The GRAVITY metrology system: narrow-angle astrometry via phase-shifting interferometry

M. Lippa, Nicolas Blind, Stefan Gillessen, Yitping Kok, Johannes Weber, Frank Eisenhauer, O. Pfuhl, A. Janssen, M. Haug, F. Haußmann, S. Kellner, O. Hans, E. Wieprecht, Thomas Ott, L. Burtscher, R. Genzel, E. Sturm, Reiner Hofmann, Stefan Huber, David Huber, Stefanie Senftleben, A. Pflüger, Roland Greßmann, G. Perrin, K. Perraut, W. Brandner, C. Straubmeier, A. Amorim, Markus SchöllerProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2014 · DOI

TL;DR

This paper presents a novel metrology system for the GRAVITY instrument, designed to achieve unprecedented 10 microarcsecond (μas) narrow-angle astrometry. The system uses phase-shifting interferometry to precisely monitor internal optical path differences down to the nanometer level, overcoming non-linearities in the phase-shifting device with a robust phase-step insensitive algorithm.

What Problem Does This Paper Address?

Achieving extremely high-precision astrometry, like the 10 μas target for the GRAVITY instrument, requires measuring tiny angular separations between astronomical objects. This translates to detecting differential optical path differences (dOPDs) on the order of a few nanometers. To achieve this, all optical paths within the interferometer, from the telescopes to the instrument itself, must be incredibly stable. The GRAVITY instrument combines light from four telescopes to achieve this precision, but it's crucial to precisely track and compensate for any internal path differences that arise within the instrument. This paper addresses the challenge of accurately measuring and calibrating these internal dOPDs using phase-shifting interferometry, specifically focusing on overcoming the limitations of non-linearities in the phase-shifting components.

Methods

The core of the GRAVITY metrology system is a phase-shifting interferometry technique. A laser beam traces the optical paths of the science light in reverse, from the instrument's beam combiners back to the telescopes. At the telescope pupils, this metrology light interferes, creating fringe patterns. By precisely shifting the phase of one of the metrology beams in a controlled sequence (a four-step algorithm), the intensity of these fringe patterns is measured.

Traditionally, extracting phase information from these intensity measurements can be problematic due to non-linearities in the phase-shifting devices themselves. To overcome this, the authors developed and implemented a phase-step insensitive algorithm. This method doesn't rely on perfectly accurate, discrete phase steps. Instead, it uses a continuous scan of phase shifts and analyzes the resulting intensity patterns as forming an ellipse. By fitting this ellipse, the algorithm can robustly extract the phase information, even in the presence of non-linearities and environmental instabilities. A key component of this algorithm is a reliable ellipse fitting routine, which they achieve using a χ²-minimization approach.

Datasets

Not discussed in the paper. The paper focuses on the calibration of a specific metrology device and the development of an algorithm, rather than on training or evaluating on large datasets.

Key Insights and Ideas

  • Non-linearities are a major hurdle: The paper highlights that standard phase-shifting interferometry is significantly hampered by non-linear behavior in the phase-shifting components, which can lead to inaccurate phase extraction.
  • Ellipse fitting as a robust solution: The phase-step insensitive algorithm, by modeling the intensity data as an ellipse, provides a powerful way to extract phase information that is less sensitive to these non-linearities and environmental drifts.
  • Calibration is paramount for precision: The accuracy of the metrology system, and thus the astrometric precision, is directly tied to the accuracy of the phase shifter calibration. Even small errors in phase calibration can translate to significant astrometric errors.
  • 1 nm accuracy is achievable: The developed method demonstrates the capability to calibrate phase shifts with an accuracy of approximately 1 nm, which is crucial for meeting the 10 μas astrometric goal.

Key Results

The paper successfully demonstrates a method for calibrating the GRAVITY metrology phase shifter with an accuracy of approximately 1 nm (or λ/2000). This is a significant improvement over simpler linear scan methods, which yielded less precise results and struggled with non-linearities. The phase-step insensitive algorithm, coupled with ellipse fitting, was shown to provide unambiguous phase extraction and a higher precision (factor of two better than linear scan) in their tests. The authors conclude that this calibration fulfills the requirements for achieving 10 μas-level astrometry.

Limitations and Future Work

  • Environmental sensitivity: While the phase-step insensitive algorithm is less sensitive to environmental instabilities than linear scans, the paper acknowledges that factors like temperature fluctuations, vibrations, and laser power fluctuations can still influence measurements.
  • Model dependence: The accuracy of the phase-step insensitive algorithm relies on the robustness of the ellipse fitting routine. While they use χ²-minimization, the paper notes that direct analytical methods for ellipse fitting have severe drawbacks.
  • Future work: The authors plan to analyze the instrument's behavior under vacuum and cryogenic conditions (as GRAVITY will operate in such an environment) to assess potential improvements in calibration accuracy. They also mention that fiber differential delay lines, which are less sensitive to transmission dependencies, are being implemented and may require adaptation of the original phase extraction formulas. The ultimate goal is to use this calibrated system for unprecedented astrometric measurements.

Summary updated

Abstract

The VLTI instrument GRAVITY will provide very powerful astrometry by combining the light from four tele- scopes for two objects simultaneously. It will measure the angular separation between the two astronomical objects to a precision of 10 μas. This corresponds to a differential optical path difference (dOPD) between the targets of few nanometers and the paths within the interferometer have to be maintained stable to that level. For this purpose, the novel metrology system of GRAVITY will monitor the internal dOPDs by means of phase- shifting interferometry. We present the four-step phase-shifting concept of the metrology with emphasis on the method used for calibrating the phase shifts. The latter is based on a phase-step insensitive algorithm which unambiguously extracts phases in contrast to other methods that are strongly limited by non-linearities of the phase-shifting device. The main constraint of this algorithm is to introduce a robust ellipse fitting routine. Via this approach we are able to measure phase shifts in the laboratory with a typical accuracy of λ=2000 or 1 nm of the metrology wavelength.
BibTeX
@inproceedings{lippa2014gravity,
  author = {M. Lippa and Nicolas Blind and Stefan Gillessen and Yitping Kok and Johannes Weber and Frank Eisenhauer and O. Pfuhl and A. Janssen and M. Haug and F. Haußmann and S. Kellner and O. Hans and E. Wieprecht and Thomas Ott and L. Burtscher and R. Genzel and E. Sturm and Reiner Hofmann and Stefan Huber and David Huber and Stefanie Senftleben and A. Pflüger and Roland Greßmann and G. Perrin and K. Perraut and W. Brandner and C. Straubmeier and A. Amorim and Markus Schöller},
  title = {{The GRAVITY metrology system: narrow-angle astrometry via phase-shifting interferometry}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {2014},
  doi = {10.1117/12.2054637},
}
Figure