Study of a large range metrological atomic force microscope applied for calibration of a vertical PZT stage

S. H. Wang, Siew Leng Tan, Guowen XuProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2011

Abstract

In this paper, a method for the calibration of vertical PZT stages using a large range metrological atomic force microscope (LRM-AFM) is described. A vertical PZT stage is mounted onto the system and an optical-flat sample is attached on the top of the PZT stage. The AFM probe working in the contact mode is used as a null indicator which is sensitive to the movement of the optical-flat sample directly driven by the vertical PZT stage. At each state of the PZT stage movement, the AFM probe approaches to the test surface without horizontal scanning in the system. The displacement of the vertical stage is measured by the laser interferometer in LRM-AFM and the corresponding laser interferometer readings and the movements of the PZT stage are recorded. All collected data are retrieved to establish the relationship of the laser interferometer reading versus the PZT stage displacement. The results show that the system is capable of calibrating PZT stage in the range of up to 250 μm with an expanded uncertainty of less than 5 nm.
BibTeX
@inproceedings{wang2011study,
  author = {S. H. Wang and Siew Leng Tan and Guowen Xu},
  title = {{Study of a large range metrological atomic force microscope applied for calibration of a vertical PZT stage}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {2011},
  doi = {10.1117/12.883334},
}