Two-Year Outcomes of Visian Implantable Collamer Lens with a Central Hole for Correcting High Myopia

Zhipeng Yan, Huamao Miao, Feng Zhao, Xiaoying Wang, Xun Chen, Meiyan Li, Xingtao ZhouJournal of Ophthalmology 2018 · DOI

TL;DR

This study demonstrates that the Visian Implantable Collamer Lens with a central hole (ICL V4c) is a safe and effective option for correcting high myopia, achieving excellent refractive outcomes and visual acuity improvements over a two-year period. Notably, despite a significant reduction in anterior chamber volume, intraocular pressure remained stable, suggesting the central hole design effectively manages aqueous humor circulation.

What Problem Does This Paper Address?

High myopia, particularly in young adults, is a significant public health concern. While laser refractive surgery is common, it has limitations for very high degrees of myopia. Phakic intraocular lenses (IOLs) offer a solution for a wider range of myopia, especially for those unsuitable for corneal surgery. The Visian ICL V4c is a newer generation of these lenses, featuring a central hole designed to improve aqueous humor circulation and eliminate the need for a peripheral iridotomy (a surgical procedure to create an opening in the iris). This study aims to evaluate the long-term (two-year) safety, efficacy, and predictability of the ICL V4c in correcting high myopia, specifically investigating its impact on anterior segment anatomy and intraocular pressure.

Methods

This was a prospective, non-randomized study involving 32 patients (61 eyes) with high myopia (spherical equivalent of -8.00 D or higher) who underwent ICL V4c implantation. The researchers evaluated several parameters two years post-surgery:

  • Refractive Outcomes: Spherical equivalent (SE) refraction, corrected distance visual acuity (CDVA), and uncorrected distance visual acuity (UDVA).
  • Efficacy and Safety: Calculated using efficacy and safety indices, which compare postoperative visual acuity to preoperative values.
  • Anterior Segment Anatomy: Measured using a Scheimpflug tomography imaging system (Pentacam) to assess anterior chamber volume (ACV), anterior chamber depth (ACD), anterior chamber angle width (ACAW), and the ICL vault (the distance between the ICL and the crystalline lens).
  • Intraocular Pressure (IOP): Measured using a noncontact tonometer.
  • ICL Haptic Position: Assessed using ultrasound biomicroscopy (UBM).

Statistical analysis employed linear mixed models to compare variables over time and correlation analysis to explore relationships between continuous variables.

Datasets

Not discussed in the paper. The study utilized patient data collected prospectively from routine ophthalmic examinations at the Refractive Surgery Center of the Eye and ENT Hospital of Fudan University.

Key Insights and Ideas

  • The central hole is key: The ICL V4c's central hole appears to be effective in maintaining stable intraocular pressure, even with a significant reduction in anterior chamber volume. This suggests it successfully facilitates aqueous humor flow, mitigating a potential risk associated with ICL implantation.
  • High myopia correction is achievable: The lens demonstrated excellent predictability, with a vast majority of eyes achieving their intended refractive correction within ±1.00 D. This is particularly impressive given the high degree of myopia in the study population.
  • Visual acuity gains are substantial: A significant portion of patients experienced improved visual acuity, with many gaining one or more lines of CDVA. This highlights the functional benefit of the ICL V4c.
  • Anterior chamber changes are expected but manageable: While the ICL V4c implantation led to a notable decrease in anterior chamber volume and angle width, these changes did not negatively impact intraocular pressure over the two-year follow-up. This suggests the eye adapts to the reduced space.
  • Stable lens position: The ICL haptics were found to be stably positioned, primarily in the ciliary processes, which is considered a safe location and did not correlate with increased IOP or anterior subcapsular opacification.

