Abstract

To evade legal controls, new psychoactive substances (NPS) are rapidly being derivatized, posing significant analytical challenges for forensic laboratories. This study developed an automated, green method for analyzing 51 indazole-type synthetic cannabinoids (SCs) in human urine and blood using online solid-phase extraction (online SPE) coupled with liquid chromatography-linear ion trap mass spectrometry (LC-LIT-MS). The method successfully distinguished five groups of isomers using multistage mass spectrometry ($MS^n$). Limits of detection (LODs) ranged from 0.02 to 1 ng/mL, with good linearity ($R^2 > 0.99$) and recoveries (83.47%–116.95%). This high-throughput approach provides a robust technical basis for rapid screening and quantitative analysis in forensic cases.

Introduction

Synthetic cannabinoids (SCs) represent a major class of New Psychoactive Substances (NPS) designed to mimic the effects of THC while bypassing legal restrictions. Following China's 2021 "class-wide" ban on SCs, indazole-type derivatives—known for their high potency—have continued to emerge through structural modifications. Forensic toxicologists face significant hurdles in detecting these substances due to sample complexity and the prevalence of structural isomers that standard mass spectrometry often fails to distinguish, leading to potential misidentifications.

Structural classification and substituent groups of the 51 indazole-type synthetic cannabinoids analyzed in this study.

Methods and Materials

The researchers developed an online SPE-LC-LIT-MS framework to automate sample processing and enhance identification accuracy.

Sample Preparation and Online SPE

Human blood and urine samples were treated with acetonitrile for protein precipitation. The resulting supernatant was diluted to 40% organic content—a "sweet spot" identified to prevent analytes from washing through the extraction column prematurely. The system utilized a valve-switching setup with an Oasis HLB column to automate enrichment and cleanup, feeding samples directly into the LC system.

LC-MS/MS Analysis

Separation was achieved on a ZORBAX Eclipse Plus C18 column using a gradient of acetonitrile and ammonium acetate. The core innovation lies in the use of Linear Ion Trap (LIT) mass spectrometry, which allows for $MS^3$ and $MS^4$ fragmentation. This "deep" fragmentation provides unique structural fingerprints necessary to differentiate between complex isomers that possess identical parent and secondary ions. The study utilized a library of 51 indazole-type SC standards, validated using matrix-matched calibration curves in blank human biological fluids.

Results and Discussion

Overcoming the Isomer Challenge

A primary achievement of this method is the successful differentiation of 10 compounds across 5 isomer groups. While $MS^2$ spectra for these compounds were nearly identical, $MS^3$ and $MS^4$ stages revealed distinct fragment ions, allowing for unambiguous identification. For example, the system could clearly distinguish between ADB-BUTINACA and AB-PINACA, which are typically difficult to separate.

Method Performance

The automated method demonstrated high sensitivity and reliability:

  • Sensitivity: Limits of detection (LOD) were 0.02–1 ng/mL; limits of quantification (LOQ) were 0.04–4 ng/mL for urine and 0.1–4 ng/mL for blood.
  • Linearity and Recovery: All analytes showed excellent linearity ($R^2 > 0.99$). Recoveries at LOQ levels ranged from 83.47% to 116.95%, with precision (RSD) between 2.29% and 13.40%.
  • Efficiency: By transitioning from offline to online SPE, the workflow eliminated labor-intensive steps like nitrogen evaporation and reconstitution, significantly reducing contamination risks and increasing sample throughput.

Conclusion

This research provides a highly sensitive, automated, and green analytical method for the forensic screening of indazole-type SCs. While LIT systems are primarily valued for qualitative identification, this study demonstrates their efficacy for quantitative analysis in complex biological matrices. Future work may expand this library to include indole-type SCs and their primary metabolites to further enhance forensic utility in real-world cases.

Summary updated

Abstract

To evade legal controls, new psychoactive substances (NPS), which have been designed as substitutes for traditional and synthetic drugs, are gradually dominating the drug market. Synthetic cannabinoids (SCs), which account for the majority of NPS, are rapidly being derivatized; consequently, controlling increasing abuse by merely listing individual compounds is difficult. Therefore, China has included the entire SC category of SCs listed as legal controlled substances since July 1, 2021. However, new SCs obtained through structural modification are still appearing and pose significant analytical challenges for forensic laboratories. Therefore, an efficient, green, and automated detection method is urgently required to provide technical support for the accurate screening actual samples. Meanwhile, the number of indazole-type SCs has increased sharply since 2013, which is ascribable to their stronger psychoactive effects. Indeed, forensic laboratories mainly analyze this key SC subclass. Therefore, in this study, we developed a new method for analyzing 51 indazole-type SCs in human urine and blood, which involves online solid-phase extraction (online SPE) as the preprocessing technology, with analysis performed using liquid chromatography-linear ion trap mass spectrometry. Deproteinization was achieved by adding acetonitrile, with dilution performed using 10 mmol/L ammonium acetate solution (pH 4.8) containing 0.1% formic acid. Samples were then analyzed directly using acetonitrile-10 mmol/L ammonium acetate aqueous solution (containing 0.1% formic acid) as the mobile phase. The mass-to-charge ratios of protonated molecular ions ([M+H]<sup>+</sup>) in the mass spectra acquired in full-scan mode, and the retention times in the chromatograms of the analytes were selected with the aim of monitoring the MS<sup>2</sup> ions of the various compounds. Characteristic fragment ions of the various SC structures were summarized, with five groups of isomers, each containing ten compounds, successfully distinguished using multistage mass spectrometry and their retention times. Multistage MS was used to qualitatively screen 51 indazole-type SCs, which were then quantitatively analyzed using MS<sup>2</sup> ion pairs (as quantitative ion pairs). The analytes exhibited limits of detection (LODs) of 0.02-1 ng/mL, with limits of quantification (LOQs) of 0.04-4 and 0.1-4 ng/mL in urine and blood, respectively. Linear fitting (weighting factor 1/<i>x</i>) revealed good linearity for each analyte within its respective linear range, with correlation coefficients (<i>R</i><sup>2</sup>) greater than 0.99 in both urine and blood. The validity of the analytical method was tested by determining precision and spiked recovery values (<i>n</i>=6). Recoveries of 83.47%-116.95% were obtained at LOQ levels, with precisions of 2.29%-13.40%. In addition, recoveries of 86.63%-113.38% and precisions of 0.58%-13.79% were obtained at low, medium, and high levels. The method described herein is not only easy to operate but also can be automated. Indeed, high-throughput sample analysis was achieved when sample extraction, enrichment, and analysis were performed in dynamic mode through valve switching. Meanwhile, the method exhibited good sensitivity and is applicable to a wider range of compounds than those previously reported; it also provides a scientific basis and technical support for the rapid screening and quantitative analysis of SCs in actual relevant cases.
BibTeX
@inproceedings{luo2025simultaneous,
  author = {Xuan Luo and Jun Zhang and Dan Zhu and Kejian Huang and Ning Yang and Xiaofeng Liu and Qiulian Luo},
  title = {{Simultaneous determination of 51 indazole-type synthetic cannabinoids in urine and blood by online solid-phase extraction-liquid chromatography-linear ion trap mass spectrometry}},
  booktitle = {Chinese},
  year = {2025},
  doi = {10.3724/sp.j.1123.2024.02008},
}