Swept-Source-Based Chromatic Confocal Microscopy

Dawoon Jeong, Se Jin Park, Hansol Jang, Hyunjoo Kim, Jaesun Kim, Chang-Seok Kim2020 · DOI

TL;DR

This paper introduces a novel "swept-source-based chromatic confocal microscopy" (SS-CCM) system that significantly boosts imaging speed and sensitivity for 3D surface profiling. By replacing the conventional broadband light source and spectrometer with a wavelength-swept laser and a single photodetector, the SS-CCM achieves rapid, high-resolution 3D imaging with minimal measurement time.

What Problem Does This Paper Address?

Accurate 3D surface profiling is crucial in various fields, from microelectronics to biology. Traditional confocal microscopy (CM) offers high resolution but suffers from slow imaging speeds due to mechanical scanning. Chromatic confocal microscopy (CCM) attempts to overcome this by using the axial chromatic aberration of a lens to encode depth information spectrally, eliminating mechanical axial scanning. However, conventional CCM systems using broadband light sources and spectrometers face limitations: low signal intensity due to spectral splitting, insufficient spectrometer sensitivity for weak signals, and the need for signal integration, which further slows down acquisition. This paper addresses the need for a faster, more sensitive, and simpler 3D surface profiling technique by proposing and validating a swept-source-based approach.

Methods

The core innovation is the integration of a wavelength-swept laser (SS) into a chromatic confocal microscopy setup, creating the SS-CCM. Instead of a broadband light source and a spectrometer, the SS-CCM uses a wavelength-swept laser that rapidly scans through a range of wavelengths. This light is then passed through a chromatic dispersive lens array, which focuses different wavelengths at different axial positions. The reflected light is collected, and instead of being analyzed by a spectrometer, its intensity is measured by a single photodetector (PD). Since the wavelength is swept over time, the temporal signal from the PD directly corresponds to spectrally encoded depth information. This eliminates the need for a spectrometer, simplifying the system and enabling faster detection. The paper also details the design of a custom chromatic dispersive lens array using aspheric lenses to maximize chromatic aberration and minimize spherical aberration, simulated using CODE V.

Datasets

Not discussed in the paper. The experiments utilized standard samples like a cover glass stacked on a glass slide and a "US cent" coin for evaluation.

Key Insights and Ideas

  • Swept-Source for Speed and Simplicity: The key insight is that a wavelength-swept laser, when combined with a chromatic dispersive element, can directly provide time-encoded spectral information to a single photodetector, bypassing the need for a spectrometer. This drastically simplifies the system and accelerates data acquisition.
  • Eliminating Spectrometer Bottlenecks: Spectrometers are often a bottleneck in CCM systems due to their limited speed and sensitivity, especially with low-intensity signals. Replacing it with a photodetector, which is inherently faster and more sensitive to overall light intensity, is a significant improvement.
  • High-Speed 3D Imaging: The SS-CCM system demonstrates the capability for extremely fast axial scanning (5 ms per pixel), enabling rapid 3D surface profiling even for complex features.
  • Stable Performance: The SS-CCM exhibits stable performance across a wide range of acquisition times, down to the minimum achievable 5 ms, highlighting its robustness.

Key Results

  • Axial Range: The SS-CCM system achieved an axial chromatic focal distance range of approximately 360 µm for the 770–820 nm swept wavelength range.
  • High-Speed Imaging: High-speed 3D surface profiling images of a cover glass and a coin were successfully obtained with a measurement time of just 5 ms per pixel.
  • Lateral Resolution: The system demonstrated excellent lateral resolution of a few microns.
  • Axial Resolution: The measured average FWHM for axial resolution was 131.452 µm for SS-CCM, comparable to conventional CCM (139.308 µm), though noted as lower than some state-of-the-art systems.
  • Stability: The SS-CCM showed stable axial position measurements even at the minimum acquisition time of 5 ms, while the conventional CCM showed larger variations and instability at shorter acquisition times.

Limitations and Future Work

  • Axial Resolution: The authors acknowledge that the axial resolution (around 130 µm FWHM) is relatively lower compared to some advanced systems (tens of nanometers). They suspect this is influenced by the objective lens design, particularly an aspheric lens that might not have been optimally designed for high NA.
  • Future Work: The authors suggest extending this technology to other spectral regions, such as the O-band (around 1310 nm) and C-band (around 1550 nm), where swept-source lasers and photodetectors are commercially available. They highlight that this approach would be particularly valuable in these longer-wavelength regions, where InGaAs-based spectrometers are expensive and have fewer pixels compared to Si-based spectrometers used for the 800 nm band.