Uncertainty Evaluation of a Broadband Attenuation Standard

Jaegab Lee, J.-h. Kim, Jong-In Park, U KangIEEE Trans. Instrumentation & Measurement 2005

Abstract

A standard step attenuator has been developed. It comprises a build-up chain of ten steps of power ratio measurement and operates in a frequency range of 10 MHz to 18 GHz. The overall system has a dynamic range of 90 dB and an expanded uncertainty of 0.006 dB (k=2). The system is used as an attenuation standard in a series radio-frequency substitution scheme. The uncertainty factors in the attenuation measurement of 40 dB at 5 GHz are fully described. For the calibration of a typical step attenuator, the expanded uncertainties are 0.011 dB for the attenuation range up to 80 dB, including the mismatch uncertainty.
BibTeX
@article{lee2005uncertainty,
  author = {Jaegab Lee and J.-H. Kim and Jong-In Park and U Kang},
  title = {{Uncertainty Evaluation of a Broadband Attenuation Standard}},
  journal = {IEEE Transactions on Instrumentation and Measurement},
  year = {2005},
  doi = {10.1109/tim.2004.843336},
}