Screening Diabetic Retinopathy Using an Automated Retinal Image Analysis System in Mexico: Independent and Assistive use Cases

Alejandro Noriega, Daniela Meizner, Dalia Camacho‐García‐Formentí, Jennifer Enciso, Hugo Quiroz-Mercado, Virgilio Morales-Cantón, Abdullah Almaatouq, Alex PentlandmedRxiv 2020

Abstract

Diabetic Retinopathy (DR) is a leading cause of preventable blindness, particularly in regions like Mexico where specialist access is limited. This study evaluates the performance of the Automated Retinal Image Analysis (ARIA) system, a deep learning-based tool, in classifying retinal fundus images as referable or non-referable. We assessed ARIA's performance both as a standalone screening tool and as an assistive aid for 17 ophthalmologists under three conditions: solo, with a binary AI answer, and with an AI explanation (including attention maps). ARIA achieved an AUROC of 98.3% on Mexican patient cases, outperforming the average clinician. Furthermore, providing clinicians with a simple AI "second opinion" significantly improved their diagnostic sensitivity from 87.3% to 93.3%, demonstrating ARIA's potential to expand monitoring capacity in resource-constrained settings.

Introduction

In Mexico, diabetes prevalence is among the highest globally, yet only a small fraction of patients receive necessary eye screenings due to a shortage of specialists. Automated screening systems offer a scalable solution to this public health challenge. This research focuses on the ARIA system, evaluating its accuracy on local patient data and its effectiveness in a collaborative human-AI diagnostic workflow.

Methods and System Architecture

The ARIA system utilizes a deep convolutional neural network (CNN) to categorize retinal images. It distinguishes between "Non-Referable" cases (No DR or Mild NPDR) and "Referable" cases (Moderate to Severe NPDR and PDR), which require specialist intervention.

The ARIA system workflow: from clinical classification logic (A) to the technical deep learning pipeline (B).

To test the system's assistive value, 17 ophthalmology residents participated in a randomized experiment. They graded 100 Mexican patient images under three distinct conditions:

  1. Solo: Independent evaluation without AI assistance.
  2. ARIA Answer: Evaluation with a simple binary AI recommendation (Referable vs. Non-referable).
  3. ARIA Explanation: Evaluation with the AI recommendation plus a confidence score and an "attention map" highlighting suspicious regions.

Results

Standalone AI Performance

ARIA demonstrated high diagnostic robustness on the Mexican dataset, achieving an AUROC of 98.3%, a sensitivity of 95.2%, and a specificity of 90%. These results were superior to the average performance of the participating clinicians, validating the system's potential for independent screening in regions where specialists are unavailable.

Assistive Impact and the "Human-AI Dyad"

The study revealed a significant synergy when clinicians were assisted by AI. In the ARIA Answer condition, clinician sensitivity rose to 93.3%, up from 87.3% in the solo condition. This boost suggests that even a simple binary prompt can help residents catch subtle signs of retinopathy they might otherwise miss.

Interestingly, the more complex ARIA Explanation condition—which included attention maps and confidence scores—did not provide additional performance gains over the simple binary answer. While attention maps help visualize the AI's focus, participants reported that the extra information could sometimes be confusing, leading to a slight increase in false positives.

Examples of referable fundus images alongside AI-generated attention maps highlighting suspicious regions.

Conclusion

The ARIA system represents a substantial opportunity for improving early DR detection in Latin America. Whether used independently to expand screening reach or as an assistive tool to improve clinician consistency and sensitivity, ARIA provides a scalable path toward reducing diabetes-related blindness. Future work will focus on real-world integration into high-pressure clinical environments and handling varied image quality in primary care settings.

Summary updated

Abstract

Abstract Background The automated screening of patients at risk of developing diabetic retinopathy (DR) represents an opportunity to improve their mid-term outcome, and lower the public expenditure associated with direct and indirect costs of common sight-threatening complications of diabetes. Objective In the present study, we aim at developing and evaluating the performance of an automated deep learning–based system to classify retinal fundus images from international and Mexican patients, as referable and non-referable DR cases. In particular, we study the performance of the automated retina image analysis (ARIA) system under an independent scheme (i.e. only ARIA screening) and two assistive schemes (i.e., hybrid ARIA + ophthalmologist screening), using a web-based platform for remote image analysis. Methods We ran a randomized controlled experiment where 17 ophthalmologists were asked to classify a series of retinal fundus images under three different conditions: 1) screening the fundus image by themselves ( solo ), 2) screening the fundus image after being exposed to the opinion of the ARIA system ( ARIA answer ), and 3) screening the fundus image after being exposed to the opinion of the ARIA system, as well as its level of confidence and an attention map highlighting the most important areas of interest in the image according to the ARIA system ( ARIA explanation ). The ophthalmologists’ opinion in each condition and the opinion of the ARIA system were compared against a gold standard generated by consulting and aggregating the opinion of three retina specialists for each fundus image. Results The ARIA system was able to classify referable vs. non-referable cases with an area under the Receiver Operating Characteristic curve (AUROC), sensitivity, and specificity of 98%, 95.1% and 91.5% respectively, for international patient-cases; and an AUROC, sensitivity, and specificity of 98.3%, 95.2%, and 90% respectively for Mexican patient-cases. The results achieved on Mexican patient-cases outperformed the average performance of the 17 ophthalmologist participants of the study. We also find that the ARIA system can be useful as an assistive tool, as significant sensitivity improvements were observed in the experimental condition where participants were exposed to the answer of the ARIA system as a second opinion (93.3%), compared to the sensitivity of the condition where participants assessed the images independently (87.3%). Conclusions These results demonstrate that both use cases of ARIA systems, independent and assistive , present a substantial opportunity for Latin American countries like Mexico towards an efficient expansion of monitoring capacity for the early detection of diabetes-related blindness.
BibTeX
@misc{noriega2020screening,
  author = {Alejandro Noriega and Daniela Meizner and Dalia Camacho‐García‐Formentí and Jennifer Enciso and Hugo Quiroz-Mercado and Virgilio Morales-Cantón and Abdullah Almaatouq and Alex Pentland},
  title = {{Screening Diabetic Retinopathy Using an Automated Retinal Image Analysis System in Mexico: Independent and Assistive use Cases}},
  year = {2020},
  doi = {10.1101/2020.07.20.20157859},
  howpublished = {medRxiv},
}
Figure