Tissue classification using depth-dependent ultrasound time series analysis: in-vitro animal study
Abstract
Time series analysis of ultrasound radio-frequency (RF) signals has been shown to be an effective tissue classification method. Previous studies of this method for tissue differentiation at high and clinical-frequencies have been reported. In this paper, analysis of RF time series is extended to improve tissue classification at the clinical frequencies by including novel features extracted from the time series spectrum. The primary feature examined is the Mean Central Frequency (MCF) computed for regions of interest (ROIs) in the tissue extending along the axial axis of the transducer. In addition, the intercept and slope of a line fitted to the MCF-values of the RF time series as a function of depth have been included. To evaluate the accuracy of the new features, an in vitro animal study is performed using three tissue types: bovine muscle, bovine liver, and chicken breast, where perfect two-way classification is achieved. The results show statistically significant improvements over the classification accuracies with previously reported features.
BibTeX
@inproceedings{imani2011tissue,
author = {Farhad Imani and Mohammad I. Daoud and Mehdi Moradi and Purang Abolmaesumi and Parvin Mousavi},
title = {{Tissue classification using depth-dependent ultrasound time series analysis: in-vitro animal study}},
booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
year = {2011},
doi = {10.1117/12.877845},
}Ask Althea to put this paper to the test.