Abstract
Background: To explore the role of nickel-titanium (NiTi) shape memory alloy embracing fixators in determining the implantation angle of hip tumor prosthesis stems and analyze its efficacy. Methods: 36 patients with proximal femur tumors were treated with extended tumor resection and prosthetic replacement. 14 patients received prosthetic replacements with the embracing fixators fixing between the junction of the prosthesis stem and the femur temporarily, while the other 22 patients received the same replacements but without the fixators. Results: There were fewer cases of hip dislocation in the group receiving prosthetic replacements with the use of embracing fixators. Occurrence of deep infection presented no difference between the two groups. Better limb function and higher active range of motion (ROM) on abduction or flexion were also found in the group using embracing fixators. Conclusion: Ni-Ti shape memory alloy embracing fixators play a key part in assisting the accurate implantation angle of the prosthesis stem in prosthetic replacement. Patients have a lower chance of dislocation, better limb function, and higher active hip ROM.
Introduction
In orthopedic oncology, surgeons often face a high-stakes geometry problem when treating tumors of the proximal femur. Removing the tumor frequently requires an extended resection that eliminates natural anatomical landmarks, such as the greater and lesser trochanters. These markers are essential for determining the correct rotational angle (version) of the prosthetic stem.
Incorrect alignment by even a few degrees can lead to joint instability and a high risk of post-operative dislocation, which historically ranges from 8% to 33% in these complex surgeries. Traditionally, surgeons have relied on experience and "gut feel," making final adjustments while bone cement hardens rapidly—a process with no "undo" button. This study evaluates a "smart" surgical tool designed to provide a reversible, iterative method for optimizing prosthesis alignment.
Methods
This retrospective observational cohort study involved 36 patients treated at Changhai Hospital between 2012 and 2019 for various conditions, including osteosarcoma, chondrosarcoma, and bone metastasis. The patients were divided into two groups:
- Fixator Group (n=14): Received temporary fixation using a nickel-titanium (Ni-Ti) shape memory alloy embracing fixator.
- Control Group (n=22): Received traditional prosthetic replacement without the fixator.
The Ni-Ti Shape Memory Alloy Embracing Fixator
Nitinol is a "smart" metal that changes properties based on temperature. The surgical workflow utilized two phases:
- Martensite Phase (Cold): At 0–4°C, the fixator becomes malleable, allowing surgeons to spread its arms around the bone-prosthesis junction.
- Austenite Phase (Warm): When warmed to 40–50°C with sterile saline, the alloy "remembers" its shape and contracts, providing a tight grip.
This allows the surgeon to "test-drive" the joint's range of motion. If the stability is suboptimal, the fixator is cooled, the angle adjusted, and the fixator re-warmed. Once the ideal angle is found and marked, the fixator is removed, and the stem is permanently cemented.
Results
The use of the Ni-Ti fixator led to significant improvements in clinical outcomes without increasing surgical risks such as operation time or blood loss.
Complications and Stability
The most striking result was the impact on joint stability. The fixator group experienced a 0% dislocation rate, whereas the control group had a 27.3% dislocation rate (6 out of 22 patients), a statistically significant difference ($P=0.032$).
Functional Outcomes and Mobility
Patients in the fixator group also demonstrated superior functional recovery. The Musculoskeletal Tumor Society (MSTS) score was significantly higher in the fixator group (80.0% vs. 72.9%, $P=0.003$). Furthermore, the active range of motion (ROM) for both flexion and abduction was significantly improved, suggesting that optimized geometry facilitates better muscle recovery.
Discussion
The core innovation of this approach is the transition from a "one-shot" cementing process to an iterative, reversible one. By utilizing material science as a surgical assistant, surgeons gain a "preview" of the mechanical outcome before making permanent changes. The continuous lateral compressive force of the Nitinol alloy provides a stable temporary environment to assess soft tissue tension and joint mechanics.
However, the study has limitations. The sample size of 36 patients is relatively small, and the retrospective design may introduce bias, as surgical techniques naturally evolve over time. Additionally, while short-term stability is improved, long-term follow-up is required to determine if this precision translates to increased prosthesis longevity and reduced aseptic loosening.
Conclusion
The Ni-Ti shape memory alloy embracing fixator is an effective tool for assisting in the accurate implantation of hip tumor prostheses. By allowing for intraoperative adjustment and testing, it significantly reduces dislocation rates and improves functional outcomes for patients undergoing complex limb-salvage surgeries.