The Application Value of FUS-200 Full-automatic Urine Sediment Analyzer in Diagnosing Urinary Tract Infection
Lists
Only you see this
Abstract
Objective To study the application value of FUS-200 full-automatic urine sediment analyzer in diagnosing urinary tract infection(UTI).Method We determined the urine speciments and cultured the clear-catch midstream urine speciments from 219 patients suspected UTI.Using the results of urine bacteria culture as golden standard in diagnosing UTI,We coculated the sensibility,specificity,positive predictive value and negative predictive value of WBC counts and bacteria counts(BACT) detected by FUS-200 in diagnosing UTI.Results The sensitivity of WBC was high(88.0%),but the specificity was minimum(only 55.9%).The specificity of BACT was the highest(91.2%),but the sensitivity was slightly lower(77.1%).When WBC was combined with BACT,the sensitivity,specificity,positive predictive value and negative predictive value were very high(89.7%,88.4%,84.7%,91.5%).Conclusion FUS-200 urine sediment analyzer can be used as screening methods for clinical diagnosis of UTI,and it is fast,economical,accurate and repeatable.But UTI diagnosis need to based on the urine bacteria identification and drug sensitivity results.
BibTeX
@inproceedings{song2013application,
author = {Yang Song},
title = {{The Application Value of FUS-200 Full-automatic Urine Sediment Analyzer in Diagnosing Urinary Tract Infection}},
booktitle = {Guide},
year = {2013},
}Loading related content...