The silicon drift detector for the IXO high-time resolution spectrometer

P. Lechner, C. Amoros, D. Barret, Pierre Bodin, Martin Boutelier, R. Eckhardt, C. Fiorini, E. Kendziorra, K. Lacombe, A. Niculae, Benjamin Pouilloux, R. Pons, D. Rambaud, Laurent Ravera, Christian Schmid, H. Soltau, L. Strüder, C. Tenzer, J. WilmsProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2010 · DOI

TL;DR

This paper introduces the Silicon Drift Detector (SDD) based High Time Resolution Spectrometer (HTRS) for the International X-ray Observatory (IXO). It's designed to handle extremely high X-ray count rates from compact astrophysical objects, enabling unprecedented studies of matter under extreme gravity and density with both high-time and energy resolution.

What Problem Does This Paper Address?

Observing the most extreme environments in the universe, such as the accretion flows around black holes and neutron stars, presents a significant challenge for X-ray astronomy. These sources are incredibly bright, producing X-ray fluxes that can overwhelm traditional detectors, leading to "dead time" where the instrument cannot record incoming photons. This limits our ability to perform precise timing and spectroscopic measurements, crucial for testing general relativity and understanding the physics of these objects. The IXO mission aims to tackle these questions, and the High Time Resolution Spectrometer (HTRS) is a key instrument designed to overcome the high count rate limitation.

Methods

The HTRS instrument utilizes a monolithic array of 31 Silicon Drift Detector (SDD) cells. SDDs are chosen for their ability to combine a large sensitive area with a small anode capacitance, which is critical for achieving both good energy resolution and high count rate capability. The core principle involves a carefully engineered electric field within the silicon that efficiently drifts charge carriers (electrons) generated by X-ray absorption towards a small collecting anode. This anode is integrated with a low-noise amplifier (a JFET) directly on the chip, minimizing parasitic capacitance and noise. The HTRS design specifically employs a circular array of these SDD cells, with the incoming X-ray spot spread out by placing the detector slightly out of focus. This distribution prevents any single cell from being saturated by the intense photon flux. The detector is logically divided into four independent "quadrants" for redundancy and readout management.

Datasets

Not discussed in the paper. The paper focuses on the instrument design and performance specifications rather than experimental data from astronomical observations.

Key Insights and Ideas

  • SDD for Extreme Count Rates: The paper highlights how the inherent design of Silicon Drift Detectors, particularly their fast charge collection and low capacitance, makes them uniquely suited for high count rate X-ray spectroscopy, a critical requirement for IXO.
  • Out-of-Focus Placement for High Flux: A clever strategy of placing the HTRS slightly out of focus is employed to spread the intense X-ray flux from bright sources across the entire 31-cell array, preventing individual cell saturation and enabling high count rate operation.
  • Integrated Readout for Low Noise: Integrating the first amplification stage (FET) directly onto the SDD chip is a key design choice to minimize noise and parasitic capacitance, crucial for achieving the required energy resolution.
  • Radiation Hardness: The paper addresses the expected radiation environment in space and presents data showing the SDDs are sufficiently radiation-hard for the mission's lifetime, with minimal degradation in energy resolution.

Key Results

  • Energy Resolution: The HTRS is specified to achieve an energy resolution of ≤ 150 eV (FWHM at 5.9 keV) at the beginning of its lifetime, with a goal of maintaining this within 2% even at high count rates.
  • Time Resolution: The instrument will provide high time resolution of 10 µsec.
  • Count Rate Capability: The HTRS is designed to handle count rates up to a goal of 2 x 10⁶ counts per second, equivalent to a 12 Crab source, with less than 10% dead time.
  • Quantum Efficiency: The detector exhibits a quantum efficiency of > 90% for energies between 0.5 and 10 keV.
  • Radiation Tolerance: Measurements show that the SDDs are not affected up to 10¹³ photons (Mo-K, 18 keV) absorbed in the sensitive area, far exceeding the expected flux. Modeled proton irradiation suggests the energy resolution will remain within specifications at the end of the mission lifetime.

Limitations and Future Work

  • Analog vs. Digital Readout: While an analog readout chain is the baseline, a digital readout strategy is under study. The analog option offers lower power consumption, while the digital option provides greater flexibility for software adaptation post-launch. A detailed tradeoff analysis is ongoing.
  • Baffle Development: The "Spider Web" baffle design is being refined to use a more absorbing material to further mitigate split events caused by photons absorbed near cell borders.
  • Prototype Development: The paper mentions that prototype versions of the sensor and readout electronics are being developed and will be available for combined testing soon, indicating ongoing development and validation.
  • Thermal Management: The sensor is cooled to -40 °C, and thermal decoupling is employed to maintain this temperature, which is critical for performance. The paper details the thermal interface and cooling strategy.

Summary updated

Abstract

The High Time Resolution Spectrometer (HTRS) is one of six scientific payload instruments of the International X-ray Observatory (IXO). HTRS is dedicated to the physics of matter at extreme density and gravity and will observe the X-rays generated in the inner accretion flows around the most compact massive objects, i.e. black holes and neutron stars. The study of their timing signature and in addition the simultaneous spectroscopy of the gravitationally shifted and broadened iron line allows for probing general relativity in the strong field regime and understanding the inner structure of neutron stars. As the sources to be observed by HTRS are the brightest in the X-ray sky and the studies require good photon statistics the instrument design is driven by the capability to operate at extremely high count rates. The HTRS instrument is based on a monolithic array of Silicon Drift Detectors (SDDs) with 31 cells in a circular envelope and a sensitive volume of 4.5 cm<sup>2</sup> &times; 450 &#956;m. The SDD principle uses fast signal charge collection on an integrated amplifier by a focusing internal electrical field. It combines a large sensitive area and a small capacitance, thus facilitating good energy resolution and high count rate capability. The HTRS is specified to provide energy spectra with a resolution of 150 eV (FWHM at 6 keV) at high time resolution of 10 &#956;sec and with high count rate capability up to a goal of 2·10<sup>6</sup> counts per second, corresponding to a 12 Crab equivalent source. As the HTRS is a non-imaging instrument and will target only point sources it is placed on axis but out of focus so that the spot is spread over the array of 31 SDD cells. The SDD array is logically organized in four independent 'quadrants', a dedicated 8-channel quadrant readout chip is in development.
BibTeX
@inproceedings{lechner2010silicon,
  author = {P. Lechner and C. Amoros and D. Barret and Pierre Bodin and Martin Boutelier and R. Eckhardt and C. Fiorini and E. Kendziorra and K. Lacombe and A. Niculae and Benjamin Pouilloux and R. Pons and D. Rambaud and Laurent Ravera and Christian Schmid and H. Soltau and L. Strüder and C. Tenzer and J. Wilms},
  title = {{The silicon drift detector for the IXO high-time resolution spectrometer}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {2010},
  doi = {10.1117/12.857260},
}
Figure