Smartphone Biosensor System with Multi-Testing Unit Based on Localized Surface Plasmon Resonance Integrated with Microfluidics Chip

Zhiyuan Fan, Zhaoxin Geng, Weihao Fang, Xiaoqing Lv, Yue Su, Shicai Wang, Hongda Chen2020 · DOI

TL;DR

This paper presents a novel smartphone-based biosensor system (SBSM) that integrates localized surface plasmon resonance (LSPR) with microfluidics to enable rapid, multi-biomarker detection at the point-of-care. The system achieves high sensitivity and low limits of detection for cancer biomarkers, demonstrating its potential as a portable and user-friendly diagnostic tool.

What Problem Does This Paper Address?

Point-of-care testing (POCT) is crucial for timely diagnosis and patient monitoring, especially for chronic diseases like cancer. However, existing POCT solutions often lack the ability to simultaneously detect multiple biomarkers, requiring separate tests and increasing complexity. Furthermore, traditional laboratory-based biosensing equipment is bulky, expensive, and requires specialized training, hindering its widespread adoption in clinical settings or for home use. This paper addresses the need for a portable, cost-effective, and multi-analyte biosensing platform that leverages the ubiquity of smartphones for data acquisition and analysis.

Methods

The core of the SBSM is a microfluidic chip featuring an array of localized surface plasmon resonance (LSPR) sensor units fabricated on a quartz substrate. Gold nanoparticles (AuNPs) are deposited and annealed to create the LSPR sensing elements. These AuNPs are then functionalized with specific antibodies to detect target biomarkers.

A key innovation is the integration of these microfluidic chips with a modular smartphone attachment. This attachment includes:

  • Light Source: A backscattering LED light source to ensure uniform illumination.
  • Optics: A series of small lenses and a larger lens to focus light onto the sensor array and then onto the smartphone camera.
  • Grating: A transmission grating to disperse the light, allowing for spectral analysis by the smartphone's camera.
  • Smartphone Stage: A customizable stage to securely hold various smartphone models.

The smartphone's camera captures the dispersed light spectrum, which is then analyzed by a custom Python program. This program calculates the wavelength shift of the LSPR peak, which directly correlates with the refractive index change caused by biomarker binding. The system is designed for ease of assembly and use, requiring only a screwdriver for attachment and a simple one-step operation for testing.

Datasets

The paper utilizes the following for evaluation:

  • Sucrose Solutions: Used to calibrate the refractive index sensitivity of the LSPR chip.
  • Pure Protein Solutions: Bovine Serum Albumin (BSA), CA125, and CA15-3 were used for interference testing and to establish reference curves.
  • Clinical Serum Specimens: Nine clinical serum samples were tested to compare the SBSM's results with hospital-provided ELISA data.
  • Patient Serum Samples: 46 patient serum samples were analyzed to research the relationship between biomarkers and early-stage cancer.

No new benchmark datasets were introduced; rather, the system was evaluated against established methods like ELISA and clinical reference concentrations.

Key Insights and Ideas

  • Smartphone as a Spectrometer: The paper demonstrates that a smartphone, augmented with simple optical attachments, can effectively function as a basic spectrometer for LSPR-based biosensing.
  • Modular Design for Versatility: The modular nature of the attachments allows for easy adaptation to different smartphone models and customization for various sensing applications.
  • Microfluidics for Sample Efficiency: The use of microfluidic channels significantly reduces sample volume requirements (10 µL per channel), making testing more efficient and less invasive.
  • Multiplexing Capability: The chip design with multiple sensor units enables simultaneous detection of different biomarkers, a significant advantage over single-analyte tests.
  • Cost-Effectiveness: The use of readily available components and simpler fabrication methods (like annealing for AuNPs) suggests a lower cost compared to traditional laboratory equipment.

Key Results

  • Sensitivity: The LSPR chip demonstrated a sensitivity of 161.0 nm/RIU.
  • Limit of Detection (LoD): The system achieved LoDs of 4.2 U/mL for CA125 and 0.87 U/mL for CA15-3.
  • Correlation with ELISA: Results from clinical serum specimens showed good correlation with ELISA measurements, indicating the SBSM's reliability.
  • Interference Testing: The system showed minimal interference from BSA, a common protein in serum, even at higher concentrations than the target biomarkers.
  • Resolution: The SBSM achieved a resolution of 0.71 nm, comparable to laboratory spectrometers, though improvements are possible.
  • Test Time: The entire testing process, including antigen reaction, took approximately 1 hour.

Limitations and Future Work

  • Resolution Improvement: While comparable to lab equipment, the resolution could be further improved by using smaller micro-holes or gratings with more grooves, though this might impact light sensitivity and cost.
  • Sensitivity Enhancement: The sensitivity could be boosted by employing AuNPs fabricated with more advanced methods (e.g., interference-lithography, electron beam lithography).
  • Light Source and Camera: Using a more powerful light source and a higher-resolution camera could enhance the system's performance.
  • Long-Term Monitoring: The paper highlights the need for long-term monitoring for accurate diagnosis and management of early-stage cancers, suggesting the SBSM's suitability for this purpose.
  • Cost Reduction: Further cost reduction can be achieved through mass production techniques like polymer injection molding for chips and 3D printing for attachments.
  • Broader Biomarker Detection: The customizable nature of the chip suggests its potential for detecting other biomarkers beyond CA125 and CA15-3.

Summary updated

BibTeX