TEM based high resolution and low-dose scanning electron nanodiffraction technique for nanostructure imaging and analysis

Kyou-Hyun Kim, Hui Xing, Jian‐Min Zuo, Peng Zhang, Haifeng WangMicron 2015 · DOI

TL;DR

This paper introduces a new low-dose, high-resolution scanning electron nanodiffraction (SEND) technique implemented in a transmission electron microscope (TEM). This method allows for detailed structural analysis of nanostructures with minimal sample damage, opening new avenues for studying delicate materials.

What Problem Does This Paper Address?

Analyzing the intricate structures of nanomaterials often requires high-resolution imaging and diffraction techniques. Traditional methods like Selected Area Electron Diffraction (SAED) and Convergent Beam Electron Diffraction (CBED) have limitations in spatial resolution or require significant beam current, which can damage sensitive samples. Scanning Electron Nanodiffraction (SEND) offers a promising alternative by combining the high spatial resolution of scanning electron microscopy with the structural information from electron diffraction. However, existing SEND techniques often require long acquisition times and high beam currents, leading to issues like sample drift, heating, and damage, especially for low-dose applications. This paper addresses the need for a robust, low-dose SEND technique that can provide high-quality diffraction patterns from nanometer-sized grains without compromising sample integrity.

Methods

The authors developed a low-dose SEND technique using a JEOL 2100 TEM equipped with a LaB6 thermionic electron source. The core of their method involves:

  • Low-Dose Electron Probe Formation: They achieved electron beams with a full-width at half-maximum (FWHM) of 2-5 nm and a beam current of 0.1 pA or less. This was accomplished by using a strong condenser 1 lens setting and adjusting the gun heating current.
  • Automated Data Acquisition: A custom script written in the DigitalMicrograph (DM) scripting language was developed to automate both scan calibration and diffraction pattern acquisition. This script controls the TEM's operation, including beam shifting and data recording.
  • Scan Calibration: A crucial step involved calibrating the scanning vectors (Vx and Vy) to convert lens setting values into real distances in nanometers. This calibration was performed using a standard magnification in TEM mode and specific commands to measure beam shifts.
  • Low-Dose Diffraction Pattern Recording: The Gatan CCD camera was used for recording diffraction patterns. The script utilizes the SSCGainNormalizedBinnedAcquire command with 8x binning, resulting in a typical exposure time of 0.1 seconds per pattern. This allows for the acquisition of 20x20 diffraction patterns in approximately 5 minutes.
  • Microscope Alignment: Specific alignment procedures were implemented to minimize diffraction pattern shifts during beam scanning, ensuring accurate data analysis. This included optimizing intermediate lens settings to maintain a stable direct beam position.

Datasets

The paper utilizes a nanostructured Au disk as a test sample to demonstrate the performance of the developed low-dose SEND technique. No new benchmark datasets were introduced.

Key Insights and Ideas

  • Achieving Low-Dose SEND: The primary breakthrough is the successful implementation of a low-dose SEND technique that significantly reduces beam current and exposure time, thereby minimizing sample damage and drift.
  • High Spatial Resolution: The method enables the acquisition of diffraction patterns from nanometer-sized grains (2-5 nm probe size), offering unprecedented spatial resolution for nanostructure analysis.
  • Automated Workflow: The development of a custom DM script streamlines the entire process from scan calibration to data acquisition, making the technique more accessible and efficient.
  • Robustness to Sample Drift: The low acquisition time and minimal sample damage observed suggest that the technique is robust against sample drift, a common issue in high-resolution electron microscopy.
  • New Way for Selected Area Diffraction: The authors propose that SEND offers a novel approach to obtaining selected area electron diffraction patterns without the need for a physical aperture, which can be advantageous for analyzing very small or complex regions.

Key Results

  • The developed SEND technique successfully acquired diffraction patterns from a nanostructured Au disk with a probe size of 2-5 nm.
  • The diffraction patterns were recorded over an area of 210x210 nm² in 30x30 pixels, with a step size of 7 nm, over a period of 10 minutes.
  • Sample drift was measured to be 3 nm or less during a 10-minute acquisition period, demonstrating the low-dose technique's stability.
  • No sample damage or contamination was observed, confirming the effectiveness of the low-dose approach.
  • The integrated diffraction intensity from the acquired patterns was mapped to create images that remarkably reproduced the geometry of the Au nanodisk, validating the technique's imaging capability.
  • Averaged diffraction patterns from different regions showed enhanced quality, improving the reliability of diffraction pattern indexing.

Limitations and Future Work

  • Limitations: The paper acknowledges that the alignment procedures were developed for a specific TEM model (JEOL 2100), although the principles are expected to apply to other TEMs. The current implementation requires manual adjustment for beam shift if automated functions are not available.
  • Future Work: The authors are actively developing an automated diffraction indexing procedure for low-dose SEND patterns. This would further enhance the utility of the technique for structural analysis.

Summary updated