The BrainPET-7T Insert for Neuroscientific Applications: Performance with and without MRI
Abstract
We present the performance of a high-resolution, high-sensitivity, and UHF-MRI-compatible BrainPET insert for in vivo multimodal and multi-parametric neuroscientific imaging applications in humans. In this performance evaluation we studied mutual compatibility, PET-relevant imaging parameters, (spatial resolution, sensitivity, noise equivalent count rate, and coincidence time resolution), and MRI-relevant imaging parameters, (image homogeneity, image SNR, field homogeneity, and quality of spectra acquired for single-voxel <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">${ }^{1} \mathrm{H}$</tex> spectroscopy). Potential imaging applications for nuclei other than <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">${ }^{1} \mathrm{H}$</tex> (e.g. <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">${ }^{13} \mathrm{C},{ }^{19} ~\mathrm{F},{ }^{23} \text{Na}$</tex>, and <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">${ }^{31} \mathrm{P}$</tex>) were also considered. In addition, we acquired PET-only in ovo images of a chick embryo and compared the images to those obtained with a Siemens Inveon small animal PET/CT. The PET insert was shown to have a homogeneous spatial resolution of <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\approx 1.6 ~\text{mm}$</tex> paired with a sensitivity of <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$57.2 \text{kcps} / \text{MBq}$</tex>. MR image SNR is <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\approx 10 \%$</tex> lower when measured with the PET insert installed and running. However, spurious RF noise spectra with running PET are free of detectable interference for <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">${ }^{1} \mathrm{H}$</tex> and <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">${ }^{19} ~\mathrm{F}$</tex> within the relevant frequency bandwidths. No PET count rate loss was observed for most typical MR sequences. Thus, the BrainPET-7T provides a unique multimodal imaging tool for neuroscience with excellent MR and PET image quality.
BibTeX
@misc{lerche2025brainpet,
author = {Christoph Lerche and D. Niekaemper and J. Scheins and Lutz Tellmann and Cláudia Régio Brambilla and C.-H. Choi and Ezequiel Farrher and Jörg Felder and Stefan Krause and M. Schoeneck and D. Arutinov and R.D. Heil and W. Silex and Stefan van Waasen and Dirk Grunwald and M. Lennartz and T. Meurer and Ghaleb Natour and Bjoern Weissler and Florian Mueller and David Schug and Eike Gegenmantel and Harald Radermacher and O. Muelhens and Michael Lang and Pierre Gebhardt and J. Lefaucheur and Zhen Chen and Gary F. Egan and Volkmar Schulz and N. Jon Shah},
title = {{The BrainPET-7T Insert for Neuroscientific Applications: Performance with and without MRI}},
year = {2025},
doi = {10.1109/nss/mic/rtsd57106.2025.11287328},
}Ask Althea to put this paper to the test.
Loading related content...