Key Results

  • Refractive Correction: The mean spherical equivalent refraction improved from -14.62 ± 4.29 D preoperatively to -0.90 ± 0.95 D at two years. 79% of eyes were within ±0.50 D and 98% within ±1.00 D of the intended correction.
  • Visual Acuity: The efficacy index was 1.03 ± 0.23, and the safety index was 1.24 ± 0.26. 75% of eyes gained one or more lines of CDVA, and only 2% lost one line.
  • Anterior Chamber Changes: ACV decreased from 193.28 to 112.48 mm³, ACD from 3.15 to 2.99 mm, and ACAW from 36.51 to 22.54 degrees (all statistically significant, p < 0.001).
  • Intraocular Pressure: IOP remained stable, averaging 15.39 mmHg preoperatively and 15.86 mmHg at two years (p = 0.11).
  • Patient Satisfaction: All patients were satisfied with their visual performance, with a mean overall satisfaction score of 9.27 ± 0.87.

Limitations and Future Work

  • Limitations:
    • The study was non-randomized, which can introduce bias.
  • The follow-up period was two years; longer-term outcomes, particularly regarding potential cataract formation, would be beneficial.
  • While the central hole design is highlighted, direct comparative studies with lenses without a central hole would further elucidate its specific benefits.
  • Future Work:
    • Longer-term follow-up studies are needed to monitor for late complications like cataract formation or glaucoma.
  • Further investigation into combining ICL V4c with corneal refractive surgery for extremely high myopia could be explored.
  • Continued monitoring of anterior chamber changes and their long-term implications is warranted.

Summary updated

Abstract

Purpose . To investigate two-year outcomes of Visian Implantable Collamer Lens with a central hole (ICL V4c) implantation for correcting high myopia. Methods . Sixty-one eyes of 32 patients went through ICL V4c implantation. Safety, efficacy, predictability, and intraocular pressure were evaluated 2 years postoperatively. Anterior chamber volume (ACV), anterior chamber depth (ACD), anterior chamber angle width (ACAW), and vault were measured using a Scheimpflug tomography imaging system. Results . At 2 years, the spherical equivalent refraction decreased from preoperative −14.62 ± 4.29 D to −0.90 ± 0.95 D, with 79% of the eyes within ±0.50 D and 98% within ±1.00 D of the intended correction. The efficacy index was 1.03 ± 0.23, and the safety index was 1.24 ± 0.26. Corrected distance visual acuity (CDVA) remained unchanged in 23% of the eyes, 75% gained one or more lines of CDVA, and 2% lost one line. The ACV, ACD, and ACAW, respectively, decreased from 193.28 ± 29.15 mm 3 , 3.15 ± 0.23 mm, and 36.51 ± 6.54 degree to 112.48 ± 17.01 mm 3 , 2.99 ± 0.23 mm, and 22.54 ± 5.27 degree (<mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="M1"><mml:mrow><mml:mi>p</mml:mi><mml:mo>=</mml:mo><mml:mn>0.0008</mml:mn><mml:mo>,</mml:mo><mml:mtext> </mml:mtext><mml:mn>0.008</mml:mn><mml:mo>,</mml:mo><mml:mtext> and </mml:mtext><mml:mn>0.0003</mml:mn></mml:mrow></mml:math>, resp.). Intraocular pressure was 15.39 ± 2.88 mmHg before surgery and was 15.86 ± 4.11 mmHg at 2 years (<mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="M2"><mml:mrow><mml:mi>p</mml:mi><mml:mo>=</mml:mo><mml:mn>0.11</mml:mn></mml:mrow></mml:math>). Conclusion . Implantation of ICL V4c is a safe, effective, and predictable procedure for correcting high myopia. Reduction of anterior chamber space after surgery did not induce intraocular pressure increase during the 2-year follow-up.
BibTeX
@article{yan2018two,
  author = {Zhipeng Yan and Huamao Miao and Feng Zhao and Xiaoying Wang and Xun Chen and Meiyan Li and Xingtao Zhou},
  title = {{Two-Year Outcomes of Visian Implantable Collamer Lens with a Central Hole for Correcting High Myopia}},
  journal = {Journal of Ophthalmology},
  year = {2018},
  doi = {10.1155/2018/8678352},
}