Spectral Domain Z-scan Technique

Xi Zeng, Pengfei Qi, Pin Chen, Lie Lin, Weiwei LiuarXiv 2020 · DOI · arXiv

Abstract

Characterizing the nonlinear optical properties of various materials plays a prerequisite role in nonlinear optics. Among different methods, the well-known Z-scan technique and its modified versions have been recognized as simple and accurate methods for measuring both the real and imaginary parts of the nonlinear refractive index. However, all Z-scan methods based on detecting small beam variations put forward a severe restriction on the roughness of materials. Therefore, measuring nonlinear optical properties of highly scattering media remains challenging. Inspired by the innovation of the conventional Z-scan method that converts the wavefront phase shift to an easily measurable spatial pattern in the far-field, an alternative spectral domain Z-scan technique is presented. It has great potential for highly scattering media, based on the fact that scattering efficiency is insensitive to wavelength for Mie scattering when wavelengths are far smaller than the roughness. To demonstrate the advantages of the spectral domain Z-scan technique, the nonlinear refraction of polished and frosted slides was measured, agreeing well with previous reports.

Introduction: Overcoming the Scattering Barrier

In nonlinear optics, the nonlinear refractive index ($n_2$) is a fundamental parameter that dictates how light behaves at high intensities, leading to phenomena such as self-focusing. For decades, the standard Z-scan technique has been the primary tool for measuring $n_2$. By translating a sample through a laser focus and monitoring the transmittance through a far-field aperture, researchers can deduce the nonlinear phase shift.

However, the traditional Z-scan relies on spatial beam profiles. If a material is rough or highly scattering—such as biological tissue or unpolished glass—the spatial information is obliterated by random scattering, making traditional measurements impossible. This limitation has hindered the study of nonlinearities in complex, real-world materials.

Methodology: The Spectral Domain Approach

The proposed spectral domain Z-scan technique shifts the measurement from spatial distortions to spectral changes. When an intense femtosecond laser pulse passes through a nonlinear medium, it undergoes Self-Phase Modulation (SPM), which broadens and reshapes its spectrum.

Experimental Setup

The setup utilizes a femtosecond laser source and a Z-scan translation stage. Crucially, an integrating sphere is employed to collect the light exiting the sample. This device captures light from all angles, effectively "averaging out" the spatial scattering and ensuring that the spectrometer receives a representative sample of the total energy regardless of how much the sample scattered the light.

Experimental setup for the spectral domain Z-scan technique, featuring an integrating sphere and spectrometer.

The Spectral Aperture

Instead of a physical aperture in space, the researchers define a "spectral window"—a specific range of wavelengths (e.g., 800 nm to 803 nm). By monitoring the transmittance within this window as the sample moves through the focus, they can replicate the characteristic Z-scan curve. This "spectral aperture" is robust against scattering because Mie scattering efficiency is relatively constant across the narrow bandwidth of the laser pulse.

Experimental Results and Validation

To validate the technique, the researchers compared a polished fused silica slide with a frosted (highly scattering) slide. One side of the frosted slide was roughened to simulate a challenging material.

For the polished slide, the measured $n_2$ was approximately $2.7 \times 10^{-16} \text{ cm}^2/\text{W}$. For the frosted slide, the result was nearly identical at $2.73 \times 10^{-16} \text{ cm}^2/\text{W}$. Both values align with established literature for fused silica, proving that the spectral domain technique is immune to the scattering that would have rendered a traditional Z-scan useless.

Experimental Z-scan curves and corresponding spectra for the frosted fused silica slide.

Conclusion

The spectral domain Z-scan technique offers a powerful alternative for characterizing nonlinear optical properties in scattering media. By leveraging spectral signatures that survive scattering, it opens new doors for studying biological tissues and unpolished optical materials. Future work may extend this method to measure two-photon absorption and refine sensitivity through optimized spectral windowing.

Summary updated

Abstract

Characterizing the nonlinear optical properties of various materials plays a prerequisite role in nonlinear optics. Among different methods, the well-known Z-scan technique and the modified versions have been recognized as a simple and accurate method for measuring both the real and imaginary parts of the nonlinear refractive index. However, all the Z-scan methods based on detecting small beam variations put forward a severe restriction on the roughness of materials. Therefore, measuring nonlinear optical properties of highly scattering media still remain challenging. Inspired by the innovation of conventional Z-scan method that converting the wavefront phase shift to the easily measurable spatial pattern in far-field, the alternative spectral domain Z-scan technique was presented in this paper. It has a great potential for highly scattering medium, based on the scattering efficiency is insensitive to the wavelength for Mie scattering as the wavelengths are far smaller than the roughness. Moreover, to demonstrate the advantages of spectral domain Z-scan technique, the nonlinear refraction of polished slides and frosted slides was measured, which agrees well with previous reports.
BibTeX
@misc{zeng2020spectral,
  author = {Xi Zeng and Pengfei Qi and Pin Chen and Lie Lin and Weiwei Liu},
  title = {{Spectral Domain Z-scan Technique}},
  year = {2020},
  eprint = {2004.04528},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2004.04528},
}

Loading related content...

Figure