Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent Diameter
Lists
Only you see this
Abstract
Using the mean CTDIvol from the whole scan range and Dw from the image at the center of the scan range provided an easily obtained estimate of SSDE for the whole scan range that agreed well with values from an image-by-image approach, with a root mean square difference less than 1.4 mGy (9%).
BibTeX
@inproceedings{leng2015size,
author = {Shuai Leng and Maria Shiung and Xinhui Duan and Lifeng Yu and Yi Zhang and Cynthia H. McCollough},
title = {{Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent Diameter}},
booktitle = {Radiology},
year = {2015},
doi = {10.1148/radiol.15142160},
}Loading related content...