TL;DR
This paper introduces a novel, lightweight soft ankle exoskeleton designed to assist individuals with dropfoot and excessive inversion. By enabling simultaneous control of ankle dorsiflexion and eversion, the device shows promise in normalizing gait kinematics and improving foot clearance, addressing limitations of existing rigid orthoses.
What Problem Does This Paper Address?
Dropfoot, a common gait disorder resulting from stroke or neurological injuries, significantly impairs a person's ability to lift their foot during the swing phase of walking. This leads to difficulties in achieving adequate foot clearance, increasing the risk of tripping and falling. Individuals with dropfoot often adopt compensatory movements like "steppage gait" (lifting the knee higher) or "circumduction" (swinging the leg outward), which reduce walking efficiency and independence. Furthermore, dropfoot can be accompanied by excessive subtalar inversion, which exacerbates instability and the risk of ankle injury.
Traditional ankle-foot orthoses (AFOs) can help by restricting plantarflexion, but they also limit any remaining voluntary plantarflexor function, potentially leading to dependence. While powered exoskeletons offer more dynamic assistance, rigid designs are often heavy, bulky, and difficult to customize. Soft exoskeletons, on the other hand, offer a more compliant and potentially lighter alternative, but many existing designs focus solely on sagittal plane motion (dorsiflexion/plantarflexion), neglecting the crucial frontal plane control needed to address excessive inversion. This paper aims to bridge this gap by developing a soft ankle exoskeleton capable of assisting in both the sagittal and frontal planes simultaneously, with a focus on lightweight design and natural interaction.
Methods
The researchers developed a 2-Degrees-of-Freedom (DoF) soft ankle exoskeleton. The core of the design involves:
- Actuation System: Two motors are housed in a waist belt to minimize the load on the ankle. This placement leverages the body's center of mass for better weight distribution.
- Transmission System: Bowden cables transmit assistive forces to the medial and lateral sides of the forefoot, allowing for simultaneous control of dorsiflexion and eversion. This dual-plane control is a key innovation.
- Control Strategy: A hierarchical controller with three levels was implemented:
- High-level: Detects gait phases (stance/swing) using foot switches.
- Mid-level: Features two modes:
- Position Control: Generates desired ankle joint trajectories for assisting dorsiflexion/eversion and controlling plantarflexion.
- Force-Free Control: Compensates for the inherent resistance of the exoskeleton's mechanical components (friction, cable drag) during passive movements like plantarflexion.
- Low-level: Embedded in the motor drivers, it ensures precise tracking of the desired motor commands using position and current control loops.
To simulate dropfoot and excessive inversion in non-disabled participants, a 1-kg weight was attached to the shoe over the lateral forefoot.
Datasets
Not discussed in the paper. The study utilized data collected from experiments with non-disabled participants.
Key Insights and Ideas
- Dual-Plane Ankle Control: The ability to simultaneously control both sagittal (dorsiflexion/plantarflexion) and frontal (inversion/eversion) ankle movements with a relatively simple soft exoskeleton design is a significant advancement. This addresses the often-overlooked need for frontal plane stability in dropfoot.
- Waist-Belt Actuation: Placing the motors on the waist belt is crucial for a lightweight and less cumbersome ankle mechanism, directly impacting user comfort and metabolic cost.
- Force-Free Controller Effectiveness: The force-free controller demonstrated a notable ability to compensate for the exoskeleton's internal resistance, allowing for more natural ankle motion during passive phases like preswing.
- Simulated Impairment Validity: Using a weighted shoe to simulate dropfoot and inversion in non-disabled individuals provided a practical way to test the exoskeleton's functionality before clinical trials.
Key Results
- Seated Tests (Tests 1-3):
- Assisting Dorsiflexion/Eversion (Test 1): The exoskeleton achieved low errors (<1.5° in sagittal, <1.4° in frontal planes) when assisting ankle movement from an initial plantarflexed and inverted position to a neutral target.
- Controlling Plantarflexion (Test 2): The exoskeleton significantly decelerated and reduced ankle motion compared to the "no cables" condition, with final angles of -7.2° ± 0.5° in the sagittal plane and -1.2° ± 0.5° in the frontal plane.
- Resistance Compensation (Test 3): The force-free controller significantly increased ankle motion velocity (more than doubled in both sagittal and frontal planes) compared to the passive mode, though it was still about half as fast as without any exoskeleton resistance.
- Gait Tests (Test 4):
- The exoskeleton normalized key gait parameters affected by the simulated dropfoot. Specifically, it improved foot inclination and ankle inversion at initial contact, increased foot clearance height during swing, and shifted the ankle angle towards dorsiflexion in late swing.
- Maximum plantarflexion in preswing was only slightly restricted, indicating minimal interference with the propulsive phase of gait.
Limitations and Future Work
- Limited Clinical Validation: The study was conducted on non-disabled individuals with simulated impairments. Future work should involve testing the exoskeleton on individuals with actual dropfoot and excessive inversion to assess its clinical efficacy.
- Incomplete Inversion Constraint: The current design did not fully constrain ankle inversion, which could be an issue for individuals with a strong tendency to invert. This may require increased tension in the lateral Bowden cable.
- Constant Force-Free Controller Parameter (K₁): A single constant value for K₁ was used for all subjects and tests. Future research should explore subject-specific optimal values for K₁ to provide more tailored assistance.
- Minor Plantarflexion Restriction: While the force-free controller helped, some restriction of maximum plantarflexion in preswing was observed during gait.
- Future Directions: The authors suggest further exploration of different control profiles and optimization methods. Addressing the limitations regarding inversion constraint and individual parameter tuning will be crucial for future iterations.