Slice-selective implementation of an adiabatic T2Prep sequence increases coronary artery conspicuity at 3T

Sahar Soleimanifard, Michael Schär, Jerry L. Prince, Robert G. Weiss, Matthias StuberJournal of Cardiovascular Magnetic Resonance 2012 · DOI

Abstract

Background: In non-contrast coronary MRA, the non-selective T2Prep 1-2 has been widely used for contrast enhancement between the coronary blood-pool and the myocardium. This non-selective pre-pulse affects the magnetization both inside and outside the field-of-view, resulting in a reduced steady-state magnetization of the inflowing blood, and consequently a penalty in SNR. We hypothesize that a slice-selective T2Prep would leave the magnetization of blood outside the imaged volume unaffected, and thereby minimize SNR penalty for inflowing blood. The purpose of this work was to implement a slice-selective T2Prep coronary MRA sequence and to assess the SNR gain quantitatively.

Introduction

Non-contrast coronary Magnetic Resonance Angiography (MRA) relies on effective contrast between the coronary arteries and the surrounding myocardium. A common technique to achieve this is the $T_2$Prep pulse, which leverages the differing $T_2$ relaxation times of blood and muscle. However, conventional $T_2$Prep pulses are non-selective, meaning they affect the magnetization of blood throughout the sensitive region of the receiver coil. This "pre-saturation" of blood before it enters the heart leads to a significant reduction in the signal-to-noise ratio (SNR) of the coronary arteries, as the blood entering the imaging volume has already had its longitudinal magnetization depleted.

Methods

To address the SNR penalty associated with non-selective pulses, a slice-selective adiabatic $T_2$Prep sequence was developed. The implementation involved several key modifications:

  1. Selective RF Pulses: Standard non-selective 90° RF pulses were replaced with slice-selective versions.
  2. Flow Compensation: To mitigate phase errors caused by blood flow during the application of selective gradients, flow-compensating gradients were integrated into the sequence.
  3. Optimized Inflow Timing: A 150ms time gap ($T_{gap}$) was introduced between the $T_2$Prep pulse and the imaging readout. This delay allows a larger volume of "fresh" blood—spins that have not been affected by the pre-pulse—to flow into the coronary arteries.
  4. Spatial Positioning: The $T_2$Prep volume was graphically aligned with the arterial axis, specifically avoiding the heart's ventricles to ensure the blood intended for imaging remained at equilibrium magnetization.

The sequence was evaluated on a Philips Achieva 3T scanner. Ten healthy volunteers underwent 3D free-breathing, navigator-gated coronary MRA under three conditions: without $T_2$Prep, with conventional non-selective $T_2$Prep, and with the proposed slice-selective $T_2$Prep. Numerical simulations based on the Bloch equations were also performed to validate the theoretical SNR gains.

Results

The slice-selective $T_2$Prep approach demonstrated superior performance across all quantitative metrics compared to the conventional non-selective method.

  • Signal-to-Noise Ratio (SNR): The slice-selective sequence achieved an SNR of $39.9 \pm 3.2$, a significant improvement over the $27.7 \pm 2.5$ observed with the non-selective version ($p < 0.02$).
  • Contrast-to-Noise Ratio (CNR): The ability to distinguish the coronary vessels from the surrounding tissue improved from $11.8 \pm 1.5$ to $16.1 \pm 1.6$ ($p < 0.05$).
  • Vessel Sharpness: Mean vessel sharpness increased from $48.5\%$ to $54.1\%$. This improvement is critical for the clear visualization of small anatomical structures and potential stenoses.

The experimental results closely matched the numerical simulations, confirming that the preservation of inflowing magnetization is the primary driver of the observed SNR increase.

Discussion and Conclusion

The transition from a global to a localized $T_2$Prep pulse effectively treats blood as a dynamic medium. By protecting the magnetization of blood in the ascending aorta, the slice-selective sequence ensures that the coronary arteries are filled with high-signal "fresh" blood during the acquisition window.

While the results are promising, certain limitations remain. The current implementation requires manual graphical selection of the $T_2$Prep volume, which may introduce operator variability. Additionally, the study was conducted on a small cohort of healthy volunteers; further validation in patients with coronary artery disease is necessary to assess performance under pathological flow conditions. Finally, the use of additional gradients and selective pulses at 3T necessitates careful monitoring of the Specific Absorption Rate (SAR).

In conclusion, the slice-selective adiabatic $T_2$Prep sequence significantly increases SNR and vessel conspicuity in 3T coronary MRA, offering a robust method for improving image quality in non-contrast cardiac imaging.

Summary updated

Abstract

Background In non-contrast coronary MRA, the non-selective T2Prep 1-2 has been widely used for contrast enhancement between the coronary blood-pool and the myocardium. This non-selective pre-pulse affects the magnetization both inside and outside the field-of-view, resulting in a reduced steady-state magnetization of the inflowing blood, and consequently a penalty in SNR. We hypothesize that a slice-selective T2Prep would leave the magnetization of blood outside the imaged volume unaffected, and thereby minimize SNR penalty for inflowing blood. The purpose of this work was to implement a slice-selective T2Prep coronary MRA sequence and to assess the SNR gain quantitatively.
BibTeX
@article{soleimanifard2012slice,
  author = {Sahar Soleimanifard and Michael Schär and Jerry L. Prince and Robert G. Weiss and Matthias Stuber},
  title = {{Slice-selective implementation of an adiabatic T2Prep sequence increases coronary artery conspicuity at 3T}},
  journal = {Journal of Cardiovascular Magnetic Resonance},
  year = {2012},
  doi = {10.1186/1532-429x-14-s1-o52},
}