The Biomechanical Response of the Cornea in Orthokeratology

Jinfang Wu, Wenxuan Fang, Huiwen Xu, Xiaode Liu, Dongliang Zhao, Qiguo RongFrontiers in Bioengineering and Biotechnology 2021 · DOI

Abstract

Orthokeratology (Ortho-k) is a widely utilized clinical method for myopia control, yet the precise biomechanical mechanisms governing corneal reshaping remain poorly understood. This study employs finite element analysis (FEA) to investigate the corneal response to Ortho-k lenses, quantifying changes in maximum stress across central and peripheral corneal regions. Numerical models were developed and validated, incorporating variations in corneal curvature, thickness, and target myopia reduction. Statistical analysis using median quantile regression and partial eta squared ($\eta^2$) values revealed that biomechanical stress is significantly influenced by lens design and initial corneal conditions. Specifically, target myopia reduction was the primary driver of central corneal stress ($\eta^2 = 0.9382$), while corneal curvature most significantly impacted peripheral stress ($\eta^2 = 0.5220$). These findings underscore the necessity of personalized lens fitting, particularly for patients with high myopia and steep corneas, to ensure safety and efficacy.

Introduction

Orthokeratology involves the overnight use of specialized rigid gas-permeable lenses to temporarily reshape the cornea, providing clear daytime vision without corrective eyewear. While its efficacy in myopia control—particularly in pediatric populations—is well-documented, the mechanical interaction between the lens and corneal tissue has largely remained a "black box." Clinicians often rely on empirical fitting guidelines, which may not account for the specific mechanical loads exerted on the cornea. Excessive stress can lead to complications such as corneal staining or structural instability. This research aims to bridge this gap by quantifying corneal stress distribution and identifying the key physiological and design factors that dictate the biomechanical response.

Methodology

The researchers utilized Finite Element Analysis (FEA) to simulate the mechanical environment of the eye during Ortho-k lens wear.

Model Development: A total of 20 idealized corneal models were constructed, featuring a range of curvatures (40.25 D to 44.25 D) and thicknesses (500 $\mu$m to 575 $\mu$m). To enhance anatomical accuracy, the cornea was modeled as a three-layer structure comprising the epithelium/Bowman’s layer, the stroma, and the Descemet’s/endothelium complex.

Lens Integration: Parameters for four-zone Ortho-k lenses (Alpha Corp) were derived from Optical Coherence Tomography (OCT) measurements. These measurements were translated into high-fidelity three-dimensional models for simulation.

Figure 1: (A) OCT B-Scan image of the ortho-k lens and (B) the corresponding three-dimensional model used for finite element analysis.

Simulation and Analysis: Eighty static simulations were conducted using Abaqus software. The models accounted for standard intraocular pressure (13 mmHg) and eyelid pressure (9 mmHg). To isolate the impact of different variables, the study employed median quantile regression and calculated partial eta squared values to rank the influence of myopia reduction targets, corneal curvature, and corneal thickness.

Results and Analysis

The study revealed a distinct shift in stress distribution based on the degree of correction. For low myopia targets (-2.00 D), the highest mechanical stress was observed in the peripheral cornea. However, as the correction power increased to -3.00 D and beyond, the maximum stress shifted to the central corneal region.

The Hierarchy of Influence: The statistical analysis identified a clear hierarchy among the factors affecting corneal stress:

  1. Myopia Reduction Target: This was the dominant factor for central corneal stress ($\eta^2 = 0.9382$). Moving from low to high correction targets resulted in a 2.42-fold increase in central stress.
  2. Corneal Curvature: Initial eye shape was the most significant predictor of peripheral stress ($\eta^2 = 0.5220$) and the second most significant for central stress ($\eta^2 = 0.5650$). Steeper corneas consistently exhibited higher stress levels.
  3. Corneal Thickness: Within normal physiological ranges, thickness had the least impact on the stress response ($\eta^2 \approx 0.19$ for both regions). This suggests that corneal curvature and lens power are more critical risk factors than corneal thickness in the context of Ortho-k biomechanics.

The maximum von Mises stress recorded across all simulations ranged from 0.00985 to 0.02513 MPa, providing a quantitative baseline for evaluating tissue safety.

Conclusion and Future Directions

This research demonstrates that the biomechanical response of the cornea is highly sensitive to both individual anatomy and lens design. The findings advocate for a shift toward personalized lens fitting, especially for high-myopia patients and those with steep corneas, to mitigate the risk of mechanical overstress.

Despite its insights, the study acknowledges certain limitations. The use of idealized spherical geometry and linear elastic material properties may not fully capture the aspheric and viscoelastic nature of the human cornea. Furthermore, the role of the tear film—specifically the "squeeze film forces" that contribute to reshaping—was not included in this static model. Future work should focus on integrating patient-specific 3D topography and dynamic fluid-structure interactions to develop more robust predictive tools for clinical practice.

Summary updated

Abstract

Orthokeratology has been widely used to control myopia, but the mechanism is still unknown. To further investigate the underlying mechanism of corneal reshaping using orthokeratology lenses via the finite element method, numerical models with different corneal curvatures, corneal thicknesses, and myopia reduction degrees had been developed and validated to simulate the corneal response and quantify the changes in maximum stress in the central and peripheral corneal areas during orthokeratology. The influence of the factors on corneal response had been analyzed by using median quantile regression. A partial eta squared value in analysis of variance models was established to compare the effect size of these factors. The results showed central and peripheral corneal stress responses changed significantly with increased myopia reduction, corneal curvature, and corneal thickness. The target myopia reduction had the greatest effect on the central corneal stress value (partial eta square = 0.9382), followed by corneal curvature (partial eta square = 0.5650) and corneal thickness (partial eta square = 0.1975). The corneal curvature had the greatest effect on the peripheral corneal stress value (partial eta square = 0.5220), followed by myopia reduction (partial eta square = 0.2375) and corneal thickness (partial eta square = 0.1972). In summary, the biomechanical response of the cornea varies significantly with the change in corneal conditions and lens designs. Therefore, the orthokeratology lens design and the lens fitting process should be taken into consideration in clinical practice, especially for patients with high myopia and steep corneas.
BibTeX
@inproceedings{wu2021biomechanical,
  author = {Jinfang Wu and Wenxuan Fang and Huiwen Xu and Xiaode Liu and Dongliang Zhao and Qiguo Rong},
  title = {{The Biomechanical Response of the Cornea in Orthokeratology}},
  booktitle = {Frontiers in Bioengineering and Biotechnology},
  year = {2021},
  doi = {10.3389/fbioe.2021.743745},
}
Figure