NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

Towards real-time alignment of 3D CT and 2D X-Ray with multi-stage CNNs

Federica Facente, Benjamin Billot, Vivek Gopalakrishnan, Manasi Kattel, Wen Wei, Polina Golland, Hervé Delingette, Isaac Cohen, Pierre Berthet-RayneSPIRE 2026

TL;DR

This paper introduces LXPose, a novel multi-stage deep learning framework that achieves real-time alignment of 3D CT scans with intraoperative 2D X-ray images. By employing a two-stage CNN and eliminating the need for manual annotations through automated landmark extraction, LXPose significantly speeds up the registration process while maintaining high accuracy, making it suitable for live image-guided interventions.

What Problem Does This Paper Address?

Image-guided interventions, particularly those using X-ray fluoroscopy, are crucial for modern surgery. While X-rays provide real-time visualization, they lack depth information and soft-tissue contrast. To overcome this, surgeons rely on preoperative 3D CT scans for detailed anatomical context. The challenge lies in mentally aligning these two modalities during surgery, which is cognitively demanding and prone to errors. Traditional methods for aligning 3D CT with 2D X-rays (3D/2D registration) often involve slow optimization processes and require manual annotations, hindering their use in real-time clinical settings where latencies of around 40ms are desired. This paper addresses the need for a fast, accurate, and automated solution for this critical registration task.

Methods

The core of LXPose is a two-stage cascaded CNN architecture.

  • Stage 1 (Pose Regressor): The first CNN takes an input X-ray image and directly predicts an initial estimate of the C-arm's pose (rotation and translation). This stage uses an axis-angle representation for rotation, which the authors found to be more robust than Euler angles.
  • Stage 2 (Registration Module): The second CNN refines this initial pose. It takes both the input X-ray and a digitally reconstructed radiograph (DRR) generated from the initial pose estimate. By learning to correct the error between the initial and ground-truth poses, this stage effectively replaces slow optimization-based refinement.

Key innovations include:

  • Multi-stage CNN: Progressively refines pose estimates for improved accuracy.
  • Automated Landmark Extraction: Eliminates the need for manual annotation of landmarks on CT scans by using automated methods (e.g., TotalSegmentator, marching cubes, farthest point sampling) to generate landmarks from CT volumes.
  • 2D Projection-Based Loss (mPD): The training objective is defined in the 2D X-ray space, directly optimizing for accurate projection of 3D anatomy, which is most relevant for image guidance.
  • Homography-based Normalization: Addresses variations in X-ray intrinsic parameters (focal length, principal point) during real-time use by resampling images to a fixed training geometry.
  • Aggressive Data Augmentation: Mimics realistic X-ray variations (Gaussian noise, blur, sharpening, gamma, salt-and-pepper, plasma intensity variations) to reduce the domain gap between synthetic DRRs and real X-rays.

Datasets

  • DeepFluoro: An open-source dataset containing CT volumes and X-ray series from six cadaveric subjects. It provides ground-truth extrinsic camera parameters and 14 manual 3D anatomical landmarks with their 2D projections. This dataset was used for both training (using simulated DRRs) and testing.
  • Ljubljana: A dataset from 10 endovascular neurointervention patients, including 3D rotational DSA and 2D DSA images. It provides ground-truth extrinsic and intrinsic parameters, and 3D automated landmarks along the center line of vascular structures. This dataset was also used for training (simulated DRRs) and testing.

Key Insights and Ideas

  • Real-time is Achievable: The paper demonstrates that achieving state-of-the-art registration accuracy is possible within strict real-time constraints (20ms inference time), which was previously a significant bottleneck.
  • Automation is Key: Eliminating manual landmark annotation is crucial for scalability and clinical adoption. The proposed automated landmark extraction method shows performance comparable to manual annotations.
  • Domain Gap is Manageable: While a domain gap exists between synthetic DRRs and real X-rays, aggressive data augmentation, particularly the novel plasma and SP augmentations, significantly bridges this gap and improves robustness.
  • Multi-stage Refinement Works: The two-stage CNN architecture effectively leverages a fast initial prediction followed by a learned refinement step, outperforming single-stage regression and offering a better speed-accuracy tradeoff than iterative optimization.

Key Results

  • Speed: LXPose achieves a mean inference time of 20ms, a two-order-of-magnitude reduction compared to methods requiring several seconds. This comfortably meets the 40ms real-time constraint.
  • Accuracy: On the DeepFluoro dataset, LXPose achieved a mean projection distance (mPD) of 0.98 mm (with manual landmarks), comparable to the best offline methods (e.g., xvr at 0.62 mm) but significantly faster. On the Ljubljana dataset, it achieved 0.87 mm mPD.
  • Comparison: LXPose significantly outperforms other real-time baselines like PoseNet and xvr-reg, with gaps of 3.09 mm and 76.72 mm in mPD on DeepFluoro, respectively.
  • Automatic Landmarks: Performance with automatically extracted landmarks was comparable to manually annotated ones, with only a marginal performance drop.
  • Domain Adaptation: Ablating the plasma and SP augmentations led to a substantial increase in projection error (up to 5 mm), highlighting their importance.

Limitations and Future Work

  • Patient-Specific Training: The current approach requires training a new model for each patient, which is time-consuming. Future work could explore training on multi-patient data to increase generalization.
  • Low-Dose CTs: The current work uses high-resolution CTs. Future research could investigate the method's robustness on low-dose CTs, which are used to reduce radiation exposure, potentially by simulating lower-quality images.
  • Surgical Instruments and Motion: The method does not explicitly handle surgical instruments or non-rigid motion (e.g., respiratory or cardiac dynamics). Future work will assess performance with surgical instruments and explore extensions for non-rigid motion.
  • Regulatory and Clinical Applicability: Patient-specific models raise regulatory challenges. Future work will focus on developing automated reliability checks and fallbacks to standard methods if confidence is low.