Tracking local volume 3D‐echo‐planar coronary artery imaging

Peter Gatehouse, Jennifer Keegan, Guang Yang, Raad Mohiaddin, David FirminMagnetic 2001 · DOI

Abstract

Aiming for robust high-resolution free respiration coronary artery imaging, localized tracking volume selective 3D-EPI at spatial resolution 1.25 × 1.25 × 3 mm and temporal resolution 65 ms was evaluated in 14 normal subjects. Subject-specific motion tracking factors were measured between diaphragm navigator and coronary artery in S-I and A-P directions. Imaging was repeated with and without tracking, accepting 128 cardiac cycles over a 20-mm range from free breathing. Reference images with minimal respiratory motion (5-mm range) used LED-guided multiple breathholds. Depending on accurate motion measurement, coronary arteries were imaged with tracking, in a 20-mm range of free breathing, with increased scan efficiency but without significant loss of image quality compared to multiple breathhold imaging. Magn Reson Med 46:1031–1036, 2001. © 2001 Wiley-Liss, Inc.
BibTeX
@inproceedings{gatehouse2001tracking,
  author = {Peter Gatehouse and Jennifer Keegan and Guang Yang and Raad Mohiaddin and David Firmin},
  title = {{Tracking local volume 3D‐echo‐planar coronary artery imaging}},
  booktitle = {Magnetic},
  year = {2001},
  doi = {10.1002/mrm.1293.abs},
}

Loading related content...

Figure