The dissipated power in atomic force microscopy due to interactions with a capillary fluid layer
Lists
Only you see this
Abstract
We study the power dissipated by the tip of an oscillating micron-scale cantilever as it interacts with a sample using a nonlinear model of the tip-surface force interactions that includes attractive, adhesive, repulsive, and capillary contributions. The force interactions of the model are entirely conservative and the dissipated power is due to the hysteretic nature of the interaction with the capillary fluid layer. Using numerical techniques tailored for nonlinear and discontinuous dynamical systems we compute the exact dissipated power over a range of experimentally relevant conditions. This is accomplished by computing precisely the fraction of oscillations that break the fluid meniscus. We find that the dissipated power as a function of the equilibrium cantilever-surface separation has a characteristic shape that we directly relate to the cantilever dynamics. Even for regions where the cantilever dynamics are highly irregular the fraction of oscillations breaking the fluid meniscus exhibits a simple trend. Using our results we also explore the accuracy of the often used harmonic approximation in determining dissipated power.
BibTeX
@article{hashemi2008dissipated,
author = {Nicole N. Hashemi and Mark Paul and Harry Dankowicz and Moonjoo Lee and Wonho Jhe},
title = {{The dissipated power in atomic force microscopy due to interactions with a capillary fluid layer}},
journal = {Journal of Applied Physics},
year = {2008},
doi = {10.1063/1.2980057},
}Loading related content...