Time-Resolved Fluorescent Immunochromatography of Aflatoxin B1 in Soybean Sauce: A Rapid and Sensitive Quantitative Analysis

Du Wang, Zhaowei Zhang, Peiwu Li, Qi Zhang, Wen Zhang2016 · DOI

TL;DR

This paper introduces a novel time-resolved fluorescent immunochromatographic (TRFIA) strip test for the rapid and sensitive detection of aflatoxin B1 (AFB1) in soybean sauce. By combining time-resolved fluorescence with immunochromatography, this method offers a cost-effective, on-site screening tool that significantly outperforms existing methods in terms of speed and sensitivity.

What Problem Does This Paper Address?

Aflatoxins, particularly aflatoxin B1 (AFB1), are highly toxic metabolites produced by certain fungi that can contaminate food products. AFB1 is a known human carcinogen, and regulatory bodies worldwide have set strict limits for its presence in food. Soybean sauce, a widely consumed condiment, is susceptible to AFB1 contamination. Current methods for AFB1 detection, such as High-Performance Liquid Chromatography (HPLC), are accurate but require extensive sample preparation, expensive equipment, and skilled personnel, making them unsuitable for rapid, on-site screening. While some immunosensing methods exist, a quantitative, rapid, and sensitive assay specifically for AFB1 in complex matrices like soybean sauce has been lacking. This paper aims to fill this gap by developing a portable, cost-effective, and highly sensitive TRFIA strip test.

Methods

The core of this work is the development of a time-resolved fluorescent immunochromatographic (TRFIA) assay. The method utilizes Europium (Eu)-nanospheres conjugated with antibodies against AFB1 as fluorescent probes. These Eu-nanospheres are designed to amplify the fluorescent signal, overcoming the inherent low fluorescence of Eu(III) ions.

The assay follows a competitive immunoassay format integrated into a lateral flow strip. Here's a breakdown of the key steps:

  1. Eu-Nanosphere Conjugation: Europium-doped nanospheres are functionalized with antibodies specific to AFB1. This conjugation is optimized for pH and antibody concentration to ensure optimal signal and stability.
  2. Immunochromatographic Strip Design: The strip consists of a sample pad, a nitrocellulose (NC) membrane with immobilized AFB2a-BSA (as the test line, T line) and rabbit anti-mouse IgG (as the control line, C line), and an absorbent pad.
  3. Assay Procedure: A prepared soybean sauce sample extract is mixed with the Eu-nanosphere-antibody probe. This mixture is then applied to the sample pad of the IC strip.
  4. Lateral Flow and Detection: If AFB1 is present in the sample, it competes with the immobilized AFB2a-BSA on the T line for binding to the Eu-nanosphere-antibody probes. This competition dictates the amount of probe captured on the T line. Unbound probes are captured by the anti-mouse IgG on the C line.
  5. Time-Resolved Fluorescence Detection: After a short incubation period (6 minutes), the strip is analyzed using a portable TRFIA apparatus. The fluorescence signal from the T and C lines is measured under specific excitation and emission wavelengths, with a delay to minimize background fluorescence.
  6. Quantitative Analysis: The ratio of the fluorescence intensity on the T line to the C line (T/C ratio) is used to quantify the AFB1 concentration. A lower T/C ratio indicates a higher concentration of AFB1 in the sample.

The method also involved optimizing sample preparation, particularly the extraction solvent (70% methanol) and dilution ratios to mitigate matrix interference from complex soybean sauce components.

Datasets

The study evaluated the TRFIA method using:

  • AFB1 standard solutions: Used to construct calibration curves and determine assay parameters like LOD and dynamic range.
  • AFB1-free soybean sauce samples: Used for spiking experiments to assess recovery and for determining the LOD.
  • Real soybean sauce samples: Purchased from local markets in Wuhan, China. A total of 53 samples were analyzed by TRFIA, and 10 of these were further analyzed by HPLC for validation.

No new benchmark datasets were introduced; rather, the study focused on validating the TRFIA method against established HPLC protocols.

Key Insights and Ideas

  • Synergistic Power of TRFIA: The combination of time-resolved fluorescence and immunochromatography proves highly effective for sensitive and rapid detection. The nanosphere encapsulation of Eu(III) is crucial for signal amplification.
  • Overcoming Matrix Complexity: Soybean sauce is a complex food matrix. The study highlights the importance of optimizing sample preparation (e.g., solvent choice and dilution) to minimize interference and ensure accurate results.
  • On-Site Screening Potential: The development of a portable TRFIA apparatus makes this method suitable for rapid, on-site screening, which is a significant advantage over laboratory-bound techniques.
  • High Specificity: The antibody used (Anti-AFB1 mAb3G1) demonstrated high specificity to AFB1, with very low cross-reactivity to other aflatoxins (B2, G1, G2), ensuring reliable detection.

Key Results

  • Sensitivity: The TRFIA method achieved a low Limit of Detection (LOD) of 0.1 µg·kg⁻¹ for AFB1 in soybean sauce.
  • Dynamic Range: The assay demonstrated a wide quantitative dynamic range of 0.3–10.0 µg·kg⁻¹.
  • Accuracy and Recovery: Recoveries ranged from 87.2% to 114.3%, indicating good accuracy. The coefficient of variation (CV) for repeatability was low, ranging from 3.4% to 8.6%.
  • Speed: The entire assay, including sample incubation, takes only 10 minutes.
  • Correlation with HPLC: The TRFIA results showed excellent correlation (R² > 0.99) with High-Performance Liquid Chromatography (HPLC) analysis, validating its accuracy.
  • Real Sample Analysis: Out of 53 analyzed soybean sauce samples, 64.2% were found to contain AFB1, with concentrations ranging from 0.31 to 12.5 µg·kg⁻¹. Only 3 samples exceeded the 5.0 µg·kg⁻¹ regulatory limit set by China.

Limitations and Future Work

  • Limitations:
    • The study focused on AFB1. While the antibody showed some specificity to other aflatoxins, further validation for simultaneous detection of multiple aflatoxins might be beneficial.
  • The sample preparation, while optimized, still involves an extraction step. Further simplification of sample preparation could enhance its on-site applicability.
  • Future Work:
    • The authors suggest that the TRFIA method can be widely utilized in rapid on-site analysis of AFB1 in soybean sauce samples.
  • Expansion to other food matrices and mycotoxins could be a logical next step.
  • Further miniaturization and integration of the portable detection device could enhance its user-friendliness for field applications.

Summary updated