Synthesized Image Reconstruction for Post-Reconstruction Resolution Recovery

Laurence Vass, Andrew J. ReaderIEEE Transactions on Radiation and Plasma Medical Sciences 2023 · DOI

TL;DR

This paper introduces a novel post-reconstruction method for improving the spatial resolution of PET images. By synthesizing new PET data from an initial reconstruction and then re-reconstructing it with a modeled point spread function (PSF), the method achieves better resolution recovery than the standard Richardson-Lucy (RL) algorithm, without requiring proprietary scanner information.

What Problem Does This Paper Address?

Positron Emission Tomography (PET) imaging is a powerful tool for visualizing biological processes, but its spatial resolution is inherently limited. This poor resolution leads to inaccurate quantification of tracer uptake and degrades overall image quality. Factors like positron range, especially for higher-energy positron emitters, significantly contribute to this resolution loss. While incorporating a model of the system's point spread function (PSF) during reconstruction can improve resolution, it's technically challenging. It requires access to scanner-specific forward and back-projectors and raw sinogram data, which are often proprietary and difficult to obtain. Post-reconstruction methods, like the Richardson-Lucy (RL) algorithm, are more practical but suffer from rapid noise amplification, leading to poor image quality. This paper aims to bridge this gap by offering a practical post-reconstruction resolution recovery (RR) technique that achieves performance comparable to more complex PSF-based methods.

Methods

The core of the proposed method is a "synthesized image reconstruction" approach. Here's how it works:

  1. Initial Reconstruction: The process starts with an initial PET image obtained from a standard reconstruction algorithm (like MLEM) available on most PET scanners. This initial image, while having resolution losses, serves as the input.
  2. Synthetic Data Generation: The initial image is then forward-projected using a virtual scanner geometry. This generates synthetic sinogram data. The key here is that the user can define this virtual geometry, meaning scanner-specific information isn't needed.
  3. Synthesized Reconstruction with PSF: The generated synthetic sinogram data is then used in a new reconstruction problem. Crucially, this reconstruction does incorporate a modeled PSF (representing factors like positron range) within its system model. This allows for resolution recovery.

This approach effectively frames the resolution recovery task as an inverse problem that can be solved more gracefully, leading to better image quality, especially when early termination is applied to mitigate noise.

Datasets

The authors evaluated their method using:

  • Simulated 2D Digital Phantoms:
    • Modified Shepp-Logan phantom
    • BrainWeb phantom
    • Zubal thorax phantom
    • Derenzo-style microPET phantom (for spatial resolution assessment)
  • Experimental Preclinical Data:
    • 68Ga-THP-PAM (a bone-seeking radiopharmaceutical) images from a mouse, acquired using a Mediso Nanoscan PET/CT.

The simulated data included varying count levels (high, mid, low) to mimic different acquisition scenarios and injected activities. The PSF model used for simulations was a 2D Gaussian function representing the positron range of Gallium-68.

Key Insights and Ideas

  • Bridging the Gap: The method ingeniously bypasses the need for proprietary scanner data by creating a synthetic reconstruction environment. This makes advanced RR accessible even when scanner-specific projectors are unavailable.
  • Decelerated Convergence: By framing RR as a synthesized inverse problem, the method allows for a more controlled convergence, leading to better image quality at earlier termination points compared to the RL algorithm.
  • PSF Modeling is Key: While the initial reconstruction might not have PSF modeling, the synthesized reconstruction step explicitly incorporates it, enabling effective resolution recovery.
  • Practicality Over Complexity: The method offers a post-reconstruction solution, making it more practical and widely applicable than integrating PSF modeling directly into complex 3D reconstruction algorithms.
  • Robustness to Input Quality: The proposed method appears less sensitive to the number of iterations used for the initial reconstruction (a proxy for input image quality) compared to the RL algorithm.

Key Results

  • Improved Image Quality vs. RL: The proposed method consistently outperformed the Richardson-Lucy (RL) algorithm in both quantitative (RMSE) and qualitative assessments, especially at mid and low-count levels.
  • Comparable to MLEM+PSF: The performance of the proposed method was often comparable to MLEM with PSF modeling (MLEM+PSF), which is considered a reference standard. In some low-count scenarios, it even showed a slight performance advantage.
  • Quantitative Improvements: Normalized RMSE was significantly reduced by the proposed method compared to RL, particularly in low-count simulations (e.g., 124% for RL vs. 33% for the proposed method in one instance).
  • Qualitative Improvements: Visual inspection of simulated and preclinical images showed that the proposed method resolved finer details and reduced blurring more effectively than RL.
  • Preclinical Validation: Experimental data with 68Ga-THP-PAM demonstrated that the proposed method could resolve more features compared to the initial reconstruction and RL, with improvements visible at various iteration levels.

Limitations and Future Work

  • 2D Implementation: The current work focused on 2D data. Further research is needed to extend the method to a 3D implementation of the synthetic projectors.
  • PSF Model Specificity: The current PSF modeling primarily addressed positron range. Future work could incorporate other resolution-degrading factors and explore spatially variant PSF models that account for different tissue types.
  • Hyperparameter Optimization: While the study explored some hyperparameters (synthetic projection angles, line thickness), further work is needed to identify optimal virtual scanner geometries and their relative importance.
  • Regularization: The synthesized reconstruction did not include regularization, which could potentially further improve the bias-standard deviation tradeoff.
  • Comparison to AI: The authors acknowledge the rise of deep learning-based methods for PET RR and suggest that future work should compare their approach to these AI-based techniques.
  • Unknown Ground Truth: In real-world applications, the ground truth is unknown, making optimal iteration number selection for both the initial reconstruction and the synthesized reconstruction a challenge. The paper notes that the proposed method appears less sensitive to this than RL.

Summary updated

Abstract

Resolution recovery (RR) techniques in positron emission tomography (PET) imaging aim to mitigate spatial resolution losses and related inaccuracies in quantification by using a model of the system's point spread function (PSF) during reconstruction or post-processing. However, including PSF modeling in fully 3-D image reconstruction is far from trivial as access to the scanner-specific forward and back-projectors is required, along with access to the 3-D sinogram data. Hence, post-reconstruction RR methods, such as the Richardson-Lucy (RL) algorithm, can be more practical. However, the RL method leads to relatively rapid noise amplification in early image iterations, giving inferior image quality compared to iterates obtained by placing the PSF model in the reconstruction algorithm. We propose a post-reconstruction RR method by synthesizing PET data by a forward projection of an initial real data reconstruction (such reconstructions are usually available via a scanner's standard reconstruction software). The synthetic PET data are then used to reconstruct an image, but crucially now including a modeled PSF within the system model used during reconstruction. Results from simulations and real data demonstrate the proposed method improves image quality compared to the RL algorithm, whilst avoiding the need for scanner-specific projectors and raw sinogram data (as required by standard PSF modeling within reconstruction).
BibTeX
@article{vass2023synthesized,
  author = {Laurence Vass and Andrew J. Reader},
  title = {{Synthesized Image Reconstruction for Post-Reconstruction Resolution Recovery}},
  journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
  year = {2023},
  doi = {10.1109/trpms.2023.3247489},
}