Tracking local volume 3D‐echo‐planar coronary artery imaging
Lists
Only you see this
Abstract
Aiming for robust high-resolution free respiration coronary artery imaging, localized tracking volume selective 3D-EPI at spatial resolution 1.25 × 1.25 × 3 mm and temporal resolution 65 ms was evaluated in 14 normal subjects. Subject-specific motion tracking factors were measured between diaphragm navigator and coronary artery in S-I and A-P directions. Imaging was repeated with and without tracking, accepting 128 cardiac cycles over a 20-mm range from free breathing. Reference images with minimal respiratory motion (5-mm range) used LED-guided multiple breathholds. Depending on accurate motion measurement, coronary arteries were imaged with tracking, in a 20-mm range of free breathing, with increased scan efficiency but without significant loss of image quality compared to multiple breathhold imaging. Magn Reson Med 46:1031–1036, 2001. © 2001 Wiley-Liss, Inc.