Ultrasonic Vocalizations test in adult rodents v1

Amirah Wright, Madison Wypyski, Susan A. Murphy, Pamela J. VandeVord2025 · DOI

TL;DR

This protocol establishes a standardized method for recording and analyzing ultrasonic vocalizations (USVs) in adult rats to assess emotional states like distress and pleasure. By providing specific software parameters and hardware configurations, the authors aim to reduce inter-laboratory variability and improve the clinical translation of Traumatic Brain Injury (TBI) research.

What Problem Does This Paper Address?

In behavioral neuroscience, understanding a subject's internal emotional state is notoriously difficult. While rats are highly social and vocal, they communicate primarily in the ultrasonic range (above 20 kHz), which is inaudible to humans. These vocalizations are critical indicators of affect: "alarm" calls (around 22 kHz) signal distress or pain, while "chirps" (around 50 kHz) are associated with positive states like play or reward anticipation.

The primary challenge addressed here is reproducibility. Different laboratories often use varying recording durations, frequency thresholds, and environmental setups, making it nearly impossible to compare results across studies. This protocol is part of the PRECISE-TBI initiative, which seeks to create a standardized "online catalog" of methods to ensure that data collected at one institution can be meaningfully compared to data from another, ultimately speeding up the development of therapeutics for brain injuries.

Methods

The protocol outlines a non-invasive approach using specialized hardware and automated analysis software. Key components include:

  • Hardware: A Pettersson M500-384 USB microphone capable of capturing high-frequency audio, placed inside a sound-attenuating chamber to eliminate background noise.
  • Environment: The animal is recorded in its home cage (with the lid removed) to minimize the stress of a novel environment, ensuring the captured vocalizations are a response to the experimental condition rather than the testing apparatus itself.
  • Software Automation: The authors utilize UltraVox XT (Noldus) software. Rather than manually "scoring" audio files—which is time-consuming and prone to human error—the protocol defines specific algorithmic parameters to automatically detect and categorize calls.
  • Standardized Duration: A fixed 5-minute recording window is established to ensure consistency across trials.

Datasets

As this is a methodology protocol rather than a primary research study, it does not introduce a new public dataset. Instead, it provides the configuration parameters (the "metadata" for future datasets) required for researchers to generate comparable data. It references the PRECISE-TBI repository (precise-tbi.org) as the hub for these standardized protocols.

Key Insights and Ideas

  • Frequency Binning: The protocol identifies two distinct "bins" for analysis.
  • 22-kHz calls are defined by a frequency range of 18–36 kHz and a longer duration (up to 300ms).
  • 50-kHz calls are defined by a higher range (40–60 kHz) and much shorter, rapid bursts (10–150ms).
  • The "Home Cage" Advantage: By recording in the home cage, researchers can capture "baseline" emotional states or responses to specific treatments without the confounding variable of "handling stress" or "novelty-induced anxiety."
  • Multi-modal Integration: A significant insight is the mention of external triggers. The protocol is designed to be paired with other behavioral tests (like the Open Field or Elevated Plus Maze), allowing researchers to sync audio data with physical movement data.

Key Results

The protocol provides the exact quantitative "Call Definitions" used for automated detection:

Parameter 22-kHz (Distress) 50-kHz (Pleasure)
Frequency Low 18,000 Hz 40,000 Hz
Frequency High 36,000 Hz 60,000 Hz
Min Duration 20 ms 10 ms
Max Duration 300 ms 150 ms
Min Gap 10 ms 10 ms

By using these specific settings, researchers can generate automated counts and mean durations of calls, providing an objective metric for "affective state" in rodent models of TBI.

Limitations and Future Work

  • Hardware Constraints: The current setup is limited to a maximum of two microphones per computer, which may restrict high-throughput screening in larger labs.
  • Environmental Sensitivity: The protocol heavily relies on a sound-attenuating chamber. Without this, ambient noise (HVAC systems, computer fans) can create "false positives" in the 20-60 kHz range.
  • Data Export: Currently, the software defaults to .txt file exports for statistics. Future iterations might benefit from more robust data formats (like JSON or CSV) that integrate more easily with modern data science pipelines (R or Python).
  • Future Directions: The authors suggest that these USV signatures could eventually serve as a "biomarker" for injury severity in TBI models, helping researchers determine if a therapeutic intervention is actually improving the animal's well-being, not just its motor function.

Summary updated

Abstract

This is a protocol to describe the materials and methods utilized to measure ultrasonic vocalizations (USVs) in adult rats. The model utilizes microphones to detect USVs produced by rats while in their home cage. The model utilizes microphones to detect USVs produced by rats while in their home cage to examine different emotional states, such as distress and pleasure. The posting of this protocol is part of the mission of the PREClinical Interagency reSearch resourcE-TBI (PRECISE-TBI, precise-tbi.org) to improve clinical translation of therapeutics by providing an online catalog and standardized protocols to reduce the variability of model usage between laboratories.
BibTeX
@misc{wright2025ultrasonic,
  author = {Amirah Wright and Madison Wypyski and Susan A. Murphy and Pamela J. VandeVord},
  title = {{Ultrasonic Vocalizations test in adult rodents v1}},
  year = {2025},
  doi = {10.17504/protocols.io.kqdg319x7l25/v1},
}
Figure