The steps about the three Dimensional Modeling of Bioprosthetic Heart Valves Based on CT Images

Dan Tang, Quan Yuan, BingShen Shen, Zhichao Wang, Hongwei Zhu2016 · DOI

Abstract

In this paper, we use reverse modeling software to complete the reconstruction of the aortic valve based on the modeling principles of the human heart valve. By using cardiac CT scanning to obtain DICOM format data and then using the 3D software Mimics for segmentation, editing, and smoothing, we generate a realistic and precise structural heart valve model and peripheral blood surface model in STL format. These models are then integrated in Geomagic Studio for further optimization. Finally, the model is imported into ANSYS Workbench for finite element analysis preparation, focusing on the leaflets and the arterial wall.

Introduction

The human heart valve presents a significant geometric challenge for engineers due to its complex, highly curved, and dynamic surfaces. Creating digital models that are both anatomically accurate and mathematically suitable for computer simulations is difficult. Standard medical imaging provides raw data, but a significant gap exists between a stack of 2D CT slices and a functional 3D mesh used for simulating blood flow or structural stress. This paper addresses that gap by providing a step-by-step methodology to reconstruct the aortic valve and surrounding arterial walls with the precision required for engineering-grade analysis.

Methodology

The authors propose a "reverse modeling" pipeline that utilizes three distinct software environments to handle different stages of the reconstruction:

  1. Image Acquisition & Segmentation (Mimics): Raw DICOM data from a cardiac CT scan is imported. The authors use thresholding (setting a minimum value of 300) to isolate the valve tissue. They then employ regional growth algorithms to separate the valve from non-connected structures and image editing to fill anatomical "holes" that would otherwise break a simulation.
Multi-planar reconstruction (MPR) view in Mimics showing axial, coronal, sagittal, and 3D views used for segmentation.
  1. Surface Optimization (Geomagic Studio): Because valve leaflets have high curvature, the initial mesh is often too noisy. The authors move the model into Geomagic to perform "reverse engineering" tasks: cleaning point clouds, repairing polygons, and smoothing surfaces to ensure the model meets the strict requirements of structural analysis.
  2. Leaflet Separation: Using "shear" and "shell" commands, the authors manually isolate the three individual valve leaflets and the arterial wall, saving them as separate STL files.
  3. Simulation Prep (ANSYS Workbench): The final optimized models are imported into ANSYS, where they are prepared for Finite Element Analysis (FEA).

Data Acquisition

The study relies on a single, high-resolution dataset from a 30-year-old healthy male with no history of cardiac disease. Scanning was performed using a GE Medical Systems Discovery CT750 HD, producing 209 slices with a thickness of 0.625mm and a pixel size of 0.488.

Results and Discussion

The primary result is a successful, reproducible pipeline for creating 3D bioprosthetic models. The process successfully converted 209 2D slices into a set of 3D STL files representing the aortic wall and three distinct valve leaflets.

The final 3D reconstructed surface model of the valve region ready for further optimization.

Key insights from this workflow include:

  • The Hybrid Software Approach: No single software tool is sufficient. Medical software (Mimics) is best for biological segmentation, but industrial reverse-engineering software (Geomagic) is necessary to handle the complex curvatures of heart leaflets for engineering purposes.
  • Thresholding Optimization: The authors identify a specific threshold value (300) as the "sweet spot" for isolating aortic valve tissue from the surrounding blood and muscle in CT data.
  • Mesh Quality: For FEA, the mesh must be optimized (smooth, hole-free, and logically divided) to prevent simulation errors, emphasizing quality over raw data fidelity.

Conclusion

While this study demonstrates a robust workflow, it was limited to a single healthy subject and requires significant manual intervention. Future work should focus on validating this workflow across different pathologies (such as calcified valves) and automating the segmentation steps. This methodology serves as a critical foundation for subsequent dynamic FEA simulations to study how these models behave under physiological blood pressure.

Summary updated

Abstract

In this paper, we use the reverse modeling softwares to complete the reconstruction of the aortic valve based on the modeling principle of the human heart valve. By using the cardiac CT scanning to get the Dicom format data and then using the three-dimension software Mimics to do the segmentation, edit, smooth processing to get the more realistic, the more precise structural heart valve model and peripheral blood in STL format surface model appearance. Then we mix up the two models in Geomagic Studio for further optimization. Finally we can get the model of blood and heart valves After importing the model into ANSYS Workbench for further treatment, we can get the model of the blood and the arterial wall preparing for the next finite element analysis about the leaflets and the arterial wall
BibTeX
@misc{tang2016steps,
  author = {Dan Tang and Quan Yuan and BingShen Shen and Zhichao Wang and Hongwei Zhu},
  title = {{The steps about the three Dimensional Modeling of Bioprosthetic Heart Valves Based on CT Images}},
  year = {2016},
  doi = {10.2991/iceeecs-16.2016.8},
}

Loading related content...

Figure