Stratus OCT Quality Control in Two Multi-Centre Multiple Sclerosis Clinical Trials

John L. Keltner, Kimberly E. Cello, Laura J. Balcer, Peter A. Calabresi, Clyde Markowitz, John S. WernerNeuro-Ophthalmology 2011

TL;DR

This paper investigates the impact of quality control (QC) procedures on optical coherence tomography (OCT) data in multiple sclerosis (MS) clinical trials. By comparing two trials with different QC protocols, the study demonstrates that certified technicians and prompt data feedback significantly improve data quality, leading to fewer errors and more usable scans. This highlights the critical role of robust QC in ensuring reliable data for MS research.

What Problem Does This Paper Address?

Clinical trials in multiple sclerosis (MS) often rely on optical coherence tomography (OCT) to objectively measure changes in retinal nerve fiber layer (RNFL) thickness and macular volume. These measurements serve as crucial indicators of axonal injury and disease progression. However, the accuracy and reliability of OCT data are highly dependent on the quality of the scans acquired. In multi-center trials, where data is collected across various sites by different technicians, ensuring consistent and high-quality data can be challenging. This paper addresses the problem of variable OCT data quality in MS clinical trials by examining the influence of different quality control (QC) protocols on error rates and data usability. Specifically, it compares two trials with distinct QC approaches: one with no technician certification and limited feedback (Trial 1), and another with certified technicians and ongoing monitoring (Trial 2).

Methods

The study retrospectively analyzed 19,961 OCT scans from 981 subjects across two multi-center MS therapeutic trials (Trial 1 and Trial 2). The core of the methodology involved a comparative analysis of QC procedures and their impact on data quality.

  • Trial 1: Technicians were not certified for OCT, and data monitoring and feedback were limited and retrospective. Data was received in batches, and any feedback was provided after scans were already acquired.
  • Trial 2: Technicians underwent certification, which included a telephone examination on data collection requirements and submission of sample scans. The Reading Center continuously monitored data and provided feedback to site coordinators.

The QC assessment involved central readers evaluating scans for various error types, including:

  • Test Parameter Errors: Low signal strength (<7), extra scans, wrong scans, scans performed on the wrong instrument, and scans requiring redraws due to segmentation artifacts.
  • Shipping Errors: Exported data missing or missing scans.
  • Unusable Scans (Objective Method): Scans identified as missing, wrong scan, or performed on the wrong instrument.
  • Scan Quality for Proper Alignment and Usability (Masked Evaluation): A subgroup of scans was randomly selected and masked for review by two readers to assess proper alignment and usability. Scans were deemed unusable if they had low signal strength, artifacts, or poor alignment.

Datasets

The study utilized data from two multi-center MS therapeutic trials:

  • Trial 1: 18,733 OCT scans from 939 relapsing-remitting MS patients across 96 sites.
  • Trial 2: 2,178 OCT scans from 40 secondary progressive MS patients across 10 sites.

No new datasets were introduced; the study focused on analyzing existing data from these two trials.

Key Insights and Ideas

  • Certification Matters: The most striking insight is the direct correlation between technician certification and improved data quality. Certified technicians in Trial 2 produced significantly fewer errors and more usable scans.
  • Proactive QC is Crucial: Ongoing monitoring and prompt feedback from a dedicated reading center, as implemented in Trial 2, are essential for maintaining high data quality throughout a clinical trial. Retrospective feedback, as in Trial 1, is far less effective.
  • Impact on Error Types: The difference in QC protocols had a substantial impact on various error categories. Trial 2 showed dramatically lower rates of low signal strength, extra scans, and wrong instruments used compared to Trial 1.
  • Usable Data is Key: The ultimate goal of QC is to ensure usable data for analysis. Trial 2's approach led to a significantly lower percentage of unusable scans, making the data more reliable for drawing conclusions.
  • Scale of Trials: The paper acknowledges that larger trials (like Trial 1) present greater logistical challenges for QC, underscoring the need for well-established and efficient QC systems from the outset.

Key Results

  • Signal Strength: Trial 1 had a significantly higher rate of low signal strength (<7) at 23.2% (4,353/18,733 scans) compared to Trial 2 at 3.7% (81/2,178 scans).
  • Manual Entry Errors: Trial 1 suffered from a substantial 22.0% (4,114/18,733) rate of missing exported data requiring manual entry, while Trial 2 had 0% errors in this category.
  • Extra Scans: Trial 1 generated 13.5% (2,537/18,733) extra scans, whereas Trial 2 had only 1.3% (29/2,178).
  • Wrong Instrument Used: Trial 1 had 1.4% (269/18,733) scans performed on the wrong instrument, while Trial 2 had 0%.
  • Unusable Scans (Objective Method): Trial 1 had 6.4% (1,192/18,733) unusable scans, while Trial 2 had a much lower rate of 3.5% (76/2,178).
  • Scan Quality (Masked Evaluation): In a masked analysis of 1200 scans, 59.3% of Trial 1 scans were properly centered, compared to 74.7% in Trial 2. Consequently, 21% of Trial 1 scans were deemed unusable, versus only 6% in Trial 2.

Limitations and Future Work

  • Trial Size Discrepancy: The authors acknowledge the significant difference in the number of subjects and sites between Trial 1 (939 subjects, 96 sites) and Trial 2 (40 subjects, 10 sites). They note that larger studies are logistically more complex, implying that the observed differences might be amplified in larger-scale trials.
  • Retrospective Nature of Trial 1 Feedback: The limited and retrospective feedback in Trial 1 meant that improvements based on this feedback were difficult to measure. This highlights the importance of real-time feedback loops.
  • Future Work: The paper strongly advocates for the involvement of OCT reading centers from the initial planning stages of MS clinical trials. They emphasize the need for certified technicians, continuous monitoring, and rapid, constructive feedback to ensure optimal data quality. While not explicitly stated as future work, the findings lay the groundwork for implementing standardized, robust QC protocols in all future OCT-based MS trials.

Summary updated

BibTeX
@inproceedings{keltner2011stratus,
  author = {John L. Keltner and Kimberly E. Cello and Laura J. Balcer and Peter A. Calabresi and Clyde Markowitz and John S. Werner},
  title = {{Stratus OCT Quality Control in Two Multi-Centre Multiple Sclerosis Clinical Trials}},
  booktitle = {Neuro-Ophthalmology},
  year = {2011},
  doi = {10.3109/01658107.2011.557760},
}
Figure