The results demonstrated task-dependent compensation, with the Clothespin Relocation Test eliciting the greatest trunk flexion and bilateral shoulder involvement. Distinct adaptive patterns emerged between dominant and non-dominant sides, with increased reliance on proximal joints as wrist DoFs were restricted. Conclusion: The findings highlight the need for prosthetic designs and rehabilitation strategies that are tailored to specific tasks and user movement patterns. Quantifying compensatory motion provides a foundation for developing user-centered control systems that enhance function and reduce long-term musculoskeletal strain.

Introduction

Current upper-limb prostheses often lack the dexterity of a natural hand, particularly at the wrist. When a prosthetic wrist cannot rotate, flex, or extend, the user must compensate by moving their shoulder or leaning their entire torso to position the hand correctly. While these "compensatory movements" allow for task completion in the short term, they lead to chronic pain, fatigue, and overuse injuries in the "sound" joints. Traditional control methods, like surface electromyography (sEMG), struggle to provide intuitive, simultaneous control over multiple joints. This study investigates how different levels of wrist mobility influence these compensatory patterns across various daily tasks, using a high-fidelity ultrasound interface to bypass the limitations of standard sEMG control.

Methods

The researchers employed a 32-channel A-mode ultrasound interface integrated into custom-fabricated sockets. Unlike sEMG, which listens to electrical signals, this ultrasound human-machine interface (HMI) tracks actual muscle deformation (morphing) inside the residual limb to predict intended movement.

Experimental Setup

Three transradial amputees and a control group of eight able-bodied individuals participated. The hardware consisted of a modified Michelangelo Hand (Ottobock) capable of multiple wrist DoFs. Participants performed three standardized tests: Drinking, Lightbulb Insertion (LB), and the Clothespin Relocation Test (CPRT). The researchers tested five wrist configurations ranging from "Open/Close only" (no wrist movement) to full "Flexion/Extension + Pronation/Supination."

Data Collection

Kinematic data was gathered using a 28-camera Vicon motion capture system, while 12-sensor sEMG recorded muscle activity to measure the physiological effort of compensation. A novel comparison framework was introduced to map specific prosthetic DoF configurations against able-bodied reference data to create "difference profiles."

Results

The study found that compensation scales with task demands. The CPRT was the most demanding, requiring significant trunk flexion and bilateral shoulder involvement. Restricting wrist flexion/extension consistently triggered the highest compensatory range of motion (RoM) in the shoulder and thorax, suggesting that wrist pitch is more critical for reducing body strain than wrist roll (pronation/supination) for many daily tasks.

Kinematic profiles for the Lightbulb (LB) task, illustrating the impact of wrist mobility on joint angles compared to able-bodied controls.

During the CPRT, prosthesis-side shoulder abduction reached 62.8°, significantly higher than the contralateral side (24.6°). Thoracic rotation peaked at 97.0° during the most complex tasks. Furthermore, the ultrasound-controlled "Open/Close only" configuration caused a massive spike in lower trapezius activity (up to 863.2 µV), indicating a heavy "scapular tax" paid when the wrist is locked.

Statistical comparison of compensatory joint angles across different wrist configurations, highlighting significant reductions in trunk and shoulder movement with increased degrees of freedom.

Discussion

Providing both flexion/extension and pronation/supination led to more stable muscle activation in postural muscles (RMSD < 4 µV), suggesting a more efficient, "natural" control state. Compensation is not just about the prosthetic side; the "sound" side shoulder showed increased axial rotation to support postural stability and reach, highlighting that unilateral amputation is a whole-body biomechanical issue. Ultrasound control provides a "neutral" context for evaluation, as users do not have the decades of "bad habits" often associated with sEMG, allowing researchers to see how they naturally adapt to new DoFs.

Conclusion

This research quantifies how prosthetic wrist limitations force amputees to use compensatory movements, leading to long-term musculoskeletal strain. The findings highlight the need for prosthetic designs and rehabilitation strategies tailored to specific tasks. Future work should focus on user-adaptive systems that automatically adjust wrist DoFs to mitigate strain and long-term studies to investigate if multi-DoF wrists reduce the incidence of clinical musculoskeletal disorders.

Summary updated

Abstract

<title>Abstract</title> <bold>Objective</bold> : Upper-limb amputation often leads to compensatory trunk and shoulder movements, increasing the risk of secondary musculoskeletal complications. This issue is exacerbated by the limited wrist functionality in current prosthetic systems, which restrict natural movement patterns during daily activities. Here, we investigate how limitations in the degrees of freedom (DoF) of the prosthetic wrist influence compensatory upper body motion during functionally relevant tasks. <bold>Methods</bold> : Three transradial amputees, experienced in the use of both body-powered and myoelectric prostheses, were fitted with custom sockets and a prosthetic hand controlled via a 32-channel A-mode ultrasound interface. Eight able-bodied participants served as controls. All participants performed three standardized tasks—drinking, lightbulb insertion, and the Clothespin Relocation Test—while trunk and shoulder kinematics and kinetics were recorded using motion capture and surface electromyography. <bold>Results</bold> : The results demonstrated task-dependent compensation, with the Clothespin Relocation Test eliciting the greatest trunk flexion and bilateral shoulder involvement. Distinct adaptive patterns emerged between dominant and non-dominant sides, with increased reliance on proximal joints as wrist DoFs were restricted. <bold>Conclusion</bold> : The findings highlight the need for prosthetic designs and rehabilitation strategies that are tailored to specific tasks and user movement patterns. Quantifying compensatory motion provides a foundation for developing user-centered control systems that enhance function and reduce long-term musculoskeletal strain.