Sure-based parameter selection for parallel MRI reconstruction using GRAPPA and sparsity

Daniel S. Weller, Sathish Ramani, Jon-Fredrik Nielsen, Jeffrey A. FesslerISBI 2013 · OpenReview

Abstract

New methods have been developed for parallel MRI reconstruction combining GRAPPA and sparsity. One impediment to the practical application of such methods is selecting a regularization parameter that acceptably balances the contributions of GRAPPA and sparsity. We propose a broadly applicable Monte-Carlo-based approximation to Stein's unbiased risk estimate (SURE) for a suitable weighted mean-squared error (WMSE) metric. Applying this approximation to predict the WMSE-optimal tuning parameter for sparsity-based reconstruction, we are able to tune our parameter to achieve nearly MSE-optimal performance. In our simulations, we vary the noise level in the simulated data and use our Monte-Carlo method to tune the reconstruction to the noise level automatically.
BibTeX
@inproceedings{weller2013sure,
  author = {Daniel S. Weller and Sathish Ramani and Jon-Fredrik Nielsen and Jeffrey A. Fessler},
  title = {{Sure-based parameter selection for parallel MRI reconstruction using GRAPPA and sparsity}},
  booktitle = {ISBI},
  year = {2013},
  url = {https://openreview.net/forum?id=TLuuB6wfOs},
}